Troubleshooting

Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.

Table 2  Calibration warnings and errors

Error/Warning

Solution

Asymmetry >
58 mV

Slope out of range (accepted values ± 58 mV).
Calibrate again.
Connect a new probe.

Buffer or electrode
in poor conditions.

Calibrate again.
Examine the probe: Clean the probe (refer to

 Clean the

probe

 on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; Connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Make sure that the buffer used
matches the buffer specified in configuration; make sure of
the temperature specification in configuration; use new
buffer solution.

Sens. (a) < 70%

BUFFER NOT
RECOGNIZED

UNSTABLE
READING
Time t> 100 s

Calibrate again.
Examine the probe: Clean the probe (refer to

 Clean the

probe

 on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Make sure that the membrane and the diaphragm are
properly immersed in the sample.

Table 2  Calibration warnings and errors (continued)

Error/Warning

Solution

Electrode in poor
conditions.

Examine the probe: Clean the probe (refer to

 Clean the

probe

 on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.

Check the
electrode

SAME BUFFERS

Calibrate again.
Examine the probe: Clean the probe (refer to

 Clean the

probe

 on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Use new buffer solution.

Table 3  Measurement warnings and errors

Error/Warning

Solution

pH 12.78 19°C

Calibrate again

pH out of range

Examine the probe: Clean the probe (refer to

 Clean the

probe

 on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.

Temp out of range
°C

Examine the temperature sensor.
Connect a different probe to verify if problem is with probe
or meter.

Time > 60 s

Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to

 Clean the

probe

 on page 13 for more information); make sure that

there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.

Time > 150 s

Time > 300 s

16

   

English

Содержание sension ph31

Страница 1: ...DOC022 97 90249 sensION pH31 03 2013 Edition 3 User Manual Manuel d utilisation Manual del usuario Manual do Usu rio...

Страница 2: ...etails Display Liquid crystal backlit 128 x 64 dots Keyboard PET with protective treatment Certification CE General information Revised editions are found on the manufacturer s website Safety informat...

Страница 3: ...ife equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipm...

Страница 4: ...1 2 3 4 English 5...

Страница 5: ...adapter plug for the power outlet from the adapter kit 2 Connect the universal power adapter to the meter Figure 2 3 Connect the universal power adapter to an AC receptacle Figure 3 4 Turn the meter...

Страница 6: ...a value 2 MEASUREMENT key confirm the selected option 6 ON OFF turn on or turn off the meter 3 UP key scroll to other options change a value 7 LEFT key change the measuring unit enter numbers and let...

Страница 7: ...t a language from a list 2 Confirm with The measurement screen shows DATA OUTPUT 3 Select Deactivated if no printer or PC is connected and confirm Refer to Select the data output on page 11 for more i...

Страница 8: ...alibration The probe calibration data is replaced at 25 C 77 F Calibration procedure This procedure is for general use with liquid calibration solutions Refer to the documents that are included with e...

Страница 9: ...r and and to select MEASURE Confirm 2 Use the to change the following settings Confirm every entry Option Description Resolution Select the resolution 1 0 1 0 01 default or 0 001 Measure Stability sel...

Страница 10: ...the display Adjust the display contrast 1 From the main menu use the or to select SYSTEM Confirm 2 Use the or to select Display contrast and confirm 3 Use the and to adjust the contrast of the display...

Страница 11: ...rm 4 Use the or to select Option Description Measurement data Measurement data stores automatically each time a sample is measured Electrode report Electrode report stores automatically the electrode...

Страница 12: ...the print report 4 When a computer is connected and LabCom Easy refer to Select the data output on page 11 for more information is selected use the or to select Option Description Users The user name...

Страница 13: ...Replace the magnetic stirrer If the magnetic stirrer does not start follow the numbered steps to replace the magnetic stirrer 14 English...

Страница 14: ...1 2 English 15...

Страница 15: ...Examine the probe Clean the probe refer to Clean the probe on page 13 for more information make sure that there are no air bubbles in the membrane Shake the probe like a thermometer connect a differe...

Страница 16: ...64 97 pH buffer solution 10 01 250 mL LZW9471 99 pH buffer solution 4 01 1000 mL LZW9466 99 pH buffer solution 7 00 1000 mL LZW9467 97 Replacement parts and accessories continued Description Item no p...

Страница 17: ...chnical buffer solutions DIN 19267 Refer to Table 4 pH and ORP mV values of specific buffer sets at varying temperatures Table 4 pH ORP mV and temperature values Temperature pH mV C F 0 32 2 01 4 01 7...

Страница 18: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compo...

Отзывы: