background image

10

IX.   

IX.   

IX.   

IX.   

IX.   Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Use of a hood and protective clothing is strongly advised.

Connect sensors to pH/mV/ISE meter

Place meter in mV mode.

Place 50 mL of deionized water into a beaker with stir
bar and 2 mL of HI 4012-00 ISA. Add 50 mL of the
methanol/Formalin mixture. Add stir bar and place
on magnetic stirrer.

Place electrodes into prepared sample.

Add 1 mL of the stock standard HI 4012-01 (or .01 M
or .001 M  standard by  dilution)  to the beaker.
Record the mV value when reading has stabilized.

Add an additional 10 mL of standard  HI 4012-01 (or
same dilute standard)  to the solution. Record the mV
when reading has stabilized.  This value should be
more positive than the previous noted.

Determine the difference between the two mV values.
An acceptable value for this slope is +27 ± 4 mV.

X.   

X.   

X.   

X.   

X.   Corrective action

Corrective action

Corrective action

Corrective action

Corrective action

Verify protective cap has been removed (HI 4012).

Verify plastic film has been removed from inner stem
(HI 4112).

Verify electrodes are connected properly to meter and
meter is powered.

Verify standard has been properly stored. Remake
standards if appropriate.

If the sensor slope just misses the suggested slope
window, soaking the sensor in a dilute standard with
the methanol/formalin added may solve the prob-
lem. (

<

10

-3 

M standard).

A scratched, pitted or oxidized sensing surface can be
polished with HI 4000-70 polishing strip.  Cut off
approximately 1 inch of the micro-abrasive strip. Wear
protectice gloves.  Wet the frosted side with deionized
water

 

and place against damaged membrane of the

electrode.  Place your gloved thumb against the shiny
backing and slowly rotate back and forth while

Содержание HI 4112

Страница 1: ...I 4012 HI 4012 HI 4012 HI 4012 HI 4012 HI 4112 HI 4112 HI 4112 HI 4112 HI 4112 Lead Ion Selective Electrode Instruction Manual Half cell Combination LEAD SULFATE HI 4012 HI 4112 LEAD SULFATE COMBINATI...

Страница 2: ...t requires a separate reference The HI 4112 is a combination ion selective electrode II II II II II Specifications Specifications Specifications Specifications Specifications Type Solid State electrod...

Страница 3: ...as a sensor or ionic conductor The HI4012requiresaseparatereferenceelectrodetocomplete itselectrolyticcircuit TheHI4112incorporates areference electrode The mixed lead sulfide silver sulfide membrane...

Страница 4: ...4112 Design elements of the HI 4012 and HI 4112 Design elements of the HI 4012 and HI 4112 electrodes electrodes electrodes electrodes electrodes Upper Cap Spring O Ring Ceramic Junction on Inner Ste...

Страница 5: ...surements 0 1 M lead standard 500 mL HI4012 01 ISA 5 X 100 mL HI4012 00 For Sulfate Titrations used to standardize lead perchlorate solutions 0 1 M sulfate standard 500 mL HI4012 21 See Section XII Ti...

Страница 6: ...on standards and sample solutions should be at same temperature The magnetic stirrer may generate heat Thermally insulate beaker containing standard or sample from magnetic stirrer by placing cork or...

Страница 7: ...solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthisoccursverifycapistightenedand the interface between the in...

Страница 8: ...lm seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Страница 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Страница 10: ...than the previous noted DeterminethedifferencebetweenthetwomVvalues An acceptable value for this slope is 27 4 mV X X X X X Corrective action Corrective action Corrective action Corrective action Cor...

Страница 11: ...ndards The meter is calibrated with two or more freshly made standards that are in the linear mea surementrangeoftheunknowns TwomLofISA HI4001 12 is added to each 50 mL volume of standard or sample Fi...

Страница 12: ...er equilibra tions are required at lower concentrations 5 To prevent carry over and contamination of samples rinse sensors with deionized and dab dry with absorbant laboratory tissue between samples X...

Страница 13: ...is recorded after the sample is mixed 2 5mL Vstd of 10 1 M standard Cstd is added to the beaker and the mV value increases as does the con centration E is calculated as mV2 mV1 The un known lead conc...

Страница 14: ...ty Addition of ISA is suggested 2mL 50mL sample pH adjustment of the sample to 4 6 range is advised If required use perchloric acid or sodium hydroxide The lead titrant should be 10X more concentrated...

Страница 15: ...sensors HI 4112 sensors HI 4112 sensors HI 4112 sensors The HI 4012 and HI 4112 sensor can be stored in very dilute standards 10 3 M for short periods of time TheHI4012 shouldbestoreddrywithit sprote...

Страница 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Отзывы: