Block 5 33
Block
Channel 1...8
Number
Range:
-100.0…100.0
Designation in %
Standard: 0.0
In case of the adjustment 0.0 % the output of the actual
present input value is set to 0 during zero adjust (taring).
In case of values other than 0.0 % the output is modified by the
adjusted value.
First point of two-point adjustment
3 5 x I I
Zero point
adjustment value *)
Level
Range:
-200.0…200.0
Designation in % (100 % is equal to an input signal movement
of +20 mV)
Standard: automatic entry of the adjustments
This parameter is optional to enter or to control of the 35x11,
35x21 and 35x22 influenced value.
It shows the zero offset of the input signal
3 5 x I 2
Manual
zero adjustment
Range:
10.0…110.0
Designation in %
Standard: 100.0
Entry of the load in % of the nominal value.
Recommended values are between 70 and 110 %.
The smallest possible value also depends on parameters 35x11
and has 10 % larger than his.
Second point of the two-point adjustment
3 5 x 2 I
Setting for
defined load*)
Range:
100.0…30000.0
Amplification factor in V/V
Standard: 666.7
This parameter is optional for the input of the gain or to display the
calculated value by 35x21
3 5 x 2 2
Manual amplifica
-
tion adjustment
(Gain)
The parameters for external triggering of the adjustment (45x12, 45x13, 45x22 and 45x23), for
coupling of amplifiers (45x11; 45x21 and 45x24) and the wrap angle correction (45x30 to 45x39 and
35x31 to 35x39) are described in the Reference Manual.
*) In Level 3 only the zero point adjustment value, resp.
the adjustment value for the defined load is adjusted. The
actual zero and gain adjustment is activated in the quick
functions (4.2.3.).