Both sync pulses differ hence as well in duration as in their 
repetition intervals. Triggering is possible with both.

Frame sync pulse triggering

Remark: 
Using frame sync triggering in dual trace chopped mode may 
result in interference, then the dual trace alternate mode 
should be chosen. It may also be necessary to turn the rea-
dout off.

In order to achieve frame sync pulse triggering call MODE, 
select video signal triggering and then FILTER to select frame 
triggering.  It may be selected further whether ”all“, ”only even“ 
or ”only odd“ half frames shall trigger. Of course, the correct tv 
standard must be selected fi rst of all (625/50 or 525/60).
The time base setting should be adapted, with 2 ms/cm a com-
plete half frame will be displayed. Frame sync pulses consist 
of several pulses with a half line rep rate. 

Line sync pulse triggering

In order to choose line snyc triggering call MODE and select 
VIDEO, enter FILTER, make sure that the correct video standard 
is selected (625/50 or 525/60) and select Line.
If ALL was selected each line sync pulse will trigger. It is also 
possible to select a line number ”LINE No.“.

In order to display single lines a time base setting of TIME/DIV. 
= 10 μs/cm is recommended, this will show 1½ lines. In general 
the composite video signal contains a high DC component which 
can be removed by ac coupling, provided the picture is steady. 
Use the POSITION control to keep the display within the screen. 
If  the video content changes like with a regular TV program 
only DC coupling is useful, otherwise the vertical position would 
continuously move.  

The sync separator is also operative with external triggering. 
Consult the specifi cations for the permissible range of trigger 
voltage. The correct slope must be chosen as the external 
trigger may have a different polarity from the composite video. 
In case of doubt display the external trigger signal.

LINE trigger

Consult SOURCE 

22

 in ”Controls and Readout“ for specifi c 

information.

If the readout shows Tr:Line the trigger signal will be internally 
taken from the line (50 or 60 Hz).

This trigger signal is independent of the scope input signals and 
is recommended for all signals synchronous with the line. Within 
limits this will also be true for multiples or fractions of the line 
frequency.  As the trigger signal is taken off internally there is 
no minimum signal height on the screen for a stable display. 
Hence even very small voltages like ripple or line frequency 
interference can be measured.

Please note that with line triggering the polarity switching will 
select either the positive or negative half period of the line, not 
the slope. The trigger level control will move the trigger point 
over most of a half wave. 

Line frequency interference may be checked using a search 
coil which preferably should have a high number of turns and 

a shielded cable. Insert a 100 Ω resistor between the center 
conductor and the BNC connector. If possible the coil should 
be shielded without creating a shorted winding. 

Alternate trigger

This mode is selected with SOURCE 

22

 

>

 Alt. 1/2. The read-

out will display Tr:alt, but no more the trigger point symbol 
indicating level and time position. Instead an arrow pointing 
upwards will indicate the trigger time position if this lies within 
the screen area. 

This trigger mode is to be used with greatest care and should be 
an exception rather than the rule, because the time relationships 
visible on the screen are completely meaningless, they depend 
only on the shape of the signals and the trigger level! 

In this mode the trigger source will be switched together with 
the channel switching, so that when CH1 is displayed in the 
dual channel alternate mode the trigger is taken from CH1 
and when CH2 is displayed the trigger is taken from CH2. This 
way two uncorrelated signals can be displayed together. If this 
mode is inadvertently chosen the time relationships between 
the signals will also be lost when both signals are correlated! 
(Except for the special case that both happen to be square waves 
with extremely fast rise times). Of course, this trigger mode is 
only possible in the dual channel alternate mode and also not 
with external or line trigger. Ac coupling is recommended for 
most cases. 

External triggering

In analog mode this trigger mode may be selected with SOURCE 

22

 

>

 Extern. In digital mode it is possible too if the logic channels 

are activated. The readout will display Tr:ext. The AUXILIARY 
INPUT 

38

 will be the input for the external trigger, all internal 

sources will be disconnected. In this mode the trigger point 
symbol (level and time position) will not be displayed, only 
the trigger time position will be indicated. External triggering 
requires a signal of 0.3 to 3 V

pp

, synchronous to the vertical 

input signal(s). 

Triggering will also be possible within limits with multiples or 
fractions of the vertical input signal frequency. As the trigger 
signal may have any polarity it may happen that the vertical 
input signal will start with a negative slope in spite of having 
selected positive slope; slope selection refers now to the ex-
ternal trigger.

Indication of triggered operation (TRIG’D LED)

Refer item 

23

 in ”Controls and Readout“. The LED labelled 

TRIG’D indicates triggered operation provided:
– Suffi cient amplitude of the internal or external trigger signal.
–  The trigger point symbol is not above or below the signal.

If these conditions are met the trigger comparator will output 
triggers to start the time base and to turn on the trigger indi-
cation. The trigger indicator is helpful for setting the trigger 
up, especially with low frequency signals (use normal trigger) 
resp. very short pulses.

The trigger indication will store and display triggers for 100 ms. 
With signals of very low rep rate the indicator will fl ash accor-
dingly. If more than one signal period is shown on the screen 
the indicator will fl ash each period.

          T r i g g e r i n g  a n d  t i m e  b a s e

Содержание CombiScope HM2008

Страница 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Страница 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Страница 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Страница 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Страница 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Страница 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Страница 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Страница 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Страница 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Страница 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Страница 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Страница 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Страница 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Страница 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Страница 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Страница 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Страница 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Страница 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Страница 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Страница 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Страница 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Страница 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Страница 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Страница 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Страница 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Страница 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Страница 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Страница 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Страница 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Страница 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Страница 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Страница 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Страница 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Страница 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Страница 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Страница 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Страница 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Страница 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Страница 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Страница 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Страница 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Страница 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Страница 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Страница 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Страница 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Страница 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...

Страница 47: ...ive is connected with the USB Stick connector 43 41 4 1 USB Stick Save Front Panel current oscilloscope parameters The current parameters are saved on the USB ash drive by pressing the function key Sa...

Страница 48: ...D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com w w w h a m e g c o m 41 2008 00E1 authori...

Отзывы: