background image

11

Subject to change without notice

of the capacitor. It may be assumed that this is negligible for 
frequencies 

>

40 Hz. 

Considering the foregoing you may measure DC signals of up to 
600 V or pure AC signals of up to 1200 V

PP

 with a HZ200 probe. 

Probes with higher attenuation like HZ53 100:1 allow to measure 
DC up to 1200 V and pure AC of up to 2400 V

PP

. (Please note 

the derating for higher frequencies, consult the HZ53 manual). 
Stressing a 10:1 probe beyond its ratings will risk destruction of 
the capacitor bridging the input resistor with possible ensuing 
damage of the scope input! 

In case the residual ripple of a high voltage is to be measured a 
high voltage capacitor may be inserted in front of a 10:1 probe, it 
will take most of the voltage as the value of the probe’s internal 
capacitor is very low, 22 to 68 nF will be suffi cient. 

If the input selector is switched to Ground the reference trace 
on the screen may be positioned at graticule center or else-
where.

DC and AC components of an input signal

The dashed curve shows an AC signal symmetrical to zero. If 
there is a DC component the peak value will be DC + AC peak.

Timing relationships

In most cases repetitive signals must be measured. The repe-
tition frequency of a signal is equal to the number of periods 
per second. Depending on the TIME/DIV setting one or more 
periods or part of a period of the signal may be displayed. The 
time base settings will be indicated on the readout in s/cm, 
ms/cm, μs/cm and ns/cm (1 cm is the equivalent of 1 div. on 
the crt graticule). Also the cursors may be used to measure the 
frequency or the period.

Without cursor the cycle duration can be determined by multi-
plying the length (cm) with the (calibrated) time coeffi cient. The 
reciprocal value is the frequency.

If portions of the signal are to be measured use delayed sweep 
(analog mode) or zoom (digital mode) or the magnifi er x 10. Use 
the HORIZONTAL positioning control to shift the portion to be 
zoomed into the screen center.

Pulse signals are characterized by their rise and fall times 
which are measured between the 10 % and 90 % portions. The 
following example uses the internal graticule of the crt, but also 
the cursors may be used for measurement. 

Measurement:

–  Adjust the rising portion of the signal to 5 cm.

–  Position the rising portion symmetrically to the graticule 

centre line, using both Y and X positioning controls.

–  Notice the intersections of the signal with the 10 and 90 % 

lines and project these points to the centre line in order to 
read the time difference. 

In the example it was 1.6 cm at 5 ns/cm equals 8 ns rise time.

When measuring very short rise times coming close to the scope 
rise time it is necessary to subtract the scope’s  (and if used the 
probe’s) rise times geometrically from the rise time as seen on 
the screen. The true signal rise time will become:

t

tot

 is the rise time seen, t

osc

 is the scope’s own rise time (1.75 

ns with the HM2008), t

t

 is the rise time of the probe, e.g. 2 ns. 

If the signal’s rise time is 

>

22 ns, the rise times of scope and 

probe may be neglected. 

For the measurement of rise times it is not necessary to proceed 
as outlined above. Rise times may be measured anywhere on 
the screen. It is mandatory that the rising portion of the signal 
be measured in full and that the 10 to 90 % are observed. In 
case of signals with over- or undershoot the 0 and 100 % levels 
are those of the horizontal portions of the signal, i.e. the over- 
resp. undershoots must be disregarded for rise and fall time 
measurements. Also, glitches will be disregarded. If signals 
are very distorted, however, rise and fall time measurements 
may be of no value. 

For most amplifi ers, even if their pulse behaviour is far from 
ideal, the following relationship holds:

 350 

350

t

a

  =   ——             B  =   ——  

 B 

t

a

tr/ns  = 350/Bandwidth/MHz  

Connection of signals

In most cases pressing the AUTOSET button will yield a satisf-
actory display (see AUTOSET). The following relates to special 
cases where manual settings will be advisable. For a description 
of controls refer to ”Controls and Readout“.

STOP

  Take care when connecting unknown signals to the 

inputs!

It is recommended to use probes whenever possible. Without 
a probe start with the attenuator set to its 20 V/cm position. 
If the trace disappears the signal amplitude may be too large 
overdriving the vertical amplifi er or/and its DC content may be 
too high. Reduce the sensitivity until the trace will reappear 

          B a s i c  s i g n a l  m e a s u r e m e n t

t

a

   

8

2

 - 2.3

2

 - 2

2

  =  7,5 ns

t

a

   

t

tot

2

  – t

osc

2

 – t

t

2

voltage

peak

AC

DC

DC

AC

DC + AC

peak

 = 400 V

max

5 cm

t

tot

100%

90%

10%

0%

Содержание CombiScope HM2008

Страница 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Страница 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Страница 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Страница 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Страница 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Страница 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Страница 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Страница 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Страница 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Страница 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Страница 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Страница 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Страница 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Страница 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Страница 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Страница 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Страница 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Страница 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Страница 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Страница 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Страница 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Страница 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Страница 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Страница 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Страница 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Страница 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Страница 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Страница 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Страница 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Страница 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Страница 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Страница 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Страница 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Страница 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Страница 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Страница 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Страница 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Страница 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Страница 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Страница 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Страница 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Страница 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Страница 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Страница 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Страница 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Страница 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...

Страница 47: ...ive is connected with the USB Stick connector 43 41 4 1 USB Stick Save Front Panel current oscilloscope parameters The current parameters are saved on the USB ash drive by pressing the function key Sa...

Страница 48: ...D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com w w w h a m e g c o m 41 2008 00E1 authori...

Отзывы: