7

Subject to change without notice

          I m p o r t a n t  h i n t s

Type of fuse:                                                                    

Size 5 x 20 mm; 250V~, C;
IEC 127, Bl. III; DIN 41 662 
(or DIN 41 571, Bl. 3).
Cut off: slow blow (T) 0,8A.

The oscilloscope may only be operated from mains outlets with 
a safety ground connector. The plug has to be installed prior to 
connecting any signals. It is prohibited to separate the safety 
ground connection. 
Most electron tubes generate X-rays; the ion dose rate of this in-
strument remains well below the 36 pA/kg permitted by law.
In case safe operation may not be guaranteed do not use the 
instrument any more and lock it away in a secure place. 

Safe operation may be endangered if any of the following 
was noticed:

–   in case of visible damage.
–  in case loose parts were noticed
–  if it does not function any more.
–  after prolonged storage under unfavourable  conditions (e.g. 

like in the open or in moist atmosphere).

–  after any improper transport (e.g. insuffi cient packing not 

conforming to the minimum standards of post, rail or trans-
port fi rm)

Proper operation

Please note: This instrument is only destined for use by person-
nel well instructed and familiar with the dangers of electrical 
measurements. For safety reasons the oscilloscope may only 
be operated from mains outlets with safety ground connector. 
It is prohibited to separate the safety ground connection. The 
plug must be inserted prior to connecting any signals.

CAT I

This oscilloscope is destined for measurements in circuits not 
connected to the mains or only indirectly. Direct measurements, 
i.e. with a galvanic connection to circuits corresponding to the 
categories II, III, or IV are prohibited!
The measuring circuits are considered not connected to the 
mains if a suitable isolation transformer fulfi lling safety class 
II is used. Measurements on the mains are also possible if 
suitable probes like current probes are used which fulfi l the 
safety class II. The measurement category of such probes must 
be checked and observed.

Measurement categories

The measurement categories were derived corresponding to 
the distance from the power station and the transients to be 
expected hence. Transients are short, very fast voltage or cur-
rent excursions which may be periodic or not. 

Measurement CAT IV:

 Measurements close to the power station, 

e.g. on electricity meters

Measurement CAT III:

 Measurements in the interior of buildings 

(power distribution installations, mains outlets, motors which 
are permanently installed).

Measurement CAT II:

 Measurements in circuits directly connec-

ted to the mains (household appliances, power tools etc).

Measurement CAT I:

 Electronic instruments and circuits which 

contain circuit breakers resp. fuses. 

Environmental conditions

The oscilloscope is destined for operation in industrial, business, 
manufacturing, and living sites.
Operating ambient temperature: +5 °C to +40 °C. During 
transport or storage the temperature may be –20 °C to +70°C. 
Please note that after exposure to such temperatures or in 
case of condensation proper time must be allowed until the 
instrument has reached the permissible temperature, resp. 
until the condensation has evaporated before it may be turned 
on! Ordinarily this will be the case after 2 hours. The oscillo-
scope is destined for use in clean and dry environments. Do 

not operate in dusty or chemically aggressive atmosphere or if 
there is danger of explosion. 
The operating position may be any, however, suffi cient ventila-
tion must be ensured (convection cooling). Prolonged operation 
requires the horizontal or inclined position. 

STOP

 Do not obstruct the ventilation holes!

Specifi cations are valid after a 30 minute warm-up period 
between 15 and 30 degr. C. Specifi cations without tolerances 
are average values.

Warranty and repair

HAMEG instruments are subjected to a strict quality control. 
Prior to leaving the factory, each instrument is burnt-in for 10 
hours. By intermittent operation during this period almost all 
defects are detected. Following the burn-in, each instrument is 
tested for function and quality, the specifi cations are checked 
in all operating modes; the test gear is calibrated to national 
standards.
The warranty standards applicable are those of the country 
in which the instrument was sold. Reclamations should be 
directed to the dealer.

Only valid in EU countries

In order to speed reclamations customers in EU countries may 
also contact HAMEG directly. Also, after the warranty expired, 
the HAMEG service will be at your disposal for any repairs.

Return material authorization (RMA):

Prior to returning an instrument to HAMEG ask for a RMA 
number either by internet (http://www.hameg.com) or fax. If 
you do not have an original shipping carton, you may obtain one 
by calling the HAMEG service dept (++49 (0) 6182 800 500) or by 
sending an email to [email protected]

Maintenance

Clean the outer shell using a dust brush in regular intervals. 
Dirt can be removed from housing, handle, all metal and plastic 
parts using a cloth moistened with water and 1 % detergent. 
Greasy dirt may be removed with benzene (petroleum ether) or 
alcohol, there after wipe the surfaces with a dry cloth. Plastic 
parts should be treated with an antistatic solution destined 
for such parts. No fl uid may enter the instrument. Do not use 
other cleansing agents as they may adversely affect the plastic 
or lacquered surfaces.

Line voltage

The instrument has a wide range power supply from 105 to 
253 V, 50 or 60 Hz ±10%. There is hence no line voltage selector.
The line fuse is accessible on the rear panel and part of the line 
input connector. Prior to exchanging a fuse the line cord must be 
pulled out. Exchange is only allowed if the fuse holder is undama-
ged, it can be taken out using a screwdriver put into the slot. The 
fuse can be pushed out of its holder and exchanged.
The holder with the new fuse can then be pushed back in place 
against the spring. It is prohibited to ”repair“ blown fuses or to 
bridge the fuse. Any damages incurred by such measures will 
void the warranty.

Содержание CombiScope HM2008

Страница 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Страница 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Страница 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Страница 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Страница 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Страница 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Страница 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Страница 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Страница 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Страница 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Страница 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Страница 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Страница 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Страница 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Страница 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Страница 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Страница 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Страница 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Страница 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Страница 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Страница 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Страница 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Страница 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Страница 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Страница 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Страница 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Страница 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Страница 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Страница 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Страница 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Страница 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Страница 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Страница 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Страница 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Страница 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Страница 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Страница 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Страница 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Страница 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Страница 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Страница 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Страница 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Страница 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Страница 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Страница 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Страница 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...

Страница 47: ...ive is connected with the USB Stick connector 43 41 4 1 USB Stick Save Front Panel current oscilloscope parameters The current parameters are saved on the USB ash drive by pressing the function key Sa...

Страница 48: ...D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com w w w h a m e g c o m 41 2008 00E1 authori...

Отзывы: