background image

22

LOT LOGGING 

(HI 2215 only)

Select “

StAb

” (for stability logging) or the desired time interval.

To start interval logging, press 

LOG 

key while the instrument is in

measurement mode.
When the selected interval is reached, the instrument will display the
current lot number on the primary LCD line, the record number on the
secondary LCD line and the “

LOG

“ tag (see example below: Lot 5 record 7)

followed by the number of free records on the corresponding memory space.

If stability logging is selected, a complete set of data is memorized every
time the reading becomes stable after an unstable condition.
To stop interval logging, press 

LOG 

key again.

Note:

 When pressing any key that is not active, while Log interval is

running, the following message is displayed for a few seconds:

VIEW LOGGED DATA

Press 

RCL

     while in measurement mode to retrive the stored information.

If no data was logged for the selected measurement range and no lots
are memorized (

HI 2215

 only), one of the next messages will be

displayed:

No records:

No pH measurement records:

No Relative mV and mV records:

27

SECOND CUSTOM BUFFER

Press 

CAL

 when 

“cb2”

 is displayed. The custom buffer (disabled – 

“no”

or “

0

” to “

16

” pH) and “

CFM

” will start blinking.

Press the 

ARROW

 keys to change the custom buffer value.

Press 

CFM

 to save the modified custom buffer value.

Press 

CAL

 to escape without saving.

ONE-POINT CALIBRATION BEHAVIOR

Press 

CAL

 when “

1 Pnt

” message is displayed on the secondary LCD.

One of the two options (“

Pnt

” or “

OFFS

”) and “

CFM

” will start

blinking (see pH CALIBRATION PROCEDURE for details, page 9).

Press the 

ARROW

 keys to toggle between “

Pnt

” and “

OFFS

” options.

Press 

CFM

 to save the behavior for one-point calibration.

Press 

CAL

 to escape without saving.

pH RESOLUTION

Press 

CAL

 when “

rES

” message is displayed on the secondary LCD. The

set resolution (“

0.1

“, “

0.01

“ or “

0.001

“) and “

CFM

” will start blinking.

Press the 

ARROW

 keys to toggle between 0.1, 0.01 and 0.001 options

for 

HI 2215

; 0.1 and 0.01 options for 

HI 2214

.

Press 

CFM

 to save the modified value.

Press 

CAL

 to escape without saving.

FIRST CUSTOM BUFFER

Press 

CAL

 when 

“cb1”

 is displayed. The custom buffer (disabled – 

“no”

or “

0

” to ”

16

” pH) and “

CFM

” will start blinking.

Press the 

ARROW

 keys to change the custom buffer value.

Press 

CFM 

to save the modified custom buffer value.

Press 

CAL

 to escape without saving.

Note:

To delete custom buffer, enter change mode by pressing 

CAL

 then

press 

CLR

 key.

Содержание HI 2214

Страница 1: ...al www hannainst com PrintedinROMANIA MAN2215 07 14 Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401...

Страница 2: ...NATION 3 GENERALDESCRIPTION 3 FUNCTIONALDESCRIPTION 4 HI2214SPECIFICATIONS 5 HI2215SPECIFICATIONS 6 OPERATIONALGUIDE 7 pHCALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 15 RELATIVEmVCALIBRATION 16 GOODLAB...

Страница 3: ...n standard buffers In addition two custom buffers can be used during calibration The GLP feature provides data consistency A calibration due alarm can be set to alert the user that too much time elaps...

Страница 4: ...lear pH calibration or to delete log records 9 RCLkey memory recall 10 ON OFFswitch 11 Powersupplysocket 12 USBconnector 13 BNCelectrodeconnector 14 Temperature probe socket 15 Electrodereferencesocke...

Страница 5: ...e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l n o i t a c i n u m m o c C P B S U d e t a l o s i o t p O y l p p u s r e w o P d e d u l c...

Страница 6: ...r p e r u t a r e p m e T 2 6 6 7 I H d e d u l c n i e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l e r u t a e f l a v r e t n I g o L b a...

Страница 7: ...until initialization is complete pH MEASUREMENTS Makesuretheinstrumenthasbeencalibratedbefore taking pH measurements Submerse the electrode tip and the temperature probe approximately 3 cm 1 into the...

Страница 8: ...SOLUTIONS HI 70004P pH 4 01 Buffer Sachets 20 mL 25 pcs HI 70007P pH 7 01 Buffer Sachets 20 mL 25 pcs HI 70010P pH 10 01 Buffer Sachets 20 mL 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL HI 7004L pH...

Страница 9: ...01assecondbuffer Ifyouaremeasuringinthealkaline range use pH 7 01 or 6 86 as first buffer and pH 10 01 or 9 18 as secondbuffer 40 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4...

Страница 10: ...l o r t c e l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C n o i t u l o s h s e r f h t i w l l i f e R s e d o r t c e l e e l b a l l i f e r r o f d n...

Страница 11: ...the LCD until the reading is stable When the reading is stable and close to the selected buffer the READY tag will be displayed and the CFM tag will blink Press CFM to confirm calibration The calibrat...

Страница 12: ...dreturnstonormalmeasurementmode 37 For refillable electrodes If the filling solution electrolyte is more than 2 cm 1 below the fill hole add HI 7082 or HI 8082 3 5M KCl Electrolyte Solution for double...

Страница 13: ...with currentbuffer pressCLR Thecalibrationwillcontinue from the current point If this procedure is performed while calibrating in the first point the instrument returns to measurement mode Press RANGE...

Страница 14: ...chars NSLx Requests the number of logged samples 4 chars Command Parameter 1 char x range see Note LODPxxx Requests the xxxth pH record logged data LODMxxx Requests the xxxth mV Rel mV record logged d...

Страница 15: ...0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 34 Reading status 2 chars R in range O over range U under range First character corresponds...

Страница 16: ...01 pH range 0 01 resolution xx 02 pH range 0 1 resolution xx 03 mV range xx 04 Relative mV range The instrument will answer for these commands with STX answer ETX where STX is 02 ASCII code character...

Страница 17: ...be done with the HI 92000 Windows compatible software optional HI 92000 also offers graphing and an on line help feature Data can be exported to the most popular spreadsheet programs for further analy...

Страница 18: ...only All the instruments are factory calibrated for mV Hanna s ORP electrodes are interchangeable and no mV calibration is needed when they are replaced If the mV measurements are inaccurate mV recal...

Страница 19: ...should be performed For an accurate recalibration contact your dealer or the nearest Hanna Customer Service Center or follow the instructions below Prepare a vessel containing ice and water and anothe...

Страница 20: ...ID 0000 to 9999 and CFM will start blinking Press the ARROW keys to change the instrument ID value Press CFM to save the modified instrument ID value Press CAL to escape without saving Note The instru...

Страница 21: ...yed to alert the user If the log space is full 100 records FULL LOG message will be displayed and no more data will be saved WhenLOG ispressed acompletesetofinformationisstored date time pH mV tempera...

Страница 22: ...CAL when cb2 is displayed The custom buffer disabled no or 0 to 16 pH and CFM will start blinking Press the ARROW keys to change the custom buffer value Press CFM to save the modified custom buffer v...

Страница 23: ...pired Calibration Alarm Log Interval HI 2215 only FirstCustomBuffer SecondCustomBuffer One point Calibration Behavior pHResolution Current Time hour minute Current Date year month day Beep Status Inst...

Страница 24: ...hange the record number To delete a lot use the ARROW keys to select the desired lot Press CLR key dEL Lot will appear on the display Press CFM key to delete 25 The dEL message is displayed on the pri...