B: Specifications
28
016-70000-000-02-201603
NIRQuest256-2.1
NIRQuest256-2.5
TEMPERATURE & THERMOELECTRIC COOLING
Temperature limits
(environmental):
10-35 ºC (0-90% non-condensing)
TEC range:
30 -- 50ºC below ambient
TEC stability:
+/-0.5 ºC of set temperature in <1 minute; typical long-term sta/-0.1 ºC
COMPUTER
Computer interfaces:
USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud (custom
configuration)
Peripheral interfaces:
I2C inter-integrated circuit
1
“Useable range” is defined in the context of the NIRQuest
model’s detector response and its typical grating response. For example,
the 512-element detector has response at 850 nm, but grating response begins at 900 nm. The G9206 256-element detector
response is sensitive to TEC temperature, and has response only to 2050 nm when the TEC is set to -20 ºC. The G9208 256-
element and 512-element detector has response to 2550 nm, but the grating efficiency drops off at 2500 nm.
2
See
NIRQuest Gratings
for more information.
3
Other filter options are available for order-sorting in the NIRQuest256-2.1, NIRQuest256-2.5 and NIRQuest 512-2.5. NIRQuest256-
2.5 and NIRQuest 512-2.5 ship with a 2
nd
-order filter. See an Applications Scientist for details.
4
Optical resolution (FWHM) depends on grating and slit selection.
5
SNR will decrease at longer integration times.
6
Maximum integration times are defined as the longest amount of time one can integrate the spectrometer before the dark level
rises to half of full scale.