20
SETUP
Highlight
SOUR configuration
and press
Select
.
Press / keys to modify the selected parameter’s value.
Press
Prev/Next
to select a different parameter.
Press
Save
to save the new SOUR configuration.
Press
ESC
to leave without changing.
Parameters:
•
Min time
- the minimum time for the SOUR test.
Range: 1 to 3600 seconds.
•
Max time
- the maximum time for the SOUR test. The test will stop automatically when the
maximum time has elapsed.
Range: 1 to 3600 seconds.
•
Min start DO
- the minimum accepted DO value for starting the SOUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
•
Min end DO
- the minimum accepted DO value at the end of the test. If the DO reading at the
end of the SOUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
•
Total volume
- the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
•
Sample volume
- the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
•
Solids weight
: Total solids or Volatile suspended solids weight.
Range: 0.1 to 300.0 g/L.
•
SOUR @ 20 °C
: If this option to enabled the SOUR value is corrected to 20 °C.