background image

33

Subject to change without notice

chosen by the INTENS knob 

2

 and after the CH1/2–CURSOR–MA/

REF–ZOOM pushbutton 

15

 

has been activated, “Math./Ref“ has 

been selected (pushbutton light green) 

The POSITION 2 knob can also serve as a Y Position control for 
mathematics signals, if  the previously mentioned conditions  
are met and in the “Mathematics“ menu the lower “Display” 
sector shows a selected equation (MA1 ... MA5) instead of off. 

13.1.5

 Y position of 2nd time base B (analog mode). 

The POSITION 1 control will assume the function of Y position 
control of each assigned channel signal displayed by time base 
B in alternate time base mode after the following procedure. 
This is convenient in order to be able to separate the displays 
of the (same) signal with both time bases on the screen. Press 
the HOR VAR 

30

 pushbutton 

>

 ”Search“. Press the CH1/2–CUR-

SOR–MA/REF–ZOOM pushbutton 

15

, select the function ”TB B“. 

The pushbutton will light up green.
 

13.1.6

 Y position for ZOOM (digital mode.) 

This is the equivalent of the former in digital mode, where the 
function is called ZOOM. The POSITION 1 control will assume 
the function of Y position control of each assigned channel 
signal displayed under ZOOM conditions after the following 
procedure. Again, the intent is to be able to separate the dis-
plays of the (same) signal on the screen in alternate time base 
mode. Press the HOR VAR 

30

 pushbutton 

>

”Search“. Press the 

CH1/2–CURSOR–MA/REF–ZOOM pushbutton 

15

 , select ”TB B“. 

The pushbutton will light up green.  

13.2.  X position in XY mode

 (analog and digital mode) 

POSITION 1 will function as X position control of CH1 in XY mode 
and provided the CH1/2–CURSOR–MA/REF–ZOOM pushbutton 

15

 is not illuminated. 

STOP

 Note: 

 

The HORIZONTAL control 

27

 will also be functional 

in XY mode. 

13.3. CURSOR position

 (analog and digital mode) 

The POSITION 1 knob 

13

 can be used as CURSOR Position con-

trol on condition that CURSOR display is switched on by pressing 
the CURSOR-MEASURE pushbutton 

4

 and “Cursors“ or “Cur. 

Track“ has been chosen after pressing the CH1/2–CURSOR–MA/
REF–ZOOM pushbutton 

15

; pushbutton lit blue.

 

STOP

 Please 

note: 

 

The function ”Cur. Track“ is only available if two 
cursors are indeed displayed, then both cursors 
can be moved simultaneously (tracking) without a 
change of their respective positions.

14

 

POSITION 2 

(knob)

 

Also this control may assume diverse functions dependent on the 
operating mode, the function selected via the CH1/2–CURSOR–
MA/REF–ZOOM-pushbutton 

15

 and the menu item activated. 

14.1. Y position 

14.1.1

 Y-Position of channel 2 (analog and digital mode)

POSITION 2 is for Y Position CH2 control, if Yt mode (time base 
mode) is present and the CH1/2–CURSOR–MA/REF–ZOOM-
pushbutton 

15

 is not lit.

14.1.2

 Y-Position of the FFT display (only in digital mode) 

measured with channel 2

Position 2 allows you to move the FFT spectrum display, the 
FFT reference indicator (arrow symbol at the left side of the 
graticule) and the FFT marker (X Symbol) in vertical direction.

14.1.3

 Y position of reference signals (digital mode) 

The POSITION 2 control will function as Y position control of 
reference signals if the following conditions are fulfilled: 

1

st

  A reference signal must on display. (SAVE/RECALL push-

button 

9

 

>

 Reference Display 

>

 (upper display area) Rex 

(x = number of memory location, select with INTENS) 

>

 On 

(with or without associated settings). 

2

nd

 Press  CH1/2–CURSOR–MA/REF–ZOOM-pushbutton 

15

 

>

 

Math./Ref. The pushbutton will light up green. 

14.1.4

 Y position of mathematics signals (digital mode) The 

POSITION 2 control functions as Y position control for mathe-
matics signals, if the following conditions are fulfilled: Press 
the MATH pushbutton 

7

 

>

 displays (upper display area), select 

an equation with the INTENS knob 

4

 (MA1.. MA5). Press the 

CH1/2–CURSOR–CH3/4–MA/REF–ZOOM-pushbutton 

15

, select 

Math./Ref. The pushbutton will light up green. 

14.1.5

 Y position of 2nd time base B (analog mode). 

The POSITION 2 control will assume the function of Y position 
control of each assigned channel signal displayed by time base 
B in alternate time base mode after the following procedure. 
This is convenient in order to be able to separate the displays 
of the (same) signal with both time bases on the screen. Press 
the HOR VAR 

30

 pushbutton 

>

 ”Search“. Press the CH1/2–CUR-

SOR–MA/REF–ZOOM pushbutton 

15

, select the function ”TB B“. 

