39
SECTION 4,
continued
4.3
Conductivity Calibration
Calibrate the meter before use. There are two ways to calibrate
the meter:
1. Use NaCl or KCl standards of known electrolytic
conductivity. See Sections 4.3.1 and 4.3.2 for instructions on
these calibration method.
Setup 7
TDS Correction
Factor
From Conductivity Reading mode, press
SETUP
.
Press the
UP ARROW
until the setup number is
7
.
Press
ENTER
.
Press the
UP
or
DOWN ARROW
keys to switch between linear and non-linear
correction functions.
To choose a non-linear conversion, scroll until a flashing
NaCl
appears, then
press
ENTER
.
To select a linear conversion, scroll until the conversion coefficient appears. Use
the numeric keypad to set the value of the coefficient, then press
ENTER
. If an
number entry error occurs, start over by pressing
CE
.
When the desired option is selected, press
EXIT
to return to the reading mode.
See Section 4.1 for more information about this setup.
Setup 8
Auto Print
Intervals
From Conductivity Reading mode, press
SETUP
.
Press the
UP ARROW
until the setup number is
8
.
Press
ENTER
.
Change the print interval by pressing the
UP ARROW
and
DOWN ARROW
keys.
Press
ENTER
to accept the print interval.
Setup 9
Reference
Temperature
From Conductivity Reading mode, press
SETUP
.
Press the
UP ARROW
until the setup number is
9
.
Press
ENTER
to toggle between 20° C and 25° C. The default is 25° C.
Press
EXIT
to return to the reading mode.
See Section 4.1 for more information about this setup.
Setup 10
Temperature
Correction
Off and On
From Conductivity Reading mode, press
SETUP
.
Press the
UP ARROW
until the setup number is
10
.
Press
ENTER
to toggle between (off) and on.
Press
EXIT
to return to the reading mode.
See Section 4.1 for more information about this setup.
Setup 11
KCl or NaCl
Calibration
Standard
Selection
From Conductivity Reading mode, press
SETUP
.
Press the
UP ARROW
until setup number is 11.
Press
ENTER
to toggle between NaCl and KCl standards.
Note: Switching conductivity standards changes the temperature correction
coefficients for calibration only. The choice of calibration standard has no effect on
the temperature correction coefficient applied when reading samples.
Table 7 How to Change Conductivity Menu Options (Continued)
Setup
How to Get There