11
• The calibrated value will be displayed on the primary LCD and the
second expected buffer value on the secondary LCD.
• After the first calibration point is confirmed, submerse the pH
electrode and the temperature probe approximately 3 cm (1¼”) into
the second buffer solution and stir gently. The temperature probe
should be close to the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected
buffer, the “
READY
” tag will be displayed and the
“
CFM
” tag will blink.
• Press
CFM
to confirm calibration.
• The calibrated value is then displayed on the primary
LCD and the third expected buffer value on the
secondary LCD.
• After the second calibration point is confirmed, submerse the pH
electrode and the temperature probe approximately 3 cm (1¼”) into
the next buffer solution and stir gently. The temperature probe
should be close to the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
• The “ ” tag will blink on the LCD until the reading is
stable.
• When the reading is stable and close to the selected
buffer, the “
READY
” tag will be displayed and the
“
CFM
” tag will blink.
• Press
CFM
to confirm calibration.
• The calibrated value is then displayed on the primary LCD and the
fourth expected value on the secondary LCD.
• After the third calibration point is confirmed, submerse the pH electrode
and the temperature probe approximately 3 cm (1¼”) into the
next buffer solution and stir gently. The temperature probe should
be close to the pH electrode.
• If necessary, press the
ARROW
keys to select a different buffer value.
38
CLEANING PROCEDURE
• General
Soak in Hanna
HI 7061
or
HI 8061
General
Cleaning Solution for approximately ½ hour.
• Protein
Soak in Hanna
HI 7073
or
HI 8073
Protein
Cleaning Solution for 15 minutes.
• Inorganic
Soak in Hanna
HI 7074
Inorganic Cleaning
Solution for 15 minutes.
• Oil/grease
Rinse with Hanna
HI 7077
or
HI 8077
Oil and Fat
Cleaning Solution.
IMPORTANT:
After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel-filled electrodes) and soak the
electrode in
HI 70300
or
HI 80300
Storage Solution for at least 1 hour
before taking measurements.