background image

8

3 cm

(1¼")

If measurements are taken successively in different samples, it is recommended
to rinse the electrode thoroughly with deionized water or tap water and
then with some of the next sample to prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the pH
accurately, the temperature effect must be compensated for. To use the
Automatic Temperature Compensation feature, connect and submerse
the 

HI 7662

 temperature probe into the sample as close as possible to

the electrode and wait for a few seconds.
If the temperature of the sample is known, manual temperature
compensation can be used by disconnecting the temperature probe.
The display will then show the last temperature
reading with the “

ºC

” tag blinking.

The temperature can now be adjusted with the

ARROW

 keys (from –20.0 ºC to 120.0 ºC).

mV/ORP MEASUREMENTS

An optional ORP electrode must be used to perform ORP measurements
(see Accessories).
Oxidation-Reduction Potential (REDOX) measurements
provide the quantification of the oxidizing or reducing
power of the tested sample.
The surface of the ORP electrode must be clean and
smooth in order to obtain an accurate measurement.
• Press 

RANGE

 to enter mV range.

• Submerse the tip of the ORP electrode 3 cm (1¼”) into the sample

to be tested and allow a few seconds for the reading to stabilize.

• The instrument displays the mV reading on the primary LCD and the

temperature on the secondary LCD.

• If the reading is out of range, the closest full-scale value will be

displayed blinking on the primary LCD.

41

pH BUFFER SOLUTIONS

HI 70004P

pH 4.01 Buffer Sachets, 20 mL, 25 pcs

HI 70007P

pH 7.01 Buffer Sachets, 20 mL, 25 pcs

HI 70010P

pH 10.01 Buffer Sachets, 20 mL, 25 pcs

HI 7001L

pH 1.68 Buffer Solution, 500 mL

HI 7004L

pH 4.01 Buffer Solution, 500 mL

HI 7006L

pH 6.86 Buffer Solution, 500 mL

HI 7007L

pH 7.01 Buffer Solution, 500 mL

HI 7009L

pH 9.18 Buffer Solution, 500 mL

HI 7010L

pH 10.01 Buffer Solution, 500 mL

HI 8004L

pH 4.01 Buffer Solution in FDA approved bottle, 500 mL

HI 8006L

pH 6.86 Buffer Solution in FDA approved bottle, 500 mL

HI 8007L

pH 7.01 Buffer Solution in FDA approved bottle, 500 mL

HI 8009L

pH 9.18 Buffer Solution in FDA approved bottle, 500 mL

HI 8010L

pH 10.01 Buffer Solution in FDA approved bottle, 500 mL

ELECTRODE STORAGE SOLUTIONS

HI 70300L

Storage Solution, 500 mL

HI 80300L

Storage Solution in FDA approved bottle, 500 mL

ELECTRODE CLEANING SOLUTIONS

HI 70000P

Electrode Rinse Sachets, 20 mL, 25 pcs

HI 7061L

General Cleaning Solution, 500 mL

HI 7073L

Protein Cleaning Solution, 500 mL

HI 7074L

Inorganic Cleaning Solution, 500 mL

HI 7077L

Oil & Fat Cleaning Solution, 500 mL

HI 8061L

General Cleaning Solution in FDA approved bottle, 500 mL

HI 8073L

Protein Cleaning Solution in FDA approved bottle, 500 mL

HI 8077L

Oil & Fat Cleaning Solution in FDA approved bottle, 500 mL

ELECTRODE REFILL ELECTROLYTE SOLUTIONS

HI 7071

3.5M KCl + AgCl Electrolyte, 4x30 mL, for single junction
electrodes

HI 7072

1M KNO

3

 Electrolyte, 4x30 mL

HI 7082

3.5M KCl Electrolyte, 4x30 mL, for double junction electrodes

HI 8071

3.5M KCl + AgCl Electrolyte in FDA approved bottle, 4x30
mL, for single junction electrodes

HI 8072

1M KNO

3

 Electrolyte in FDA approved bottle, 4x30 mL

HI 8082

3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for
double junction electrodes

ACCESSORIES

Содержание HI 2214

Страница 1: ...al www hannainst com PrintedinROMANIA MAN2215 07 14 Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401...

Страница 2: ...NATION 3 GENERALDESCRIPTION 3 FUNCTIONALDESCRIPTION 4 HI2214SPECIFICATIONS 5 HI2215SPECIFICATIONS 6 OPERATIONALGUIDE 7 pHCALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 15 RELATIVEmVCALIBRATION 16 GOODLAB...