The pushbutton will light up green.
 

14.1.6

 Y position for ZOOM (digital mode.) 

This is the equivalent of the former in digital mode, where the 
function is called ZOOM. The POSITION 2 control will assume 
the function of Y position control of each assigned channel 
signal displayed under ZOOM conditions after the following 
procedure. Again, the intent is to be able to separate the dis-
plays of the (same) signal on the screen in alternate time base 
mode. Press the HOR VAR 

30

 pushbutton 

>

”Search“. Press the 

CH1/2–CURSOR–MA/REF–ZOOM pushbutton 

15

 , select ”TB B“. 

The pushbutton will light up green.  

14.2  Y position of CH2 in XY mode

 (Analog and digital mode.) 

POSITION 2 will function as the Y position control of CH2 in XY 
mode provided the CH1/2–CURSOR–MA/REF–ZOOM pushbutton 

15

 is not illuminated. 

14.3 CURSOR position

 (analog and digital mode) 

The POSITION 2 knob 

14

 can be used as CURSOR Position con-

          C o n t r o l s  a n d  R e a d o u t

VOLTS / DIV

SCALE 

·

 VAR

VOLTS / DIV

SCALE 

·

 VAR

AUTO

MEASURE

5 V

1 mV

5 V

1 mV

CH 1/2

CURSOR

MA/REF

ZOOM

POSITION 1

POSITION 2

POWER

MENU

OFF

13

15

14

17

16

18

Содержание CombiScope HM2008

Страница 1: ...200 MHz Mixed Signal CombiScope HM2008 Manual English...

Страница 2: ...irective des equipements basse tension 73 23 CEE amend e par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonis es utilis es Sicherheit Safety S curit EN 61010 1 2...

Страница 3: ...ajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggering MODE menu 15 Normal trigger mode...

Страница 4: ...t voltage Time Base 2ns div 50s div Acquisition modes Single Refresh Average Envelope Roll Peak Detect Front USB Stick Connector for Screenshots USB RS 232 optional IEEE 488 or Ethernet USB Signal dis...

Страница 5: ...range 0 250 MHz Coupling DC Level control range 10 10 div Horizontal Deflection Analog Time Base Operating modes A ALT alternating A B B Time base A 20 ns div 0 5 s div 1 2 5 sequence Time base B 20 n...

Страница 6: ...rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked A A B B C C D D E E T F PUkT PUkT PUk PUk PUk PUk PUk PUk PUkT...

Страница 7: ...may be 20 C to 70 C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature resp until...

Страница 8: ...V SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling setting 18 AUTO MEASURE p...

Страница 9: ...selection of PROBE ADJ square wave signal hard ware and software information and details about interface rear side and USB Stick ash drive connector 42 COMPONENT TESTER 2 sockets with 4 mm 47 Connecto...

Страница 10: ...st be taken however when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty c...

Страница 11: ...ursors may be used for measurement Measurement Adjust the rising portion of the signal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning con...

Страница 12: ...e ground connections of probes are especially critical they should be as short as possible and of large size STOP If a probe is to be connected to a BNC connector use a probe tip to BNC adapter If rip...

Страница 13: ...mind The oscillosco pe vertical ampli ers are two separate ampli ers and do not constitute a true difference ampli er with as well a high CM rejection as a high permissible CM range Therefore please o...

Страница 14: ...and Y signals af ter reading a and b off the screen is possible using the following formulas and a pocket calculator with trigonometric functions This calculation is independent of the signal amplitud...

Страница 15: ...ime base variable or external triggering Reading a and b off the screen the modulation degree will result a b a b m bzw m 100 a b a b a UT 1 m and b UT 1 m When measuring the modulation degree the amp...

Страница 16: ...ange the slope Positive or negative slope may be selected in auto or normal trigger modes Also a setting both may be selected which will cause a trigger irrespective of the polarity of the next slope...

Страница 17: ...Insert a 100 resistor between the center conductor and the BNC connector If possible the coil should be shielded without creating a shorted winding Alternate trigger This mode is selected with SOURCE...

Страница 18: ...time for most of the time the digital will not see the signal Also in analog mode the signal itself will be seen on the screen in real time whereas a digital can only show a reconstruction of the sig...

Страница 19: ...er COMPONENT PROBE 40 and COMPONENT TESTER 41 The scope has a built in component tester The test object is connected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned of...

Страница 20: ...pend strongly on the kind of FET With depletion type MOSFETs and all JFETs the channel will conduct if prior to testing the gate was connected to the source The Rdson will be shown As this can be very...

Страница 21: ...ependent single event recording with Yt display Single Readout display sgl 4 Trigger independent continuous recording menu VERT XY with XY display XY Readout display XY 5 XY display of previously in Y...

