Change the stirring settings

The magnetic stirrer can be turned on and the stirring speed can be
changed in the Stirring menu.

1.

From the main menu use the   or   to select SYSTEM. Confirm.

2.

Use the   or   to select Stirring and confirm.

3.

To turn on/off the stirrer push  .

4.

When the stirrer is turned on, use the   or   to change the stirring
speed in %.

Note: Use the   or   to change the stirring speed during calibration and
during a measurement.

Change the temperature units

The temperature units can be changed to Celsius or Fahrenheit.

1.

From the main menu use the   or   to select SYSTEM. Confirm.

2.

Use the   or   to select Temperature units and confirm.

3.

Use the   or   to select between Celsius or Fahrenheit and confirm.

Data logger

Display data

The Display data log contains Measurement data, Electrode report and
Calibration data. The stored data can be send to a printer or to a PC.
When the data log becomes full (400 data points), the oldest data point
is deleted when a new data point is added.

1.

From the main menu use the   or   to select SYSTEM. Confirm.

2.

Use the   or   to select DATA LOGGER and confirm.

3.

Use the   or   to select Display data and confirm.

4.

Use the   or   to select

Option

Description

Measurement data

Measurement data—stores automatically each time a
sample is measured

Electrode report

Electrode report—stores automatically the electrode
history and measurement conditions

Calibration data

Calibration data—stores automatically the current
calibration

Delete data

The entire measurement data or electrode report log can be deleted to
remove data that has already been sent to a printer or PC.

1.

From the main menu use the   or   to select SYSTEM. Confirm.

2.

Use the   or   to select DATA LOGGER and confirm.

3.

Use the   or   to select Erase and confirm.

4.

Use the   or   to select Measurement data or Electrode report and
confirm. Confirm again to delete the data.
The entire log is deleted at once.

Send data to a printer or to a computer

N O T I C E

 

The data output (printer or PC) needs to be selected first, so that the Print menu
is available (refer to 

Select the data output

 on page 11).

Note: Refer to 

Report output

 on page 13 to select the report output type.

1.

From the main menu use the   or   to select SYSTEM. Confirm.

2.

Use the   or   to select DATA LOGGER and confirm.

3.

Use the   or   to select Print and confirm. Select one of the
following options and confirm with   to print the data: Measurement
data, Electrode data, Calibration data, Calibration report or
Instrument condit.

12

   

English

Содержание sension ph31

Страница 1: ...DOC022 97 90249 sensION pH31 03 2013 Edition 3 User Manual Manuel d utilisation Manual del usuario Manual do Usu rio...

Страница 2: ...etails Display Liquid crystal backlit 128 x 64 dots Keyboard PET with protective treatment Certification CE General information Revised editions are found on the manufacturer s website Safety informat...

Страница 3: ...ife equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipm...

Страница 4: ...1 2 3 4 English 5...

Страница 5: ...adapter plug for the power outlet from the adapter kit 2 Connect the universal power adapter to the meter Figure 2 3 Connect the universal power adapter to an AC receptacle Figure 3 4 Turn the meter...

Страница 6: ...a value 2 MEASUREMENT key confirm the selected option 6 ON OFF turn on or turn off the meter 3 UP key scroll to other options change a value 7 LEFT key change the measuring unit enter numbers and let...

Страница 7: ...t a language from a list 2 Confirm with The measurement screen shows DATA OUTPUT 3 Select Deactivated if no printer or PC is connected and confirm Refer to Select the data output on page 11 for more i...

Страница 8: ...alibration The probe calibration data is replaced at 25 C 77 F Calibration procedure This procedure is for general use with liquid calibration solutions Refer to the documents that are included with e...

Страница 9: ...r and and to select MEASURE Confirm 2 Use the to change the following settings Confirm every entry Option Description Resolution Select the resolution 1 0 1 0 01 default or 0 001 Measure Stability sel...

Страница 10: ...the display Adjust the display contrast 1 From the main menu use the or to select SYSTEM Confirm 2 Use the or to select Display contrast and confirm 3 Use the and to adjust the contrast of the display...

Страница 11: ...rm 4 Use the or to select Option Description Measurement data Measurement data stores automatically each time a sample is measured Electrode report Electrode report stores automatically the electrode...

Страница 12: ...the print report 4 When a computer is connected and LabCom Easy refer to Select the data output on page 11 for more information is selected use the or to select Option Description Users The user name...

Страница 13: ...Replace the magnetic stirrer If the magnetic stirrer does not start follow the numbered steps to replace the magnetic stirrer 14 English...

Страница 14: ...1 2 English 15...

Страница 15: ...Examine the probe Clean the probe refer to Clean the probe on page 13 for more information make sure that there are no air bubbles in the membrane Shake the probe like a thermometer connect a differe...

Страница 16: ...64 97 pH buffer solution 10 01 250 mL LZW9471 99 pH buffer solution 4 01 1000 mL LZW9466 99 pH buffer solution 7 00 1000 mL LZW9467 97 Replacement parts and accessories continued Description Item no p...

Страница 17: ...chnical buffer solutions DIN 19267 Refer to Table 4 pH and ORP mV values of specific buffer sets at varying temperatures Table 4 pH ORP mV and temperature values Temperature pH mV C F 0 32 2 01 4 01 7...

Страница 18: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compo...

Отзывы: