background image

12

FOUR, THREE OR TWO-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.

• Press 

CAL

 after the fourth, third or second calibration point was

confirmed. The instrument will memorize the calibration data and
return to measurement mode.

ONE-POINT CALIBRATION

Two SETUP selectable options are available: “

Pnt

” and “

OFFS

”.

If the “

Pnt

” option is selected, the new calibration point overrides an

existing one. The adjacent slopes will be reevaluated.

If the “

OFFS

” option is selected, an electrode offset correction is

performed. The adjacent slopes will remain unchanged.

• Proceed as described in “FOUR, THREE or TWO-POINT CALIBRATION”

section.

• Press 

CAL

 after the first calibration point was confirmed. The

instrument will memorize the one-point calibration data and return
to measurement mode.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected

buffer, the “

READY

” tag will be displayed and the

CFM

” tag will blink.

• Press 

CFM

 to confirm calibration.

• The calibrated value is then displayed on the

primary LCD and the fifth expected value on the
secondary LCD.

• After the fourth calibration point is confirmed, submerse the pH

electrode and the temperature probe approximately 3 cm (1¼”)
into the next buffer solution and stir gently. The temperature
probe should be close to the pH electrode.

• If necessary, press the 

ARROW

 keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected

buffer, the “

READY

” tag will be displayed and the

CFM

” tag will blink.

• Press 

CFM

 to confirm the fifth calibration point.

• The instrument stores the calibration value of the

fifth point and returns to normal measurement mode.

37

For refillable electrodes:

If the filling solution (electrolyte) is more than 2½ cm (1”) below the fill
hole, add 

HI 7082

 or 

HI 8082

 3.5M KCl Electrolyte Solution for double

junction or 

HI 7071

 or 

HI 8071

 3.5M KCl+AgCl Electrolyte Solution for

single junction electrodes.
For faster response, unscrew the fill hole screw during measurements.

For AMPHEL

®

 electrodes:

If the electrode does not respond to pH changes, the battery is run down
and the electrode should be replaced.

MEASUREMENT

Rinse the electrode tip with distilled water. Submerse the tip (3 cm /1¼”)
in the sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested, before
taking measurements.

STORAGE PROCEDURE

To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of 

HI 70300

or 

HI 80300

 Storage Solution or, in its absence, Filling Solution

(

HI 7071

 or 

HI 8071

 for single junction and 

HI 7082

 or 

HI 8082

 for

double junction electrodes). Follow the Preparation Procedure  before
taking measurements.

Note:

NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cable used for connection to the
instrument must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb. Connectors
must be perfectly clean and dry. If any scratches or cracks are present,
replace the electrode. Rinse off any salt deposits with water.

For refillable electrodes:

Refill the reference chamber with fresh electrolyte (

HI 7071

 or 

HI 8071

 for

single junction and 

HI 7082

 or 

HI 8082

 for double junction electrodes).

Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.

Содержание HI 2214

Страница 1: ...al www hannainst com PrintedinROMANIA MAN2215 07 14 Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401...

Страница 2: ...NATION 3 GENERALDESCRIPTION 3 FUNCTIONALDESCRIPTION 4 HI2214SPECIFICATIONS 5 HI2215SPECIFICATIONS 6 OPERATIONALGUIDE 7 pHCALIBRATION 9 pHBUFFERTEMPERATUREDEPENDENCE 15 RELATIVEmVCALIBRATION 16 GOODLAB...

Страница 3: ...n standard buffers In addition two custom buffers can be used during calibration The GLP feature provides data consistency A calibration due alarm can be set to alert the user that too much time elaps...

Страница 4: ...lear pH calibration or to delete log records 9 RCLkey memory recall 10 ON OFFswitch 11 Powersupplysocket 12 USBconnector 13 BNCelectrodeconnector 14 Temperature probe socket 15 Electrodereferencesocke...

Страница 5: ...e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l n o i t a c i n u m m o c C P B S U d e t a l o s i o t p O y l p p u s r e w o P d e d u l c...

Страница 6: ...r p e r u t a r e p m e T 2 6 6 7 I H d e d u l c n i e c n a d e p m i t u p n I 0 1 2 1 s m h o e r u t a e f g o L s d r o c e r 0 0 1 d n a m e d n o g o l e r u t a e f l a v r e t n I g o L b a...

Страница 7: ...until initialization is complete pH MEASUREMENTS Makesuretheinstrumenthasbeencalibratedbefore taking pH measurements Submerse the electrode tip and the temperature probe approximately 3 cm 1 into the...

Страница 8: ...SOLUTIONS HI 70004P pH 4 01 Buffer Sachets 20 mL 25 pcs HI 70007P pH 7 01 Buffer Sachets 20 mL 25 pcs HI 70010P pH 10 01 Buffer Sachets 20 mL 25 pcs HI 7001L pH 1 68 Buffer Solution 500 mL HI 7004L pH...

Страница 9: ...01assecondbuffer Ifyouaremeasuringinthealkaline range use pH 7 01 or 6 86 as first buffer and pH 10 01 or 9 18 as secondbuffer 40 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4...

Страница 10: ...l o r t c e l e w o L s e d o r t c e l e e l b a l l i f e r y l n o e d o r t c e l e e h t n a e l C n o i t u l o s h s e r f h t i w l l i f e R s e d o r t c e l e e l b a l l i f e r r o f d n...

Страница 11: ...the LCD until the reading is stable When the reading is stable and close to the selected buffer the READY tag will be displayed and the CFM tag will blink Press CFM to confirm calibration The calibrat...

Страница 12: ...dreturnstonormalmeasurementmode 37 For refillable electrodes If the filling solution electrolyte is more than 2 cm 1 below the fill hole add HI 7082 or HI 8082 3 5M KCl Electrolyte Solution for double...

Страница 13: ...with currentbuffer pressCLR Thecalibrationwillcontinue from the current point If this procedure is performed while calibrating in the first point the instrument returns to measurement mode Press RANGE...

Страница 14: ...chars NSLx Requests the number of logged samples 4 chars Command Parameter 1 char x range see Note LODPxxx Requests the xxxth pH record logged data LODMxxx Requests the xxxth mV Rel mV record logged d...

Страница 15: ...0 9 4 9 1 9 7 1 9 1 4 8 8 6 3 0 7 5 8 8 5 7 9 2 8 0 1 5 9 3 0 2 1 8 1 0 2 4 9 8 6 4 0 7 3 8 8 6 7 9 3 7 0 1 34 Reading status 2 chars R in range O over range U under range First character corresponds...

Страница 16: ...01 pH range 0 01 resolution xx 02 pH range 0 1 resolution xx 03 mV range xx 04 Relative mV range The instrument will answer for these commands with STX answer ETX where STX is 02 ASCII code character...

Страница 17: ...be done with the HI 92000 Windows compatible software optional HI 92000 also offers graphing and an on line help feature Data can be exported to the most popular spreadsheet programs for further analy...

Страница 18: ...only All the instruments are factory calibrated for mV Hanna s ORP electrodes are interchangeable and no mV calibration is needed when they are replaced If the mV measurements are inaccurate mV recal...

Страница 19: ...should be performed For an accurate recalibration contact your dealer or the nearest Hanna Customer Service Center or follow the instructions below Prepare a vessel containing ice and water and anothe...

Страница 20: ...ID 0000 to 9999 and CFM will start blinking Press the ARROW keys to change the instrument ID value Press CFM to save the modified instrument ID value Press CAL to escape without saving Note The instru...

Страница 21: ...yed to alert the user If the log space is full 100 records FULL LOG message will be displayed and no more data will be saved WhenLOG ispressed acompletesetofinformationisstored date time pH mV tempera...

Страница 22: ...CAL when cb2 is displayed The custom buffer disabled no or 0 to 16 pH and CFM will start blinking Press the ARROW keys to change the custom buffer value Press CFM to save the modified custom buffer v...

Страница 23: ...pired Calibration Alarm Log Interval HI 2215 only FirstCustomBuffer SecondCustomBuffer One point Calibration Behavior pHResolution Current Time hour minute Current Date year month day Beep Status Inst...

Страница 24: ...hange the record number To delete a lot use the ARROW keys to select the desired lot Press CLR key dEL Lot will appear on the display Press CFM key to delete 25 The dEL message is displayed on the pri...

Отзывы: