speed will be increased by a factor of 10, hence the time base 
B display will be expanded 10 fold in X direction. 

29.3 ”B only“

 

 The time defl ection coeffi cient reduces by factor 10 causing a 
tenfold X magnifi cation (signal expansion). 

30

 HOR 

/

 VAR 

(pushbutton) 

This pushbutton will open the ”Time base“ menu the contents 
of which depend on the operating mode selected. 

30.1 Analog mode 

The following modes are available: 

30.1.1

 A only 

In this setting only time base A is active and thus in the top left 
position the readout displays only “A....“. The TIME/DIV-SCA-
LE-VAR knob 

28

 only effects time base A. By aid of the MAG 

x10 pushbutton 

29

 the signal display can be expanded 10 fold; 

reducing the time defl ection coeffi cient by factor 10.  

If the mode is changed from time base A to ”Search“ or ”B only“ 
all settings of time base A remain intact including triggering. 

30.1.2

 Search 

This mode implies alternate time base operation. The readout 
will show the speeds of both time bases (”A...“ and ”B...“). The 
TIME/DIV-SCALE-VAR knob will set the time base B speed. 
In alternate time base mode part of the time base A display will 
be intensified. The horizontal position of the intensified portion 
may be shifted using the knob HORIZONTAL 

27

, provided its 

function was set to ”Delay“, this is the case if the pushbutton 
X-POS DELAY 

26

 is illuminated. The length of the intensified 

sector is determined by the speed of time base B and is equal 
to the run time of B. Hence this intensified portion of A will be 
spread over the full screen as displayed by B and thus expanded. 
The Y position of the signal is the same when displayed with 
either A or B. This means, however, that both displays will be 
written over each other. 

This can be avoided by changing the time base B trace position in 
the following way: Press the CH1/2–CURSOR–MA/REF–ZOOM-
Taste 

15

 to call the “Pos./Scale“ menu. Then press the function 

key “TB B“ so that – depending on the channel mode - the PO-
SITION 1 and /or POSITION 2 knob(s) 

13

 

14

 become(s) the trace 

separation control(s) (see 13.1.5 Y Position - 2

nd

 time base). As 

there is only a need for trace separation in “Search mode“ this 
function is only offered in this time base mode.

Also in ”Search“ the 10 x magnifier is available by pressing the 
pushbutton MAG x10 

29

. The magnifier will affect solely time 

base B. 

30.1.3 

”B only“ 

In this mode only time base B will be displayed. Thus the readout 
only displays in the top left position only “B“. The TIME/DIV-SCA-
LE-VAR knob then only affects time base B. By aid of the MAG 
x10 pushbutton 

29

 the signal display can be expanded 10 fold; 

reducing the time defl ection coeffi cient by factor 10.  

30.1.4

 B trigger – 

 Edge 

In this mode time base B will not start immediately after the set 
delay time has elapsed, but will be only set ready, waiting for a 
signal trigger. This has the advantage that any jitter is removed, 
but the delay time adjustment will now only have the effect that 
the time base B display will jump from signal period to period 
if several periods are displayed. In this setting a positive slope 
will trigger. 

The (trigger) LEVEL A/B 

19

 knob will set the trigger level for B. 

Only normal triggering and DC coupling are possible. All para-
meters of time base A remain stored and preserved. (LEVEL, 
auto or normal, Slope, coupling). In addition to the delay time 
(”Dt:...“) also the B trigger parameters are shown in the readout: 
”BTr:slope, DC“. In ”Search“ mode the trigger point symbol will 
be preceded by ”B“. As mentioned changing the delay time will 
not cause a continuous move of the intensified portion of the 
time base A display and the time base B display, but jumps from 
signal period to period. 

If the trigger level symbol of time base B is shifted outside the 
signal representation by time base A there will be no triggering 
of time base B any more and thus no time base B display. The 
same holds in time base B only mode. 

30.1.5

 B trigger – 

 Edge 

Except for the negative edge the function is identical to the one 
described above (30.1.4). 

30.1.6

 B trigger – OFF 

Time base B will be started upon the end of the delay time set. 
The delay time can be changed continuously in this mode which 
can be seen as the intensified sector of the time base A display. 
The disadvantage here is that with very long delay times, jitter 
of the time base B display may occur. As time base B is not 
operated in the signal triggered mode, the controls for time 
base B trigger will be disabled. 

30.1.7

 A variable – On Off 

If ”On“ was selected the TIME/DIV-SCALE-VAR knob 

28

 will 

function as variable control for the time base A speed. Only 
in time base A only mode will this option be available in the 
menu. 
For a full description see ”28.1.3 Variable“. 

30.1.8

 B variable – On Off

If ”On“ was selected the TIME/DIV-SCALE-VAR knob 

28

 will 

          C o n t r o l s  a n d  R e a d o u t

VAR

VAR

VAR

x10

AUXILIARY INPUT

LC

0 ... LC3

LOGIC PROBE

Use recommended

probe only!

TRIG. EXT. / Z-INP.

LC/AUX

FFT

1M

Ω

 II

15pF

max

100 Vp

INPUTS

1M

Ω

II15pF

max

250 Vp

X-INP

CH 1

VERT/XY

CH 2

HOR

MAG

CAT I

!

50

Ω

 

5V RMS

50

Ω

 

5V RMS

CH 1

CH 2

!

PROBE

ADJ

00 

POWER

MENU

MENU

OFF

OFF

44

31

34

32

33

35

36

38

37

39

29

30

Содержание CombiScope HM2008

Страница 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Страница 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Страница 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Страница 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Страница 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Страница 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Страница 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Страница 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Страница 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Страница 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Страница 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Страница 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Страница 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Страница 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Страница 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Страница 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Страница 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Страница 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Страница 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Страница 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Страница 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Страница 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Страница 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Страница 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Страница 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Страница 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Страница 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Страница 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Страница 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Страница 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Страница 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Страница 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Страница 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Страница 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Страница 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Страница 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Страница 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Страница 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Страница 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Страница 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Страница 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Страница 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Страница 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Страница 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Страница 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Страница 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...

Страница 47: ...ive is connected with the USB Stick connector 43 41 4 1 USB Stick Save Front Panel current oscilloscope parameters The current parameters are saved on the USB ash drive by pressing the function key Sa...

Страница 48: ...D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com w w w h a m e g c o m 41 2008 00E1 authori...

Отзывы: