background image

11

• The calibrated value will be displayed on the primary LCD and the

second expected buffer value on the secondary LCD.

• After the first calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the second buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the 

ARROW

 keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected

buffer, the “

READY

” tag will be displayed and the

CFM

” tag will blink.

• Press 

CFM

 to confirm calibration.

• The calibrated value is then displayed on the primary

LCD and the third expected buffer value on the
secondary LCD.

• After the second calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”) into
the next buffer solution and stir gently. The temperature probe
should be close to the pH electrode.

• If necessary, press the 

ARROW

 keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is

stable.

• When the reading is stable and close to the selected

buffer, the “

READY

” tag will be displayed and the

CFM

” tag will blink.

• Press 

CFM

 to confirm calibration.

• The calibrated value is then displayed on the primary LCD and the

fourth expected value on the secondary LCD.

• After the third calibration point is confirmed, submerse the pH electrode

and the temperature probe approximately 3 cm (1¼”) into the
next buffer solution and stir gently. The temperature probe should
be close to the pH electrode.

• If necessary, press the 

ARROW

 keys to select a different buffer value.

38

CLEANING PROCEDURE

• General

Soak in Hanna 

HI 7061

 or 

HI 8061

 General

Cleaning Solution for approximately ½ hour.

• Protein

Soak in Hanna 

HI 7073

 or 

HI 8073

 Protein

Cleaning Solution for 15 minutes.

• Inorganic

Soak in Hanna 

HI 7074

 Inorganic Cleaning

Solution for 15 minutes.

• Oil/grease

Rinse with Hanna 

HI 7077

 or 

HI 8077

 Oil and Fat

Cleaning Solution.

IMPORTANT:

 After performing any of the cleaning procedures, rinse the

electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel-filled electrodes) and soak the
electrode in 

HI 70300

 or 

HI 80300

 Storage Solution for at least 1 hour

before taking measurements.

Содержание HI 2214

Страница 1: ...al www hannainst com PrintedinROMANIA MAN2215 07 14 Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401...

Страница 2: ...NATION 3 GENERALDESCRIPTION 3 FUNCTIONALDESCRIPTION 4 HI2214SPECIFICATIONS 5 HI2215SPECIFICATIONS 6 OPERATIONALGUIDE 7 pHCALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 15 RELATIVEmVCALIBRATION 16 GOODLAB...

Страница 3: ...n standard buffers In addition two custom buffers can be used during calibration The GLP feature provides data consistency A calibration due alarm can be set to alert the user that too much time elaps...

Страница 4: ...lear pH calibration or to delete log records 9 RCLkey memory recall 10 ON OFFswitch 11 Powersupplysocket 12 USBconnector 13 BNCelectrodeconnector 14 Temperature probe socket 15 Electrodereferencesocke...

Страница 5: ...e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l n o i t a c i n u m m o c C P B S U d e t a l o s i o t p O y l p p u s r e w o P d e d u l c...

Страница 6: ...r p e r u t a r e p m e T 2 6 6 7 I H d e d u l c n i e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l e r u t a e f l a v r e t n I g o L b a...

Страница 7: ...until initialization is complete pH MEASUREMENTS Makesuretheinstrumenthasbeencalibratedbefore taking pH measurements Submerse the electrode tip and the temperature probe approximately 3 cm 1 into the...

Страница 8: ...SOLUTIONS HI 70004P pH 4 01 Buffer Sachets 20 mL 25 pcs HI 70007P pH 7 01 Buffer Sachets 20 mL 25 pcs HI 70010P pH 10 01 Buffer Sachets 20 mL 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL HI 7004L pH...

Страница 9: ...01assecondbuffer Ifyouaremeasuringinthealkaline range use pH 7 01 or 6 86 as first buffer and pH 10 01 or 9 18 as secondbuffer 40 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4...

Страница 10: ...l o r t c e l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C n o i t u l o s h s e r f h t i w l l i f e R s e d o r t c e l e e l b a l l i f e r r o f d n...

Страница 11: ...the LCD until the reading is stable When the reading is stable and close to the selected buffer the READY tag will be displayed and the CFM tag will blink Press CFM to confirm calibration The calibrat...

Страница 12: ...dreturnstonormalmeasurementmode 37 For refillable electrodes If the filling solution electrolyte is more than 2 cm 1 below the fill hole add HI 7082 or HI 8082 3 5M KCl Electrolyte Solution for double...

Страница 13: ...with currentbuffer pressCLR Thecalibrationwillcontinue from the current point If this procedure is performed while calibrating in the first point the instrument returns to measurement mode Press RANGE...

Страница 14: ...chars NSLx Requests the number of logged samples 4 chars Command Parameter 1 char x range see Note LODPxxx Requests the xxxth pH record logged data LODMxxx Requests the xxxth mV Rel mV record logged d...

Страница 15: ...0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 34 Reading status 2 chars R in range O over range U under range First character corresponds...

Страница 16: ...01 pH range 0 01 resolution xx 02 pH range 0 1 resolution xx 03 mV range xx 04 Relative mV range The instrument will answer for these commands with STX answer ETX where STX is 02 ASCII code character...

Страница 17: ...be done with the HI 92000 Windows compatible software optional HI 92000 also offers graphing and an on line help feature Data can be exported to the most popular spreadsheet programs for further analy...

Страница 18: ...only All the instruments are factory calibrated for mV Hanna s ORP electrodes are interchangeable and no mV calibration is needed when they are replaced If the mV measurements are inaccurate mV recal...

Страница 19: ...should be performed For an accurate recalibration contact your dealer or the nearest Hanna Customer Service Center or follow the instructions below Prepare a vessel containing ice and water and anothe...

Страница 20: ...ID 0000 to 9999 and CFM will start blinking Press the ARROW keys to change the instrument ID value Press CFM to save the modified instrument ID value Press CAL to escape without saving Note The instru...

Страница 21: ...yed to alert the user If the log space is full 100 records FULL LOG message will be displayed and no more data will be saved WhenLOG ispressed acompletesetofinformationisstored date time pH mV tempera...

Страница 22: ...CAL when cb2 is displayed The custom buffer disabled no or 0 to 16 pH and CFM will start blinking Press the ARROW keys to change the custom buffer value Press CFM to save the modified custom buffer v...

Страница 23: ...pired Calibration Alarm Log Interval HI 2215 only FirstCustomBuffer SecondCustomBuffer One point Calibration Behavior pHResolution Current Time hour minute Current Date year month day Beep Status Inst...

Страница 24: ...hange the record number To delete a lot use the ARROW keys to select the desired lot Press CLR key dEL Lot will appear on the display Press CFM key to delete 25 The dEL message is displayed on the pri...

Отзывы: