B: Specifications
016-70000-000-02-201603
29
NIRQuest Spectrometer
–
Slit Size vs. Resolution
and Throughput
The following chart illustrates the effect that varying slit sizes have on NIRQuest512 Spectrometer
resolution and throughput:
NIRQuest Gratings
The following tables show the NIRQuest gratings available for preconfigured (standard) setups and for all
options. Additional grating options, adjustments to starting and ending wavelengths and similar
customization may be available. Please contact an Applications Scientist for details.
Spectrometer
Standard
Grating
Groove
Density
(lines/mm)
Spectral
Range
Blaze
Wavelength
Best Efficiency
(>30%)
NIRQuest512-1.7
NIR3
150
~800 nm
1100 nm
900-1700 nm
NIRQuest512-1.9
NIR3
150
~800 nm
1100 nm
1100-1900 nm
NIRQuest512-2.2
NIR2
100
1150 nm
1600 nm
900-2200 nm
NIRQuest512-2.5
NIR1
75
1425 nm
1700 nm
1075-2500 nm
NIRQuest256-2.1
NIR2
100
1150 nm
1600 nm
900-2050 nm
NIRQuest256-2.5
NIR1
75
1425 nm
1700 nm
1075-2500 nm
0
10
20
30
40
50
60
0
2
4
6
8
10
12
14
16
0
50
100
150
200
R
e
la
ti
v
e
I
n
te
n
s
it
y
R
e
s
o
lu
ti
o
n
(
F
WH
M)
Slit Size (microns)
Содержание Ocean Optics NIRQuest512-1.7
Страница 8: ...About This Manual vi 016 70000 000 02 201603...
Страница 14: ...1 Introduction 6 016 70000 000 02 201603...
Страница 22: ...2 Installing the NIRQuest Spectrometer 14 016 70000 000 02 201603...
Страница 48: ...C NIRQuest512 2 5 and NIRQuest256 2 5 Sensitivity 40 016 70000 000 02 201603...