Get a sample measurement
Use this option to get sample measurements with one of the installed applications.
1.
From the main menu, select the measurement application, then push
Start
. Application
information shows on the display.
2.
Read the related "Application Note" from the USB applications key for more instructions.
3.
If necessary, select an icon for more information or to change some data.
4.
Fill a beaker with the recommended sample amount that shows on the display. If necessary, add
sufficient deionized water to make sure the sensor is correctly installed in the sample.
5.
Carefully put a magnetic stir bar into the beaker. Make sure there is no liquid spill.
6.
Attach the beaker to the sensor holder.
7.
Make sure that the icon at the bottom of the display
is highlighted. Do the instructions that
show on the display adjacent to this icon. Refer to
on page 13 to make sure
that the tubes and sensor are correctly aligned.
8.
Push
Start
to start the measurement. Measurement data shows on the display.
9.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
10.
If the options are available during the measurement procedure, push
Skip
to ignore the current
step or push
Stop
to abort the measurement.
11.
When the measurement is complete, push the arrow keys to see the different measurement
views.
12.
Push
Next
for the options that follow:
Option
Description
Replicate
sample
Use this option to start the same titration on the same sample. This is used to study the
repeatability by successively analyzing several parts of the same sample. At the end of
each measurement, a window shows the average value, the standard deviation and the
relative standard deviation.
New sample
Use this option to start the same titration on a new sample. No standard deviation or
relative standard deviation measurements will be done.
13.
Push
Exit
to go back to the main menu.
Manage the data log
To select data to view, delete or export, specify data filters
1.
From the main menu, select
Data log
.
2.
Select an option, then push
Select
.
Option
Description
View data log
Views measurement data. Select individual lines of data to view more content.
Export data log
Exports measurement data from the system to an external device. Preview data selection
before it is exported. Make sure that an external device is connected to the instrument
(e.g., a USB key, external hard drive, etc.).
Delete data log
Removes measurement data from the system. Previews data selection before it is
removed.
3.
Specify the data extraction parameters. Push the left and right arrow keys to make a selection.
Push the up and down arrow keys to select an option.
Option
Description
Result type
Sets the type of result available.
18
English
Содержание TitraLab AT1102
Страница 2: ...English 3 Fran ais 24 Espa ol 47 Portugu s 70 93 112 135 2...
Страница 31: ...Fran ais 31...
Страница 44: ...44 Fran ais...
Страница 54: ...54 Espa ol...
Страница 67: ...Espa ol 67...
Страница 77: ...Portugu s 77...
Страница 90: ...90 Portugu s...
Страница 95: ...4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 2 0 1 2 1 1 AT1102 1 0 AT1112 1 1 AT1122 1 2 AT1222 2 2 95...
Страница 96: ...2 USB USB USB USB 2 1 24V 4 7 USB 2 1 5 8 3 2 6 3 1 3 3x 5 2 1 4 1 96...
Страница 98: ...GFCI GFI 1 2 2 96 3 180 5 7 98...
Страница 99: ...1 95 99...
Страница 100: ...100...
Страница 101: ...102 1 2 5 5 1 2 3 101...
Страница 102: ...102...
Страница 103: ...MSDS SDS USB 103...
Страница 104: ...6 6 1 4 2 5 3 2 104...
Страница 105: ...MSDS SDS 1 Settings 2 Select 105 105 3 Back 1 Settings 2 3 4 5 105...
Страница 109: ...2 Toggle Syringe replacement 98 1 Pump activation Start 2 Toggle 109...
Страница 110: ...1 Select Reset 110...
Страница 111: ...Edit PtPt ORP 111...
Страница 114: ...1 2 3 4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 2 1 2 1 1 AT1102 1 0 AT1112 1 1 114...
Страница 115: ...1 AT1122 1 2 AT1222 2 2 2 USB USB USB USB 2 1 24 V 4 7 USB 2 1 5 8 3 2 6 HACH Japan 115...
Страница 116: ...3 1 3 3 5 2 1 4 1 1 116...
Страница 117: ...4 1 10 50mL 10 150 L 7 5 2 2 8 1 114 3 2 9 USB 4 10 10 6 5 3 11 GL45 2 GL25 1 6 4 12 3 1 114 1 1 5 1 2 6 3 4 117...
Страница 118: ...15 cm 6 AC GFCI GFI 1 2 2 115 3 2 1 2 180 2 2 5 7 118...
Страница 119: ...1 114 119...
Страница 120: ...120...
Страница 121: ...122 1 2 5 121...
Страница 122: ...5 1 2 3 122...
Страница 123: ...MSDS SDS USB 123...
Страница 124: ...124...
Страница 125: ...6 6 1 4 2 5 3 2 125...
Страница 126: ...MSDS SDS 1 Settings 2 Select 126 5 PC 126 3 Back 1 Settings 2 Applications Edit 3 Edit 126...
Страница 128: ...MSDS SDS 1 1 Start 2 USB 3 4 5 6 7 122 8 Start 9 10 Skip Stop 11 12 Next 13 Exit 128...
Страница 130: ...4 5 6 Continue 7 Yes No 8 Exit 127 Maintenance 1 Syringe activation 130...
Страница 131: ...2 2 Toggle 2 Syringe replacement 118 1 Pump activation 2 Toggle 131...
Страница 132: ...1 Select Reset 132...
Страница 134: ...Burette fails to fill Burette failed to read delivered volume 134...
Страница 137: ...1 2 3 4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 2 0 1 2 1 1 AT1102 1 0 AT1112 1 1 AT1122 1 2 AT1222 2 2 137...
Страница 138: ...2 USB USB USB USB 2 1 24V 4 7 USB 2 1 5 8 3 2 6 138...
Страница 139: ...3 1 3 3x 5 2 1 4 1 1 139...
Страница 141: ...AC GFCI GFI 1 2 2 138 3 180 5 7 141...
Страница 142: ...1 137 142...
Страница 143: ...143...
Страница 144: ...145 1 2 5 5 1 2 3 144...
Страница 145: ...145...
Страница 146: ...MSDS SDS USB 146...
Страница 147: ...6 6 1 4 2 5 3 2 147...
Страница 148: ...MSDS SDS 1 2 148 PC 148 3 1 2 3 148...
Страница 149: ...4 5 USB 5 1 2 22 154 1 2 USB USB USB 3 4 5 USB 1 2 3 4 2 5 6 MSDS SDS 149...
Страница 150: ...1 2 USB 3 4 5 6 7 145 8 9 10 11 12 13 1 2 USB 3 150...
Страница 151: ...1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 Purge 149 151...
Страница 152: ...1 2 2 2 141 1 2 152...
Страница 153: ...153...
Страница 154: ...1 PtPt ORP 154...
Страница 155: ...155...
Страница 156: ......
Страница 157: ......