11
SPECIFICATIONS,
continued
Resolution:
Accuracy:
Conductivity: ±0.5% of range
TDS: ±0.5% of full scale
Salinity: ±0.1 ppt (-2 to 35°C)
Temperature: ±0.3 °C from 0–70 °C;
±1.0 °C from 70–110 °C
Conversion Factor for TDS: automatic or user adjustable
Temperature Compensation:
Manual (user selected coefficient, % per °C) or
Automatic (non-linear based on NaCl solutions)
Dissolved Oxygen Mode
Range: 0–20 mg/L (ppm), 0–200% sat.
Accuracy: ±1% full scale
Temperature: 0–50 °C
Resolution:
Instrument drift: < 1%/day
TDS:
0.00–199.9 mg/L
0.1 mg/L
200–1999 mg/L
1 mg/L
2.00–19.99 g/L
0.01 g/L
20.0–50 g/L
0.1 g/L
Salinity:
0.1 ppt (‰)
Temperature
±0.1 °C
Oxygen Concentration: 0.01 or 0.1 ppm (mg/L)
% Saturation:
0.1%
Temperature:
0.1 °C