20
RELATIVE mV CALIBRATION DATA
Last Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press
GLP
while in Relative mV
measurement mode.
The instrument will display the Relative mV GLP information and the
time (hh:mm:ss) of the last Rel mV calibration.
Use the
ARROW
keys to scroll through the calibration data:
• The Relative mV calibration date.
• The Relative mV calibration offset.
• The instrument ID.
Notes:
• Press
GLP
to return to measurement mode.
• If calibration has not been performed, the instrument
displays “
no CAL
” message blinking.
29
INSTRUMENT ID
Press
CAL
when the “
InId
” is displayed. The instrument ID (“
0000
“ to
“
9999
“) and “
CFM
” will start blinking.
Press the
ARROW
keys to change the instrument ID value.
Press
CFM
to save the modified instrument ID value.
Press
CAL
to escape without saving.
Note:
The instrument ID is downloaded to a PC as part of a logged
data, set to identify its origin.
TEMPERATURE UNIT
Press
CAL
when “
tnP
“ is displayed. The temperature unit and “
CFM
”
will start blinking.
Press the
ARROW
keys to change the option.
Press
CFM
to save the modified temperature unit.
Press
CAL
to escape without saving.
Press the
ARROW
keys to change the beep status (On or OFF).
Press
CFM
to save the modified beep status.
Press
CAL
to escape without saving.
When enable, beep sounds as a short beep every time a key is pressed or
when the calibration can be confirmed.
A long beep alerts that the pressed key is not active or a wrong condition
is detected while in calibration.