63
SECTION 5,
continued
9.
Press the
READ/ENTER
key. The
stabilizing icon will
appear while the meter
completes the
calibration.
10.
When the
calibration is complete,
the meter will return to
the reading mode. Press
the
EXIT
key during the
calibration sequence to
back out of the
calibration routine, one
screen at a time, without
completing a calibration.
11.
Add the weight
assembly to the probe if
required (3- or 15-m
cable versions only).
12.
If necessary, enter
the sample salinity value
into DO Setup 7. See
Section 4.6 on page 44.
13.
Insert the probe
into the sample. The
probe must be deep
enough to cover the
thermistor (metallic
button) located on the
side of the probe.
14.
Agitate the probe
in the sample to dislodge
air bubbles from the
sensing area of the
probe tip.
15.
Stir the sample
vigorously with the
probe or use a stir stand
and stir bar. When
measuring deep bodies
of water, create
sufficient flow across the
probe tip by pulling on
the cable to move the
probe up and down.
16.
When the reading
on the meter stabilizes,
record or store the value
in the meter memory.
enter
exit
7 8 9
4 5 6
1 2 3
CAL
Stabilizing...
C
DO
enter
exit
7 8 9
4 5 6
1 2 3
DO
enter
exit
7 8 9
4 5 6
1 2 3
CORRECTION
Sal
setup
DO
store
8