Страница 22: ...value of the sampling interval 1 sampling interval sampling rate 3 The signal frequency value is related to the highest sine wave signal frequency that can be captured with 10 samples per period realt...

Страница 23: ...s shows a straight line in the position of the positive peak value and consequently suggest a positive DC voltage An alias may also take the form of a signal of much lower frequency beat frequency bet...

Страница 24: ...rogram for USB connection test RS 232 The RS 232 interface has the usual 9 pole SubD connector Via this bidirectional interface the scope can be controlled remo tely or its settings may be transferred...

Страница 25: ...ilable submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If a menu is shown some other information dis played in...

Страница 26: ...to leave the menu Thereafter only the signal display is activated and the FOCUS TRACE MENU pushbutton will blink as long as the readout is off Pressing the ashing pushbutton calls the Int Knob menu wh...

Страница 27: ...hannel mode Yt in the section First time operation and pre settings One period of a sine wave is equal to 2 hence the dis tance between the two long CURSOR lines must be set to one period If the dista...

Страница 28: ...will be lost after turn off Mathematics offers 7 1 Equations set Using the INTENS knob 5 sets of formulas can be selected for editing This way 5 user de ned formula sets may be created Each set of fo...

Страница 29: ...he display of the former The display will remain on screen until the next acquisition is started by the trigger and causes a new recording This mode is available over the full time base range 50 s cm...

Страница 30: ...size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak Detect is that signals are sampled with the highest possible sampling rate to reduce the gap size so that eve...

Страница 31: ...RE1 RE1 the signal will be displayed twice on the same spot Associated settings Pressing this function key causes the oscilloscope to take over the settings parameter stored when the signal was store...

Страница 32: ...s will remain The digital modes Roll Envelope or Average will be changed to refresh mode AUTOSET will further set the intensity to an average value if it was set too low If a menu was opened it will b...

Страница 33: ...control if Yt mode time base mode is present and the CH1 2 CURSOR MA REF ZOOM pushbutton 15 is not lit 14 1 2 Y Position of the FFT display only in digital mode measured with channel 2 Position 2 allo...

Страница 34: ...he CH1 menu the de ection coef cient becomes calibrated the CH1 VAR pushbutton 31 does not light any longer and the VOLTS DIV SCALE VAR knob 16 returns to 1 2 5 sequence 16 3 Scaling the FFT display d...

Страница 35: ...at V rms the range is from 5mV cm to 20V cm also switched in 1 2 5 sequence Please note that in contrast to Yt and XY mode signal display no peak to peak values but rms values are displayed 18 AUTO ME...

Страница 36: ...me base set B trigger to positive or negative slope Thereafter the LEVEL A B control will control the time base B trigger a second trigger point symbol will be shown and marked with B 19 2 FFT mode Th...

Страница 37: ...mpli er bandwidth and consequently less trigger signal noise Readout Tr Source Slope AC or DC NR 21 1 2 Slope SLOPE determines whether the rising or falling portion of a signal shall trigger the level...

Страница 38: ...5 External In this mode the trigger signal comes from the AUXILIARY INPUT TRIG EXT 38 Readout Tr ext Slope Filter 22 1 6 AC Line The trigger signal is taken from the line mains supply which feeds the...

Страница 39: ...ither reference nor mathematic signals are displayed The expanded portion of the signal will be shown on the normal display as an intensi ed sector The length of this sector is dependent upon the sett...

Страница 40: ...ave form become digitised and at least show spectrum displays spectra that do not exist in reality If the sampling rate is too low the readout displays ALS if the signal is too high overrange will be...

Страница 41: ...nly B The TIME DIV SCA LE VAR knob then only affects time base B By aid of the MAG x10 pushbutton 29 the signal display can be expanded 10 fold reducing the time de ection coef cient by factor 10 30 1...

Страница 42: ...sensitivity setting is for AC and is for DC 31 1 1 DC coupling The signal will be directly coupled from the BNC connector via the attenuator to the vertical ampli er The input resistance is 1 M in all...

Страница 43: ...DUAL trace alt chop In dual trace mode both channels are turned on and the para meters of both are shown in the readout Between the sensi tivity indications there is an indication whether alternate al...

Страница 44: ...It is recommended rst to look at the signals in DUAL mode and to set the sampling rate such that at least one signal period will be displayed Then XY should be selected In XY digital mode both channe...

Страница 45: ...he VOLTS DIV SCALE VAR knob 17 returns to 1 2 5 sequence switching and CH2 to calibrated condition 34 INPUT CH1 BNC connector This is the CH1 signal input connector In Yt mode it is a Y input in XY mo...

Страница 46: ...accuracy of the FFT calculations in the following averageing process Numbers between 2 and 512 can be chosen by aid of the INTENS 2 knob which are displayed by the readout e g avg 256 The accuracy in...