V.  Equipment required:

Equipment required:

Equipment required:

Equipment required:

Equipment required:

Hanna HI 5315 Double Junction Reference Electrode
with HI 7072 Fill Solution for HI 4012.

Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is
useful if an ISE (ion) meter is not available).

Hanna HI 180 Magnetic Stirrer or equivalent with TFE
coated stirring bars (HI 731320).  Note: isolate bea-
kers from stirrer motor heat by placing insulating
material such as foam or cork between them.

Hanna HI 76404 Electrode Holder or equivalent.

Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.

VI.  

VI.  

VI.  

VI.  

VI.  Solutions Required

Solutions Required

Solutions Required

Solutions Required

Solutions Required

For Lead Measurements
0.1 M lead standard, (500 mL)

HI 4012-01

ISA, (5 X 100 mL)

HI 4012-00

For Sulfate Titrations
(used to standardize lead perchlorate solutions)
0.1 M sulfate standard, (500 mL)

HI 4012-21

See Section XII ; Titrations

User supplied solutions:
A user supplied Methanol/Formalin solution is  required to
decrease pellet solubility and oxidative effects. To prepare,
add 200 

µ

L 37 wt. % formalin in H

2

O to 500 mL of

reagent grade methanol. Stir and store in tightly closed
glass bottle. This solution is used at a dose of  50 parts of
the Methanol/Formalin solution to 50 parts standard or
sample.

For Direct lead measurements- Using volumetric pipettes
and glassware, make dilutions of HI 4012-01 to bracket
the concentration of the samples.  Note: 0.1M lead stan-
dard  is equivalent to 20,720 ppm. Standards with con-
centrations 

<

 10

-3 

M (207ppm) should be prepared daily.

Store in tightly closed plastic bottle. To 50 parts standard or
sample add 50 parts Methanol/Formalin solution plus 2
mL of ISA.

Содержание HI 4112

Страница 1: ...I 4012 HI 4012 HI 4012 HI 4012 HI 4012 HI 4112 HI 4112 HI 4112 HI 4112 HI 4112 Lead Ion Selective Electrode Instruction Manual Half cell Combination LEAD SULFATE HI 4012 HI 4112 LEAD SULFATE COMBINATI...

Страница 2: ...t requires a separate reference The HI 4112 is a combination ion selective electrode II II II II II Specifications Specifications Specifications Specifications Specifications Type Solid State electrod...

Страница 3: ...as a sensor or ionic conductor The HI4012requiresaseparatereferenceelectrodetocomplete itselectrolyticcircuit TheHI4112incorporates areference electrode The mixed lead sulfide silver sulfide membrane...

Страница 4: ...4112 Design elements of the HI 4012 and HI 4112 Design elements of the HI 4012 and HI 4112 electrodes electrodes electrodes electrodes electrodes Upper Cap Spring O Ring Ceramic Junction on Inner Ste...

Страница 5: ...surements 0 1 M lead standard 500 mL HI4012 01 ISA 5 X 100 mL HI4012 00 For Sulfate Titrations used to standardize lead perchlorate solutions 0 1 M sulfate standard 500 mL HI4012 21 See Section XII Ti...

Страница 6: ...on standards and sample solutions should be at same temperature The magnetic stirrer may generate heat Thermally insulate beaker containing standard or sample from magnetic stirrer by placing cork or...

Страница 7: ...solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthisoccursverifycapistightenedand the interface between the in...

Страница 8: ...lm seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Страница 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Страница 10: ...than the previous noted DeterminethedifferencebetweenthetwomVvalues An acceptable value for this slope is 27 4 mV X X X X X Corrective action Corrective action Corrective action Corrective action Cor...

Страница 11: ...ndards The meter is calibrated with two or more freshly made standards that are in the linear mea surementrangeoftheunknowns TwomLofISA HI4001 12 is added to each 50 mL volume of standard or sample Fi...

Страница 12: ...er equilibra tions are required at lower concentrations 5 To prevent carry over and contamination of samples rinse sensors with deionized and dab dry with absorbant laboratory tissue between samples X...

Страница 13: ...is recorded after the sample is mixed 2 5mL Vstd of 10 1 M standard Cstd is added to the beaker and the mV value increases as does the con centration E is calculated as mV2 mV1 The un known lead conc...

Страница 14: ...ty Addition of ISA is suggested 2mL 50mL sample pH adjustment of the sample to 4 6 range is advised If required use perchloric acid or sodium hydroxide The lead titrant should be 10X more concentrated...

Страница 15: ...sensors HI 4112 sensors HI 4112 sensors HI 4112 sensors The HI 4012 and HI 4112 sensor can be stored in very dilute standards 10 3 M for short periods of time TheHI4012 shouldbestoreddrywithit sprote...

Страница 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Отзывы: