B: Specifications
24
016-70000-000-02-201603
NIRQuest512-1.7
NIRQuest512-1.9
NIRQuest512-2.2
NIRQuest512-2.5
Longpass filter
3
:
OF1-RG830
longpass NIR
filter (optional)
OF1-CGA1000
longpass NIR filter
(standard)
OF1-RG830 longpass NIR filter (optional)
2
nd
Order filter
3
:
N/A
Standard
Collimating and
focusing mirrors:
Gold-coated for enhanced NIR reflectivity
Fiber optic
connector:
SMA 905 to 0.22 numerical aperture single-strand optical fiber
SPECTROSCOPIC
Wavelength range:
900-1700 nm
w/Grating NIR3
1100-1900nm
w/Grating NIR3
900-2200nm
w/Grating NIR2
900-2500nm w/Grating
NIR1
Optical resolution
(FWHM)
4
:
~3.1 nm w/25 µm slit
~5 nm w/25 µm slit
~6.3 nm w/25 µm slit
Signal-to-noise ratio
at full signal
5
:
>15000:1 @ 100 ms integration
10000:1 @ 100 ms integration
>13000:1 @ 1000 ms integration
A/D resolution:
16-bit
Dark noise:
6 RMS counts @ 100 ms
16 RMS counts@
10ms
12 RMS counts
@
1000 ms
12 RMS counts @ 250 ms
24 RMS counts@
30ms
Dynamic range:
150 x 10
6
(system);
15K:1 for a single
acquisition
7.5M (system); 10K:1 for a single acquisition
100K (system); 7.5K:1 for
a single acquisition
Integration time
6
:
1 ms
–
120 s
1 ms
–
1 s
1 ms
–
200 ms
Corrected linearity:
>99.8%
>99.6%
ELECTRONICS
Power consumption
DC input jack +5V, 3 A maximum
Data transfer speed:
Full scan to memory every 5 ms with USB 2.0 port
Inputs/ Outputs:
External trigger input + single strobe output