9.5 Strap change RiseAtlas. Spare part no: 81000003
1.
Remove covers (see chapter 3)
2.
Remove h-adapter
3.
Extract strap with hand controll or emergency lowering
4.
Deactivate micro switch in order to extract all strap until opening in shaft is revealed
5.
Remove batteries and main board (see chapter 9.2, 9.3)
6.
Remove emergency lowering strap
7.
Remove interface plate
8.
Unscrew all four screws (pilar!)