Parallel operation

In order to increase the total available current the outputs of
supplies can be paralleled. The output voltages of the supplies
involved are adjusted as accurately as possible to the same value.
In this mode it is possible that one or more supplies enter the
current limit mode. The output voltage remains in regulation as
long as still at least supply is in the voltage control mode. It is
recommended but not absolutely necessary to fine adjust the
voltages such that the individual current contributions remain
nearly equal. Of course, the maximum available output current is
the sum of the individual supplies‘ maximum currents.

Example:

A load draws 0.7 A at 12 V. Each output of the HM 8040-3 is
specified for 0.5 A. In order to supply the 0.7 A both outputs have
to be set to 12 V. Then the respective output safety terminals have
to be connected in parallel (black to black, red to red). The load is
connected to one of the supplies. By depressing OUPUT 

 the

outputs are activated . As a rule one output will supply appr. 500
mA in current limit mode, the other the balance of 200 mA in
voltage control mode.

If you parallel power supplies made by HAMEG and
those of other manufacturers please make sure that the
individual current contributions of the supplies involved
are as equal as possible. HAMEG supplies are designed
for series and parallel operation. If supplies of other
manufacturers are used in a parallel connection which
are not overload protected damage may occur if the
individual currents are unequal.

Current limiting

means that a maximum current can be set. This is e.g. useful in
order to protect a sensitive test circuit. In case of an inadvertent
short in the test circuit the current will be limited to the value set
which will in most cases prevent damage.

The picture shows that the output voltage V

out

 remains stable,

while the current I increases until the current limit selected will be
reached. At this moment the instrument will change from constant
voltage regulation to constant current regulation. Any further load
increase will cause the current to remain stable while the voltage
decreases ultimately to zero.

Electronic fuse

This feature provides a still better protection of sensitive loads,
especially of such loads which require more than one voltage: if
one current limit is reached all output voltages will be disconnected
immediately. In order to turn these back on the pushbutton
OUTPUT 

 must be depressed.

Concept of the HM 8040-3

The triple power supply HM8040-3 features 3 independent output
voltages. In addition to supplying 3 voltages the 0 .. 20 V outputs
may be connected in series or parallel.

Exceeding 42 V
If all outputs are series connected the sum can exceed
42 V. In this case touching of live parts may cause
danger of life! It is assumed that only qualified and
extensively instructed personnel is allowed to operate
these supplies and their loads.

In series connection the maximum current is 0.5 A. By paralleling
the outputs a maximum of 1 A is possible. The maximum voltage
remains 20 V. Please note that the specifications (e.g. impedance,
noise, regulation) listed for individual use of the outputs may
change if series or parallel connected.

Output power

As the HM 8040-3 is fed from the mainframe supplies in the
HM 8001-2 (predecessor HM 8001) its maximum output power is
limited to 25 W. Sufficient ventilation is necessary.

The operation of 2 HM 8040-3 in the HM 8001-2 is
only permitted if the sum of all output currents

(these are 6) remains 

 2 A.

Overtemperature protection

If the internal temperature exceeds 75 .. 80 degrees C all outputs
and displays will be deactivated. This will be indicated by 3
blinking dashes in the left and right displays. As soon as the
displays will light up again and display the values set the instrument
will be ready for operation. In order to turn the outputs back on the
pushbutton OUTPUT 

 has to be depressed.

Switching the outputs on/off

It is a feature of all HAMEG power supplies that the outputs can
be turned on and off by pushing a button. The power supply
remains turned on so that all voltages and currents may be set
prior to turning the outputs on.

STOP

+

-

+

-

+

-

U

=  U =  U =  U

1

2

3

total

I

=  I +  I +  I

total

1

2

3

U

3

U

2

U

1

I

2

I

3

I

1

Power supply 1
Output 1

Power supply 1

2

Output

Power supply 2

1

Output

I

out

U

out

U

const

I

max

Current limit

Adjustment of current

Adjustment

of

volt

age

STOP

STOP

Concept of HM 8040-3

Содержание HM8040-3

Страница 1: ...t HM 8040 3 DEUTSCH ENGLISH MANUAL HANDBUCH MANUEL 0 20 V 0 5 A 1 A 0 5 A 0 20 V 5 V VOLTAGE CURRENT VOLTAGE CURRENT V OUTPUT TRIPLE POWER SUPPLY HM8040 3 Instruments F F ON PUSH LONG FUSE ON OFF mA...

Страница 2: ...2 nderungen vorbehalten...

Страница 3: ...icherheitshinweise 6 Bestimmungsgem er Betrieb 6 Garantie 7 Wartung 7 Temperatursicherung 7 Bezeichnung der Bedienelemente 8 Netzger tegrundlagen 8 Lineare Netzteile 8 Parallel und Serienbetrieb 8 Par...

Страница 4: ...lt abgeschirmte Koaxialkabel RG223 U RG214 U verwendet werden 3 Auswirkungen auf die Ger te Beim Vorliegen starker hochfrequenter elektrischer oder magnetischer Felder kann es trotz sorgf ltigen Messa...

Страница 5: ...t HM 8040 3 Bedienungsanleitung Garantiekarte Als weiteres Zubeh r empfehlen wir HZ 10 Silikonmessleitung Technische Daten bei 23 C nach 30 min zum Erreichen der spezifizierten Genauigkeit Ausg nge 2...

Страница 6: ...ssisteile des Einschubmoduls w h rend dem Betrieb im Grundger t mit dem Netzschutzleiter ver bunden Das Einschubmodul zusammen mit dem Grundger t darf aus Sicherheitsgr nden nur an vorschriftsm igen S...

Страница 7: ...noch ist es m glich dass ein Bauteil erst nach l ngerer Betriebsdauer ausf llt Daher wird auf alle Ger te eine Funktions garantie von 2 Jahren gew hrt Vorausetzung ist dass am und im Ger t keine Ver n...

Страница 8: ...ur galvanischen Trennung von Prim rkreis Netzspannung und Sekund rkreis Ausgangsspannung Der nachfolgende Gleichrichter erzeugt eine ungeregelte Gleichspannung Kondensatoren vor und nach dem Regelglie...

Страница 9: ...en Ausgleichsstr me innerhalb der Netzger te flie en HAMEG Netzger te sind f r Parallel und Serien betrieb dimensioniert Verwenden Sie Netzger te ei nesanderenHerstellersalsHAMEG welchenicht ber lasts...

Страница 10: ...ellen und danach mit der Taste OUTPUT an den Verbraucher zuschalten Einf hrung in die Bedienung des HM 8040 3 Achtung Bedienungsanleitung beachten Inbetriebnahme Beachten Sie bitte besonders bei der e...

Страница 11: ...sam auf 0 Volt ab Ist die elektroni sche Sicherung aktiviert werden die Ausg nge beim Zu schalten mit OUTPUT sofort wieder ausgeschaltet Display 7 Segment LED Display mit 3 stelliger Anzeige von Ausga...

Страница 12: ...12 Subject to change without notice...

Страница 13: ...uctions 16 Correct operation 16 Warranty 17 Maintenance 17 Overtemperature protection 17 Designition of operating controls 18 Basics of power supplies 18 Linear regulators 18 Parallel and series opera...

Страница 14: ...t to guard against the intrusion of undesired signals This will not cause destruction or malfunction of HAMEG instruments however small deviations from the guranteed specifications may occur under suc...

Страница 15: ...ssories HZ10 Silicon insulated cable Specifications Valid at 23 degrees C after a 30 minute warm up period Outputs 2 x 0 20 V and 5 V Single pushbutton control of all outputs Linear regulators overtem...

Страница 16: ...n the mainframe The module and its mainframe may only be connected to a mains outlet with a safety ground conductor or to an isolation transformer In case of doubt whether a mains outlet carries a saf...

Страница 17: ...h have not been tampered with Also the warranty will be voided for any instruments damaged by negligence or by unsuitable transport packaging Only the original packaging may be used for transport by p...

Страница 18: ...same value If in any of the series connected power supplies the current limit will be reached this supply will enter the current limiting mode so the total output voltage will decrease and become unre...

Страница 19: ...he current to remain stable while the voltage decreases ultimately to zero Electronic fuse This feature provides a still better protection of sensitive loads especially of such loads which require mor...

Страница 20: ...2 outputs have to be set according to the procedure outlined above The pushbutton V mA Fuse is held depressed for a long time after the LED F lights up the electronic fuse mode is active If any of th...

Страница 21: ...pertinent only for the 20 V outputs The range is 0 500 mA If the control is fully CCW the outputs will be disabled immediately if the mode Electronic Fuse was selected In Current Limit mode the LED I...

Страница 22: ...800 0 Telefax 0 61 82 800 100 E mail sales hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers Power Supplies CurveTrace...

Отзывы: