A list of possible warnings is shown in
Table 3 Warning list for dissolved oxygen sensors
Warning
Description
Resolution
DO TOO HIGH
The measured value is > 40 ppm
Make sure that the DO level in the
process water is within the
operating limits of the sensor.
Calibrate or replace the sensor.
DO TOO LOW
The measured value is < 0 ppb
Calibrate or replace the sensor.
TEMP TOO HIGH
The measured temperature is >
50 °C
Reduce sample temperature.
TEMP TOO LOW
The measured temperature is <
0 °C
Increase sample temperature.
CURRENT TOO HIGH
The measured current > 200 µA
Make sure that the DO level in the
process water is within the
operating limits of the sensor.
Calibrate or replace the sensor.
CURRENT TOO LOW
The measured current < -0.5 mA
Calibrate or replace the sensor.
CAL OVERDUE
The Cal Reminder time has expired
Calibrate the sensor.
REPLACE SENSOR
The sensor has been in operation >
365 days
Replace the sensor cartridge and
calibrate the sensor. If the
calibration result is pass, reset the
membrane days in the DIAG/TEST
menu.
NOT CALIBRATED
The sensor has not been calibrated
Calibrate the sensor.
CAL IN PROGRESS
A calibration was started but not
completed
Return to calibration.
9.3.2 Error list
Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the
menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in
.
Table 4 Error list for dissolved oxygen sensors
Error
Description
Resolution
ADC FAILURE
The analog to digital conversion
failed
Power off and power on the
controller. Call technical support.
TEMP SENSOR MISSING
The temperature sensor is missing
or disconnected
Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.
Section 10 Replacement parts and accessories
Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.
Note: Product and article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
20
English
Содержание Polymetron 9582sc
Страница 1: ...DOC023 53 93067 Hach 9582sc Dissolved Oxygen 02 2020 Edition 5 User Manual...
Страница 2: ......
Страница 4: ...Table of Contents 2...
Страница 11: ...English 9...
Страница 27: ......