VII. General Guidelines

General Guidelines

General Guidelines

General Guidelines

General Guidelines

•    

Use of a hood and protective clothings is advised

        when handling lead solutions.

Concentrated samples (>0.10 M)  should be diluted
before measurement.  Multiply the final result by the
corresponding dilution factor.

Calibration standards and sample solutions should
have the same ionic strength. ISA and Methanol/
Formalin should be added to both samples and stan-
dards in the same ratio. 2 parts ISA to 50 parts stan-
dard (or sample)/50 parts Methanol/Formalin is the
normal dosing. Note: no correction for dilution factor is
required for this because standards and samples are
treated in the same way.

For high ionic strength samples,  use standard addi-
tion or titration methods.

Calibration standards and sample solutions should
be at same temperature.

The magnetic stirrer may generate heat. Thermally
insulate beaker containing standard or sample from
magnetic stirrer by placing cork or other insulative
sheet between beaker and stirrer plate.

Calibration standards and sample solutions should
be stirred at the same rate using identical sized TFE
coated stir bars.

Rinse electrode pair with distilled or deionized water
between samples and gently dab dry with soft dis-
posable absorbent toweling. Do not rub electrodes.

Presoaking lead sensor in a dilute standard will opti-
mize response. Use concentrations approximately
10

-3 

M or less with Methanol/Formalim and ISA added.

A scratched, pitted, or oxidized pellet surface can  cause
drift, a loss of low level response, or poor repeatability.
Optimum response can be restored by removing the
damaged surface with the microabrasive strip HI 4000-
70. Use gloves to protect skin.

Avoid large changes in temperature (thermal shock)
as it may damage the sensor.

Содержание HI 4112

Страница 1: ...I 4012 HI 4012 HI 4012 HI 4012 HI 4012 HI 4112 HI 4112 HI 4112 HI 4112 HI 4112 Lead Ion Selective Electrode Instruction Manual Half cell Combination LEAD SULFATE HI 4012 HI 4112 LEAD SULFATE COMBINATI...

Страница 2: ...t requires a separate reference The HI 4112 is a combination ion selective electrode II II II II II Specifications Specifications Specifications Specifications Specifications Type Solid State electrod...

Страница 3: ...as a sensor or ionic conductor The HI4012requiresaseparatereferenceelectrodetocomplete itselectrolyticcircuit TheHI4112incorporates areference electrode The mixed lead sulfide silver sulfide membrane...

Страница 4: ...4112 Design elements of the HI 4012 and HI 4112 Design elements of the HI 4012 and HI 4112 electrodes electrodes electrodes electrodes electrodes Upper Cap Spring O Ring Ceramic Junction on Inner Ste...

Страница 5: ...surements 0 1 M lead standard 500 mL HI4012 01 ISA 5 X 100 mL HI4012 00 For Sulfate Titrations used to standardize lead perchlorate solutions 0 1 M sulfate standard 500 mL HI4012 21 See Section XII Ti...

Страница 6: ...on standards and sample solutions should be at same temperature The magnetic stirrer may generate heat Thermally insulate beaker containing standard or sample from magnetic stirrer by placing cork or...

Страница 7: ...solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthisoccursverifycapistightenedand the interface between the in...

Страница 8: ...lm seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Страница 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Страница 10: ...than the previous noted DeterminethedifferencebetweenthetwomVvalues An acceptable value for this slope is 27 4 mV X X X X X Corrective action Corrective action Corrective action Corrective action Cor...

Страница 11: ...ndards The meter is calibrated with two or more freshly made standards that are in the linear mea surementrangeoftheunknowns TwomLofISA HI4001 12 is added to each 50 mL volume of standard or sample Fi...

Страница 12: ...er equilibra tions are required at lower concentrations 5 To prevent carry over and contamination of samples rinse sensors with deionized and dab dry with absorbant laboratory tissue between samples X...

Страница 13: ...is recorded after the sample is mixed 2 5mL Vstd of 10 1 M standard Cstd is added to the beaker and the mV value increases as does the con centration E is calculated as mV2 mV1 The un known lead conc...

Страница 14: ...ty Addition of ISA is suggested 2mL 50mL sample pH adjustment of the sample to 4 6 range is advised If required use perchloric acid or sodium hydroxide The lead titrant should be 10X more concentrated...

Страница 15: ...sensors HI 4112 sensors HI 4112 sensors HI 4112 sensors The HI 4012 and HI 4112 sensor can be stored in very dilute standards 10 3 M for short periods of time TheHI4012 shouldbestoreddrywithit sprote...

Страница 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Отзывы: