PERMA PURE LLC
Model 3300 User’s Manual
Doc. #532: Revision: 000
Page 9 of 14
Computer Control Terminals Summary
Terminal 2 = System Calibration
Terminal 1 = Cal Gas 1
Terminal 4 = Direct Calibration
Terminal 3 = Cal Gas 2
Terminal 6 = Blowback
Terminal 5 = Cal Gas 3
Terminal 8 = NA
Terminal 7 = Cal Gas 4
Terminal 10 = +5 VDC
Terminal 9 = Cal Gas 5
Terminal 12 = +5 VDC
Terminal 11 = Cal Gas 6
Terminal 14 = GND
Terminal 13 = Cal Gas 7
Terminal 16 = GND
Terminal 15 = Cal Gas 8
Computer Sense CS:
Along with the computer control mode of operation, a computer verification
method is also incorporated into the system. A 40 pin vertical terminal block on the back panel
allows the computer to verify the actions initiated by the computer. This is done through a set of
normally open dry contacts that are associated with each particular function.
Computer Sense Terminals Summary
FUNCTION
TERMINALS
Cal Gas Valve 1
1 & 2
Cal Gas Valve 2
3 & 4
Cal Gas Valve 3
5 & 6
Cal Gas Valve 4
7 & 8
Cal Gas Valve 5
9 & 10
Cal Gas Valve 6
11 & 12
Cal Gas Valve 7
13 & 14
Cal Gas Valve 8
15 & 16
System
Calibration
17
&
18
Direct
Calibration
19
&
20
Blowback*
31
&
32
Local
Mode*
39
&
40
*Note: The sense contacts are normally closed when activated except for blowback