1a)
Change of circuit board RiseAtlas,
The picture shows an example of RiseAtlas circuit
board RiseAtlas450MQ (manual motor). Important is
to disconnect and connect all wirings properly – use
marking if necessary.
After removing the RiseAtlas cover, please check and perform the following operations:
•
Check if all cables and connectors are correct attached.
-
If loose connector is found, attach correctly and reassembly the RiseAtlas covers. Check functionality in
order to establish if it is necessary to change PCB.
•
Disconnect the battery or take away the fuse. Important that the power supply is disconnected.
•
Remove the old PCB-card from the RiseAtlas by:
-
disconnecting all wirings from PCB-card.
-
Take away the old PCB-card by pressing the plastic clips and leave the plastic clips.
21
S E R V I C E M A N U A L