background image

 
 
 

 

7

 

COPYRIGHT 2008, ALL RIGHTS RESERVED, Original Instructions 

 PERMA PURE LLC 

Model 3300 User’s Manual 
Doc. #532: Revision: 000 

Page 7 of 14 

 

Operation 

 

A sample conditioning system supplies a clean, dry sample to the back panel of the Model 3300 
Flow Control Drawer. The stack probe filter and sample pump performance are indicated by the 

Flow Control Drawer Vacuum/Pressure Gauges.  
 

Note: Please refer to the attached Model 3300 Flow Diagram for an example of an integrated 
continuous emissions monitoring system flow schematic using the Model 3300 Flow Control 
Drawer.    

 
Flow from the conditioning system enters the Model 3300 Flow Drawer sample/calibration manifold. 

Here, either sample or calibration gas (for direct calibration) can be directed to the atmospheric 
pressure sample manifold and then on to the individual analyzer gas channels. Individual sample 

flowmeters indicate the flow rate of each gas channel. A bypass flowmeter vents excess sample gas 
not used by the continuous gas channels and also acts as an atmospheric vent.  

 
The Model 3300 Flow Control Drawer has the ability to switch the sample and as many as eight (8) 
calibration gases to the analyzer rack directly or via the total sampling system. This is done 

through two (2) sets of solenoid valves: (1) the calibration gas solenoid valves which are mounted 
on a common manifold at the rear of the drawer, and (2) the system/direct solenoid valves located 

in the front of the drawer. A block & bleed solenoid valve, which prevents the manifolds from being 
pressurized during sampling, is also part of the system/direct solenoid valve set.  

 
The calibration function will only take place when both a calibration gas solenoid valve and either 

the system solenoid valve or direct solenoid valve have been selected. In Local mode, the 
calibration gas solenoid valves are cascaded, that is if there are more than one (1) calibration gas 
solenoid valve switches in the on position, only the highest numerical calibration gas solenoid valve 

will be active. This is not the case with computer control. The programmer can select any or all of 
the calibration gas solenoids. Also, computer control is not allowed in the Local mode.  While in the 

Local/Direct Calibration mode, manual blowback is allowed, but not while in the Local/System 
Calibration mode. While under computer control, blowback is allowed when nothing else is 

activated, but not in any Calibration mode.  
 

The Model 3300 has a provision for system bias checks. System integrity is verified by comparison 
of calibration gas flow through the stack filter assembly to calibration gas flow directly to the gas 
analyzers. This check indicates problems such as system leakage and gas component loss.  

 
A probe filter blowback relay contact is provided on the Flow Control Drawer to facilitate Manual or 

computer initiated blowback. For Blowback there is a 250V, 6Amp dry contact form C relay output 
provided on the back panel for blowback implementation. There are two (2) vertical terminal block 

assemblies on the back panel provided for computer interface. The 16-point terminal block is used 
for computer control and the 40- point terminal block is used for computer verification of flow 

control drawer operation. 
 

Pressure Sensor Board Option 

 
Upstream, (stack filter/and sample line) the pressure drop is actively monitored by a vacuum 

transmitter, the sample gas flow is verified by a sample manifold pressure transmitter and 
presence of adequate calibration gas is verified by a pressure transmitter on the calibration gas 

manifold. 
 

Содержание PERMA PURE 3300

Страница 1: ...C 732 244 0010 800 337 3762 fax 732 244 8140 www permapure com info permapure com PERMA PURE LLC 8 Executive Drive P O Box 2105 Toms River NJ 08754 Baldwin Series Model 3300 3300P Flow Control Drawer...

Страница 2: ...ard Option 7 Local Mode Operation 8 Computer Control Mode Operation 8 Computer Control Terminals Summary 9 Computer Sense Terminals Summary 9 COMPUTER CONTROL TRUTH TABLE 10 Pressure Sensor Model 3300...

Страница 3: ...ly a carrier must be notified within 24 hours of the delivery to make a hidden damage claim Items in the carton include 1 Model 3300 Flow Control Drawer 1 User s Manual If any of the above parts are m...

Страница 4: ...to the stack filter head and through the entire sample system to the calibration gas introduced directly to the continuous gas analyzer common sample manifold This check indicates problems such as sy...

Страница 5: ...ack Specify 12 VDC line voltage or dry contact Electrical requirements 90 240 VAC 50 60 Hz 100 watts Connections refer to Appendix A3 for back panel drawing 1 All Sample Gas Connections are via 1 4 st...

Страница 6: ...ed prior to operation Electrical current may be present behind covers or doors even if tools are not necessary to access these components NEVER attempt service on this equipment without first disconne...

Страница 7: ...gas solenoid valve and either the system solenoid valve or direct solenoid valve have been selected In Local mode the calibration gas solenoid valves are cascaded that is if there are more than one 1...

Страница 8: ...calibration toggle the Direct System switch to the left Then toggle any of the Calibration Gas switches to the right This will allow the selected calibration gas to flow to the analyzer suite directl...

Страница 9: ...ong with the computer control mode of operation a computer verification method is also incorporated into the system A 40 pin vertical terminal block on the back panel allows the computer to verify the...

Страница 10: ...n 7 Direct Calibration 8 Blowback D Disabled Pressure Sensor Model 3300P The pressure sensor option provides Vacuum pressure sensor 0 15 psig Sample pressure sensor 0 30 psig Calibration gas pressure...

Страница 11: ...tle Cal gas solenoid not operating Flow drawer regulator set 10psig Cal gas Flowmeter closed Block bleed solenoid constantly vents System cal gas solenoid valve defective Direct cal gas solenoid valve...

Страница 12: ...001 2 way solenoid valve 12 VDC 2VSB 002 3 way solenoid valve 12 VDC 2FTS 021 Cal Gas connector 3DCB 006 Circuit board switch board FCD 4D 3300P 2FTS 021 Cal Gas connector 3DCB 006 Circuit board swit...

Страница 13: ...PYRIGHT 2008 ALL RIGHTS RESERVED Original Instructions PERMA PURE LLC Model 3300 User s Manual Doc 532 Revision 000 Page 13 of 14 Appendix A Flow Control Drawer Drawings A1 Panel drawing A2 Flow schem...

Страница 14: ...e suitable or desirable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysi...

Страница 15: ...Form C contact 1 Amp 2 Outputs are terminated on the Flow Control Drawer back panel on the Pressure Transducer Terminal Strip Connections N O Common N C Vacuum Pin 11 Pin 12 Pin 13 Sample Pin 14 Pin 1...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Отзывы: