Troubleshooting and Maintenance Guide 

 

Symptom 

 

Diagnostic

 

Fix 

No LED display 

Not plugged in 
 

Power switch is off 

 

Fuse blown 
 

Power supply defect 

     12 VDC red-black 
 

Check jumper JP1 

 
Ribbon cable loose 

Defective control board 

Plug in instrument 
 

Turn on power 

 

Replace fuse 
 

Replace power supply 

 
 

Jumper 1-2; 3-4; 5-6 

 
Tighten ribbon cables 

Replace control board 

Manual Control 
(Local) 

Switches have no effect 

 
Inadequate sample flow 

 
No system cal gas flow 

 
 

 
 
 

 
 

 
 

No direct cal gas flow 

 
 
Remote/Local switch in remote 

 
Low sample pressure 

 
No cal gas from bottle 

Cal gas solenoid not operating 
 

Flow drawer regulator set <10psig 
Cal gas Flowmeter closed 
Block/bleed solenoid constantly 

vents 
System cal gas solenoid valve 

defective 
 

Direct cal gas solenoid valve 
defective 

 
 
Select local 

 
Adjust sample pressure 

Cylinder to 15 psig 
Check cabling 

Replace cal gas solenoid 
Replace control board 

Set regulator to 10 psig 
Adjust Flowmeter 
Replace solenoid SV9 

 
Replace solenoid SV10 

 
 

Replace solenoid SV11 
Check cables and connector 

boards for good connections 

Computer Control 
(Remote) 

Computer control has 
no effect 
 

 
 

No computer sense 

 
 

Remote/Local switch in local 
Computer control wiring to 
terminal block is wrong 

 
 

Ribbon cable is loose 
Defective control board 

Computer sense wiring to terminal  
     block is wrong 

 
Ribbon cable is loose 

 
 

Select remote 
 
 

 
 

Check cables. 
Replace control board 

See computer control wiring 
section in manual 

 
Check cables 

 

Содержание PERMA PURE 3300

Страница 1: ...C 732 244 0010 800 337 3762 fax 732 244 8140 www permapure com info permapure com PERMA PURE LLC 8 Executive Drive P O Box 2105 Toms River NJ 08754 Baldwin Series Model 3300 3300P Flow Control Drawer...

Страница 2: ...ard Option 7 Local Mode Operation 8 Computer Control Mode Operation 8 Computer Control Terminals Summary 9 Computer Sense Terminals Summary 9 COMPUTER CONTROL TRUTH TABLE 10 Pressure Sensor Model 3300...

Страница 3: ...ly a carrier must be notified within 24 hours of the delivery to make a hidden damage claim Items in the carton include 1 Model 3300 Flow Control Drawer 1 User s Manual If any of the above parts are m...

Страница 4: ...to the stack filter head and through the entire sample system to the calibration gas introduced directly to the continuous gas analyzer common sample manifold This check indicates problems such as sy...

Страница 5: ...ack Specify 12 VDC line voltage or dry contact Electrical requirements 90 240 VAC 50 60 Hz 100 watts Connections refer to Appendix A3 for back panel drawing 1 All Sample Gas Connections are via 1 4 st...

Страница 6: ...ed prior to operation Electrical current may be present behind covers or doors even if tools are not necessary to access these components NEVER attempt service on this equipment without first disconne...

Страница 7: ...gas solenoid valve and either the system solenoid valve or direct solenoid valve have been selected In Local mode the calibration gas solenoid valves are cascaded that is if there are more than one 1...

Страница 8: ...calibration toggle the Direct System switch to the left Then toggle any of the Calibration Gas switches to the right This will allow the selected calibration gas to flow to the analyzer suite directl...

Страница 9: ...ong with the computer control mode of operation a computer verification method is also incorporated into the system A 40 pin vertical terminal block on the back panel allows the computer to verify the...

Страница 10: ...n 7 Direct Calibration 8 Blowback D Disabled Pressure Sensor Model 3300P The pressure sensor option provides Vacuum pressure sensor 0 15 psig Sample pressure sensor 0 30 psig Calibration gas pressure...

Страница 11: ...tle Cal gas solenoid not operating Flow drawer regulator set 10psig Cal gas Flowmeter closed Block bleed solenoid constantly vents System cal gas solenoid valve defective Direct cal gas solenoid valve...

Страница 12: ...001 2 way solenoid valve 12 VDC 2VSB 002 3 way solenoid valve 12 VDC 2FTS 021 Cal Gas connector 3DCB 006 Circuit board switch board FCD 4D 3300P 2FTS 021 Cal Gas connector 3DCB 006 Circuit board swit...

Страница 13: ...PYRIGHT 2008 ALL RIGHTS RESERVED Original Instructions PERMA PURE LLC Model 3300 User s Manual Doc 532 Revision 000 Page 13 of 14 Appendix A Flow Control Drawer Drawings A1 Panel drawing A2 Flow schem...

Страница 14: ...e suitable or desirable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysi...

Страница 15: ...Form C contact 1 Amp 2 Outputs are terminated on the Flow Control Drawer back panel on the Pressure Transducer Terminal Strip Connections N O Common N C Vacuum Pin 11 Pin 12 Pin 13 Sample Pin 14 Pin 1...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Отзывы: