background image

11

       applying gentle pressure.  Continue polishing until

you are satisfied with the surface.   If dark deposits
appear on polishing strip move the paper slightly
and continue polishing.

  

If the membrane is damaged, the response becomes
extremely sluggish, or the slope of the electrode has
decreased significantly, and procedures above have
not helped, the sensor  should be replaced.

XI.   

XI.   

XI.   

XI.   

XI.   Direct Calibration and Measurement

Direct Calibration and Measurement

Direct Calibration and Measurement

Direct Calibration and Measurement

Direct Calibration and Measurement

This method is a simple procedure for measuring many
samples. Use of protective clothing is advised when han-
dling lead samples and standards. A direct reading ISE
meter (HI 4222 or equivalent) determines concentration of
the unknown by a direct reading after calibrating the meter
with the standards.  The meter is calibrated with two or
more freshly made standards that are in the linear mea-
surement range of the unknowns.  Two mL of ISA (HI 4001-
12) is added to each 50 mL volume of standard or sample.
Fifty milliliters of the methanol/formalin mixture is also
added to each sample or standard.   More calibration stan-
dards are required in non-linear regions.  Unknowns are
read directly.
Samples with concentration greater than 0.1 M should be
diluted to be within the working range of the electrodes.
The final result must be multiplied by the corresponding
dilution factor to determine the actual concentration.

A pH/mV meter in mV mode with semi-log graph paper
may also be used.  Two or more freshly prepared standards
that are in the measurement range of the unknowns are
measured in mV mode on the meter.
These values are plotted on the semi-log paper and the
points are connected to form a straight-line curve. When
samples are measured, their mV values are converted to
concentration by following the mV to the concentration axis
on the semi-log plot.

Содержание HI 4112

Страница 1: ...I 4012 HI 4012 HI 4012 HI 4012 HI 4012 HI 4112 HI 4112 HI 4112 HI 4112 HI 4112 Lead Ion Selective Electrode Instruction Manual Half cell Combination LEAD SULFATE HI 4012 HI 4112 LEAD SULFATE COMBINATI...

Страница 2: ...t requires a separate reference The HI 4112 is a combination ion selective electrode II II II II II Specifications Specifications Specifications Specifications Specifications Type Solid State electrod...

Страница 3: ...as a sensor or ionic conductor The HI4012requiresaseparatereferenceelectrodetocomplete itselectrolyticcircuit TheHI4112incorporates areference electrode The mixed lead sulfide silver sulfide membrane...

Страница 4: ...4112 Design elements of the HI 4012 and HI 4112 Design elements of the HI 4012 and HI 4112 electrodes electrodes electrodes electrodes electrodes Upper Cap Spring O Ring Ceramic Junction on Inner Ste...

Страница 5: ...surements 0 1 M lead standard 500 mL HI4012 01 ISA 5 X 100 mL HI4012 00 For Sulfate Titrations used to standardize lead perchlorate solutions 0 1 M sulfate standard 500 mL HI4012 21 See Section XII Ti...

Страница 6: ...on standards and sample solutions should be at same temperature The magnetic stirrer may generate heat Thermally insulate beaker containing standard or sample from magnetic stirrer by placing cork or...

Страница 7: ...solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthisoccursverifycapistightenedand the interface between the in...

Страница 8: ...lm seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Страница 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Страница 10: ...than the previous noted DeterminethedifferencebetweenthetwomVvalues An acceptable value for this slope is 27 4 mV X X X X X Corrective action Corrective action Corrective action Corrective action Cor...

Страница 11: ...ndards The meter is calibrated with two or more freshly made standards that are in the linear mea surementrangeoftheunknowns TwomLofISA HI4001 12 is added to each 50 mL volume of standard or sample Fi...

Страница 12: ...er equilibra tions are required at lower concentrations 5 To prevent carry over and contamination of samples rinse sensors with deionized and dab dry with absorbant laboratory tissue between samples X...

Страница 13: ...is recorded after the sample is mixed 2 5mL Vstd of 10 1 M standard Cstd is added to the beaker and the mV value increases as does the con centration E is calculated as mV2 mV1 The un known lead conc...

Страница 14: ...ty Addition of ISA is suggested 2mL 50mL sample pH adjustment of the sample to 4 6 range is advised If required use perchloric acid or sodium hydroxide The lead titrant should be 10X more concentrated...

Страница 15: ...sensors HI 4112 sensors HI 4112 sensors HI 4112 sensors The HI 4012 and HI 4112 sensor can be stored in very dilute standards 10 3 M for short periods of time TheHI4012 shouldbestoreddrywithit sprote...

Страница 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Отзывы: