C A U T I O N

 

Indicates a potentially hazardous situation that may result in minor or moderate
injury.

N O T I C E

 

Indicates a situation which, if not avoided, may cause damage to the instrument.
Information that requires special emphasis.

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol, if noted
on the instrument, will be included with a danger or caution statement in
the manual.

This symbol, if noted on the instrument, references the instruction
manual for operation and/or safety information.

Electrical equipment marked with this symbol may not be disposed of
in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU Directive
2002/98/EC), European electrical equipment users must now return
old or end-of-life equipment to the Producer for disposal at no charge
to the user.

Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment, producer-supplied
electrical accessories, and all auxillary items for proper disposal.

Product overview

The sensION

+ meters are used with probes to measure various

parameters in water.
The sensION

+ PH31 meter measures pH, ORP (mV) or temperature.

Measurement data can be stored and transferred to a printer or PC.

Product components

Refer to 

Figure 1

 to make sure that all components have been received.

If any items are missing or damaged, contact the manufacturer or a
sales representative immediately.

Figure 1  Meter components

1   

Electrolyte for the probe

5   

Power supply

2   

Buffer solutions (pH 4.01, pH
7.00 and pH 10.01)

6   

Rod with o-ring

3   

Calibration beakers (with magnetic
bar inside)

7   

Probe (included with kits only)

4   

Probe holder

8   

Meter

Installation

Assemble the probe holder

Follow the numbered steps to assemble the probe holder and to connect
the magnetic stirrer.

4

   

English

Содержание sension ph31

Страница 1: ...DOC022 97 90249 sensION pH31 03 2013 Edition 3 User Manual Manuel d utilisation Manual del usuario Manual do Usu rio...

Страница 2: ...etails Display Liquid crystal backlit 128 x 64 dots Keyboard PET with protective treatment Certification CE General information Revised editions are found on the manufacturer s website Safety informat...

Страница 3: ...ife equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipm...

Страница 4: ...1 2 3 4 English 5...

Страница 5: ...adapter plug for the power outlet from the adapter kit 2 Connect the universal power adapter to the meter Figure 2 3 Connect the universal power adapter to an AC receptacle Figure 3 4 Turn the meter...

Страница 6: ...a value 2 MEASUREMENT key confirm the selected option 6 ON OFF turn on or turn off the meter 3 UP key scroll to other options change a value 7 LEFT key change the measuring unit enter numbers and let...

Страница 7: ...t a language from a list 2 Confirm with The measurement screen shows DATA OUTPUT 3 Select Deactivated if no printer or PC is connected and confirm Refer to Select the data output on page 11 for more i...

Страница 8: ...alibration The probe calibration data is replaced at 25 C 77 F Calibration procedure This procedure is for general use with liquid calibration solutions Refer to the documents that are included with e...

Страница 9: ...r and and to select MEASURE Confirm 2 Use the to change the following settings Confirm every entry Option Description Resolution Select the resolution 1 0 1 0 01 default or 0 001 Measure Stability sel...

Страница 10: ...the display Adjust the display contrast 1 From the main menu use the or to select SYSTEM Confirm 2 Use the or to select Display contrast and confirm 3 Use the and to adjust the contrast of the display...

Страница 11: ...rm 4 Use the or to select Option Description Measurement data Measurement data stores automatically each time a sample is measured Electrode report Electrode report stores automatically the electrode...

Страница 12: ...the print report 4 When a computer is connected and LabCom Easy refer to Select the data output on page 11 for more information is selected use the or to select Option Description Users The user name...

Страница 13: ...Replace the magnetic stirrer If the magnetic stirrer does not start follow the numbered steps to replace the magnetic stirrer 14 English...

Страница 14: ...1 2 English 15...

Страница 15: ...Examine the probe Clean the probe refer to Clean the probe on page 13 for more information make sure that there are no air bubbles in the membrane Shake the probe like a thermometer connect a differe...

Страница 16: ...64 97 pH buffer solution 10 01 250 mL LZW9471 99 pH buffer solution 4 01 1000 mL LZW9466 99 pH buffer solution 7 00 1000 mL LZW9467 97 Replacement parts and accessories continued Description Item no p...

Страница 17: ...chnical buffer solutions DIN 19267 Refer to Table 4 pH and ORP mV values of specific buffer sets at varying temperatures Table 4 pH ORP mV and temperature values Temperature pH mV C F 0 32 2 01 4 01 7...