7. INITIAL PREPARATION
Install the supplied batteries into the meter. See BATTERY REPLACEMENT for details.
To prepare the meter for field measurements close the USB communication socket with the attached
stopper.
Turn the meter ON by pressing ON/OFF key.
At start-up the meter will show the Hanna Instruments logo for a few seconds, followed by the
percentage indication of the remaining battery life.
Before connecting the probe for the first time, press
SETUP
and using / arrows navigate to
Date/Time. Press
Modify
and set the current date and time.
Note:
Verify time and date are properly set on meter prior to probe initialization.
1. Remove the
HI764113
from the carrying case. Remove stainless
steel guard from probe body if it was installed. Use care not to
leave fingerprints on the optical window.
2. Remove the syringe plunger; cut top off supplied sachet with
silicone grease and empty contents into the syringe. Using the
syringe, sparingly lubricate the o-ring with a thin film of the
supplied grease. Avoid getting any kind of grease or fingerprints
onto the optical window.
3. Remove the
HI764113-1
optical cap from the container. Align
the notched cutout arrow on the Smart Cap with the matching
guide on the probe body.
4. Slide and press the Smart Cap onto the
HI764113
body
until the cap snaps in place. Once the cap is installed,
it should not be removed unless a new cap is required.
5. Connect the
HI764113
probe to the
HI98198
meter
by plugging the DIN connector to the socket located on
the top of the meter.
6. Power the meter to initiate the cap timer.
Note:
Turn off the meter before connecting or
disconnecting the probe.