9.4 Replace bottom plate incl. display, RiseAtlas. Spare part no: 81000017
In order to replace bottom plate there is a need to open the RiseAtlas cover, see chapter 3 Access inside RiseAtlas.
Handling of electronics shall be performed with caution and the use of ESD protection.
The bottom plate incl. display, spare part no: 81000017, is complete in its delivery.
Remove the bottom plate by:
1. Remove batteries.
2. Disconnect PCB cables from bottom plate.
3. Remove H-Adapter
4 A. Disassemble the chassis from the bottom plate by removing the for screws from inside the chassie, see picture, and
check the emergency stop, to be reused, and also the end-of-strap limiting switch, to be reused.
4 B. Remove interface plate, unscrew the three screws underneath the bottom plate and both screws that holds the
lower limit switch.
5. Change the bottom plate and start by fitting the emergency stop and limit switch, see item 4.
6. Tighten the chassis screws.
7. Connect the cables.
8. Assemble the batteries.
9. Attach new inter face plate and remount h-adapter. Don´t forget the safety pin!
Make an initial function test before closing the cover.
Before reassembly the RiseAtlas covers, ensure that all connectors are correctly attached and that there are no
obstacles.
A functional, safety and load test, according to 2.3.2, 2.3.3 and 2.3.4, shall be performed of RiseAtlas to
ensure proper function and notes of change should be presented in the lift logbook.
The old parts can be recycled at the nearest recycling station
A: Older version
B: New version
23
S E R V I C E M A N U A L
System
RoMedic
TM
Содержание SystemRoMedic RiseAtlas625
Страница 2: ......
Страница 11: ...11 S E R V I C E M A N U A L SystemRoMedic TM 3c 2j 2k 2i 3b 2c ...
Страница 12: ...12 S E R V I C E M A N U A L SystemRoMedic TM 2b 2a 2h 3a 2g 2l 2d ...
Страница 47: ...47 S E R V I C E M A N U A L SystemRoMedic TM ...