Страница 3: ...n standard buffers In addition two custom buffers can be used during calibration The GLP feature provides data consistency A calibration due alarm can be set to alert the user that too much time elaps...

Страница 4: ...lear pH calibration or to delete log records 9 RCLkey memory recall 10 ON OFFswitch 11 Powersupplysocket 12 USBconnector 13 BNCelectrodeconnector 14 Temperature probe socket 15 Electrodereferencesocke...

Страница 5: ...e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l n o i t a c i n u m m o c C P B S U d e t a l o s i o t p O y l p p u s r e w o P d e d u l c...

Страница 6: ...r p e r u t a r e p m e T 2 6 6 7 I H d e d u l c n i e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l e r u t a e f l a v r e t n I g o L b a...

Страница 7: ...until initialization is complete pH MEASUREMENTS Makesuretheinstrumenthasbeencalibratedbefore taking pH measurements Submerse the electrode tip and the temperature probe approximately 3 cm 1 into the...

Страница 8: ...SOLUTIONS HI 70004P pH 4 01 Buffer Sachets 20 mL 25 pcs HI 70007P pH 7 01 Buffer Sachets 20 mL 25 pcs HI 70010P pH 10 01 Buffer Sachets 20 mL 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL HI 7004L pH...

Страница 9: ...01assecondbuffer Ifyouaremeasuringinthealkaline range use pH 7 01 or 6 86 as first buffer and pH 10 01 or 9 18 as secondbuffer 40 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4...

Страница 10: ...l o r t c e l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C n o i t u l o s h s e r f h t i w l l i f e R s e d o r t c e l e e l b a l l i f e r r o f d n...

Страница 11: ...the LCD until the reading is stable When the reading is stable and close to the selected buffer the READY tag will be displayed and the CFM tag will blink Press CFM to confirm calibration The calibrat...

Страница 12: ...dreturnstonormalmeasurementmode 37 For refillable electrodes If the filling solution electrolyte is more than 2 cm 1 below the fill hole add HI 7082 or HI 8082 3 5M KCl Electrolyte Solution for double...

Страница 13: ...with currentbuffer pressCLR Thecalibrationwillcontinue from the current point If this procedure is performed while calibrating in the first point the instrument returns to measurement mode Press RANGE...

Страница 14: ...chars NSLx Requests the number of logged samples 4 chars Command Parameter 1 char x range see Note LODPxxx Requests the xxxth pH record logged data LODMxxx Requests the xxxth mV Rel mV record logged d...

Страница 15: ...0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 34 Reading status 2 chars R in range O over range U under range First character corresponds...

Страница 16: ...01 pH range 0 01 resolution xx 02 pH range 0 1 resolution xx 03 mV range xx 04 Relative mV range The instrument will answer for these commands with STX answer ETX where STX is 02 ASCII code character...

Страница 17: ...be done with the HI 92000 Windows compatible software optional HI 92000 also offers graphing and an on line help feature Data can be exported to the most popular spreadsheet programs for further analy...

Страница 18: ...only All the instruments are factory calibrated for mV Hanna s ORP electrodes are interchangeable and no mV calibration is needed when they are replaced If the mV measurements are inaccurate mV recal...

Страница 19: ...should be performed For an accurate recalibration contact your dealer or the nearest Hanna Customer Service Center or follow the instructions below Prepare a vessel containing ice and water and anothe...

Страница 20: ...ID 0000 to 9999 and CFM will start blinking Press the ARROW keys to change the instrument ID value Press CFM to save the modified instrument ID value Press CAL to escape without saving Note The instru...

Страница 21: ...yed to alert the user If the log space is full 100 records FULL LOG message will be displayed and no more data will be saved WhenLOG ispressed acompletesetofinformationisstored date time pH mV tempera...

Страница 22: ...CAL when cb2 is displayed The custom buffer disabled no or 0 to 16 pH and CFM will start blinking Press the ARROW keys to change the custom buffer value Press CFM to save the modified custom buffer v...

Страница 23: ...pired Calibration Alarm Log Interval HI 2215 only FirstCustomBuffer SecondCustomBuffer One point Calibration Behavior pHResolution Current Time hour minute Current Date year month day Beep Status Inst...

Страница 24: ...hange the record number To delete a lot use the ARROW keys to select the desired lot Press CLR key dEL Lot will appear on the display Press CFM key to delete 25 The dEL message is displayed on the pri...

Отзывы: