4.
Push the arrow keys to scroll through the application parameters. The
Edit
key is only available
when a parameter can be changed.
Note: More parameters can be changed with the instrument in expert mode.
5.
Push
Edit
to change the parameter. Enter the new details or select from a list.
Install the applications
Use the supplied USB key to install the applications. The instrument can install a maximum of five
applications for each syringe installed. For two syringes, the installed applications shown on the top
line of the display refer to syringe one and the installed applications shown on the second line refer to
syringe two. If any errors occur during installation, refer to
1.
Push
Home
to go to the main menu.
2.
Connect the USB key to the USB port on the side of the instrument. The applications on the USB
key show on the display.
3.
Push the arrow keys to highlight and select an application to install. Push the left or right arrow
key to select it. Do this step again to select additional applications to install.
4.
Push
Import
to install the selected applications.
5.
Push
OK
to complete the installation. The installed applications show on the main menu.
Note: To install more applications, push Home to go to the main menu, then remove the USB key and
reconnect it.
Prepare the instrument for measurement
1.
From the main menu, select
Purge
, then push
Start
. All attached devices are listed.
2.
Select
All elements
to purge all the attached devices, or select one device to purge. Push
Select
. Air is removed from the device and filled with liquid from the bottle.
3.
Push
OK
when the operation has completed.
4.
Make sure that there are no air bubbles in the device. Do step
again if there are any air
bubbles.
5.
Select the next device to purge if individual devices are being selected.
6.
Push
Exit
when all the tubes are filled with reagent and the device has no air bubbles.
Note: If a few small air bubbles can be seen on the inner wall and/or piston of the syringe, they can be left
without effecting system performance.
Standard operations
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Personal injury hazard. Never use the instrument without the syringe protection cover in place.
C A U T I O N
Chemical exposure hazard. Never remove the stir bar from the beaker before the end of a titration.
English
17
Содержание TitraLab AT1102
Страница 2: ...English 3 Fran ais 24 Espa ol 47 Portugu s 70 93 112 135 2...
Страница 31: ...Fran ais 31...
Страница 44: ...44 Fran ais...
Страница 54: ...54 Espa ol...
Страница 67: ...Espa ol 67...
Страница 77: ...Portugu s 77...
Страница 90: ...90 Portugu s...
Страница 95: ...4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 2 0 1 2 1 1 AT1102 1 0 AT1112 1 1 AT1122 1 2 AT1222 2 2 95...
Страница 96: ...2 USB USB USB USB 2 1 24V 4 7 USB 2 1 5 8 3 2 6 3 1 3 3x 5 2 1 4 1 96...
Страница 98: ...GFCI GFI 1 2 2 96 3 180 5 7 98...
Страница 99: ...1 95 99...
Страница 100: ...100...
Страница 101: ...102 1 2 5 5 1 2 3 101...
Страница 102: ...102...
Страница 103: ...MSDS SDS USB 103...
Страница 104: ...6 6 1 4 2 5 3 2 104...
Страница 105: ...MSDS SDS 1 Settings 2 Select 105 105 3 Back 1 Settings 2 3 4 5 105...
Страница 109: ...2 Toggle Syringe replacement 98 1 Pump activation Start 2 Toggle 109...
Страница 110: ...1 Select Reset 110...
Страница 111: ...Edit PtPt ORP 111...
Страница 114: ...1 2 3 4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 2 1 2 1 1 AT1102 1 0 AT1112 1 1 114...
Страница 115: ...1 AT1122 1 2 AT1222 2 2 2 USB USB USB USB 2 1 24 V 4 7 USB 2 1 5 8 3 2 6 HACH Japan 115...
Страница 116: ...3 1 3 3 5 2 1 4 1 1 116...
Страница 117: ...4 1 10 50mL 10 150 L 7 5 2 2 8 1 114 3 2 9 USB 4 10 10 6 5 3 11 GL45 2 GL25 1 6 4 12 3 1 114 1 1 5 1 2 6 3 4 117...
Страница 118: ...15 cm 6 AC GFCI GFI 1 2 2 115 3 2 1 2 180 2 2 5 7 118...
Страница 119: ...1 114 119...
Страница 120: ...120...
Страница 121: ...122 1 2 5 121...
Страница 122: ...5 1 2 3 122...
Страница 123: ...MSDS SDS USB 123...
Страница 124: ...124...
Страница 125: ...6 6 1 4 2 5 3 2 125...
Страница 126: ...MSDS SDS 1 Settings 2 Select 126 5 PC 126 3 Back 1 Settings 2 Applications Edit 3 Edit 126...
Страница 128: ...MSDS SDS 1 1 Start 2 USB 3 4 5 6 7 122 8 Start 9 10 Skip Stop 11 12 Next 13 Exit 128...
Страница 130: ...4 5 6 Continue 7 Yes No 8 Exit 127 Maintenance 1 Syringe activation 130...
Страница 131: ...2 2 Toggle 2 Syringe replacement 118 1 Pump activation 2 Toggle 131...
Страница 132: ...1 Select Reset 132...
Страница 134: ...Burette fails to fill Burette failed to read delivered volume 134...
Страница 137: ...1 2 3 4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 2 0 1 2 1 1 AT1102 1 0 AT1112 1 1 AT1122 1 2 AT1222 2 2 137...
Страница 138: ...2 USB USB USB USB 2 1 24V 4 7 USB 2 1 5 8 3 2 6 138...
Страница 139: ...3 1 3 3x 5 2 1 4 1 1 139...
Страница 141: ...AC GFCI GFI 1 2 2 138 3 180 5 7 141...
Страница 142: ...1 137 142...
Страница 143: ...143...
Страница 144: ...145 1 2 5 5 1 2 3 144...
Страница 145: ...145...
Страница 146: ...MSDS SDS USB 146...
Страница 147: ...6 6 1 4 2 5 3 2 147...
Страница 148: ...MSDS SDS 1 2 148 PC 148 3 1 2 3 148...
Страница 149: ...4 5 USB 5 1 2 22 154 1 2 USB USB USB 3 4 5 USB 1 2 3 4 2 5 6 MSDS SDS 149...
Страница 150: ...1 2 USB 3 4 5 6 7 145 8 9 10 11 12 13 1 2 USB 3 150...
Страница 151: ...1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 Purge 149 151...
Страница 152: ...1 2 2 2 141 1 2 152...
Страница 153: ...153...
Страница 154: ...1 PtPt ORP 154...
Страница 155: ...155...
Страница 156: ......
Страница 157: ......