Handicare Freecurve
68
The LED flashes upon joystick operation in both directions
Connect the short-circuit connectors K6 and K9
A. The lift moves after operation of the direction switch (DS)
The fault is located in the chair
•
The seat position switch is switched off/faulty
•
Wire break between K6-6 and E1-6 (K6-6 and C1-6)
•
Wire break between K6-7 and E1-7 (K6-7 and C1-7)
•
Wire break between E11-3 and E11-4 (C14-3 and C14-4)
•
Wire break between E9-2 and E9-3 (C12-2 and C12-3)
•
Wire break between C8-2 and K4-5 (powered swivel only)
•
Earth short at K6-6, K6-7, K9-1, K9-2, K9-3, K9-4
B. The lift does not move after operation of the direction switch (DS)
The fault is located in the unit
The lift does not operate in the traction mode
•
A final limit switch is off/activated
•
Wire break between K5-5 and K5-6
•
Wire break between K4-8, K7-3, K7-1, K8-1, K8-3 and K6-6
•
Wire break between K4-9 and K6-7
The lift does operate in the traction mode
•
The tilt switch is activated/faulty/not connected
•
Wire break between K3-1/2/3/4 and the tiltswitch C5 (E8)
•
Earth short K3-1, K3-2, K3-4, K4-4, K4-5, K4-6, K4-7, K4-8, K4-9, K5-4, K2-4
1
|
The lift does not move in both directions
Содержание FREECURVE Basic
Страница 1: ...Handicare Freecurve Technical manual...
Страница 2: ......
Страница 9: ...Installation Version 1 0 9 9 Technical manual...
Страница 14: ...14 Handicare Freecurve...
Страница 15: ...15 Technical manual 11P00177 _ 141021 C25 C22 C23 C24 R R POWERED SWIVEL C26 GRAPHICS NO GRAPHICS...
Страница 42: ...Handicare Freecurve 42...
Страница 47: ...47 Technical manual Classic Alliance Elegance Unit Hinge Sparepart book complete Spareparts...
Страница 84: ...84 Handicare Freecurve...
Страница 85: ......