35
SECTION 4
CONDUCTIVITY OPERATION MODE
4.1
Conductivity Setup Menu Options
Table 5 describes the options available in the Conductivity
Setup menu.
Temperature Correction Value — Allows selection of a linear
or non-linear temperature correction function. The non-linear
coefficient has been determined from measurements using
aqueous NaCl solutions; for most freshwater samples, this is the
best setting. If the linear function is chosen, the measured
conductivity values are automatically temperature-corrected
based on the specified temperature coefficient and the selected
Table 5 Setup Options
Setup
Number
Setup Description
Option Description
Default Setting
1
Time
00:00 to 23:59
00:00
2
Date
01/01 to 12/31
01/01
3
Year
2000–2099
2000
4
Temperature units
°C, °F
°C
5
Display lock
On:
Lock
icon
Off:
Lock
icon and
(Off)
icon
Off
6
Temperature
Correction Factor
(Thermometer icon)
Non-linear NaCl or
Non-linear NaCl
7
TDS Correction Factor
Non-linear NaCl or numeric value for converting
µ
S/cm to TDS
NaCl in units
field
8
Auto-print interval
Off, 10 sec., 30 sec., 1 min., 5 min., 20 min.,
1 hr., 2 hrs., and 6 hrs.
Off
9
Reference
Temperature Selection
(Thermometer icon)
20° C or 25° C
25° C
10
Temperature
Correction
(Thermometer icon)
On, Off
(Off)
icon for off
On. If “raw”
conductivity is
desired, such
as with a soil
cup, turn off
temperature
correction.
11
NaCl or KCl
Calibration Standards
toggles NaCl, KCl
NaCl
Numeric value
[
]
%
°C
------------------------------------------------