Figure 4  Replace the membrane

8.2 Electrode rejuvenation

After some months of operation (3 to 12 depending on sample oxygen concentration, plant shutdown
frequency, etc.), a dark coating of silver bromide (AgBr) may cover part of the silver anode. This
coating does not affect the measurement unless more than 90% of the surface is contaminated.
When changing the electrolyte and membrane, visually check the silver anode. If more than 2/3 of
the surface is coated with silver bromide then an electrode rejuvenation is required. To do this, very
softly polish the areas coated with the deposit using a soft abrasive (N° 400 to 600). After polishing,
rinse with demineralized water and wipe dry with a soft cloth. Some of the dark coloring may be left
to improve sensor stabilization time.
After cleaning, replace the sensor in the sample and leave for 30 minutes for the measurement to
stabilize. Once stable, the sensor must be calibrated.

16

   

English

Содержание Polymetron 9582sc

Страница 1: ...DOC023 53 93067 Hach 9582sc Dissolved Oxygen 02 2020 Edition 5 User Manual...

Страница 2: ......

Страница 3: ...iguration 10 7 2 Configure the sensor 10 7 3 Calibrate the sensor 11 7 3 1 About sensor calibration 11 7 3 2 Temperature calibration 11 7 3 3 Zero calibration 11 7 3 4 Calibration in air 12 7 3 5 Cali...

Страница 4: ...Table of Contents 2...

Страница 5: ...pressure Atmospheric pressure Relative humidity 0 90 Certifications EN 61326 1 2006 EN 61010 1 2010 Sensor module Specification Details Measuring range 0 2 ppm Sensitivity 0 5 ppb Repeatability 0 5 p...

Страница 6: ...if not avoided could result in death or serious injury C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury N O T I C E Indicates a situation which if...

Страница 7: ...t transfer liquid 3 3 Product components Make sure that all components have been received If any items are missing or damaged contact the manufacturer or a sales representative immediately Components...

Страница 8: ...il there is resistance Do not force the electrode 4 Install and tighten the holding nut by hand 5 Install the gasket and filling screw to prevent electrolyte leakage or external contamination from the...

Страница 9: ...connector not provided 3 Assembled sensor 7 M4 screws not provided 4 Gasket 8 1 4 NPT sample output connector not provided 4 4 Install the controller Refer to the controller documentation to see mount...

Страница 10: ...c components can be damaged by static electricity resulting in degraded performance or eventual failure Refer to the illustrated steps that follow and Table 1 Be sure to connect all sensor ground shie...

Страница 11: ...English 9...

Страница 12: ...on and push enter To enter numbers characters or punctuation push and hold the up or down arrow keys Push the right arrow key to advance to the next space Option Description EDIT NAME Changes the name...

Страница 13: ...tilled water 2 Push the menu key and select SENSOR SETUP CALIBRATE 3 If the pass code is enabled in the security menu for the controller enter the pass code 4 Select ZERO CAL and push enter 5 Select t...

Страница 14: ...rops of water Place the sensor vertically in the calibration cap with the membrane downwards Tighten the calibration cap and push enter 8 Wait for the value to stabilize and push enter 9 Review the ca...

Страница 15: ...libration options on page 13 10 On the NEW SENSOR screen select whether the sensor is new Option Description YES The sensor was not calibrated previously with this controller The days of operation and...

Страница 16: ...STANDBY Section 8 Maintenance W A R N I N G Multiple hazards Only qualified personnel must conduct the tasks described in this section of the document 8 1 Membrane replacement C A U T I O N Chemical e...

Страница 17: ...embrane 1 Unscrew the filling screw and remove along with the gasket 2 Unscrew the holding nut 3 Carefully remove the electrode from the sensor body 4 Empty any electrolyte left in the sensor body 5 U...

Страница 18: ...e and membrane visually check the silver anode If more than 2 3 of the surface is coated with silver bromide then an electrode rejuvenation is required To do this very softly polish the areas coated w...

Страница 19: ...nd install finger tight The gold surface is scratched or damaged Change the electrode Sludge or particles on the cathode Clean the cathode with a soft and absorbent tissue Rinse the membrane Cable or...

Страница 20: ...membrane filling screw and change the electrolyte and the membrane Electrolyte leaking Check the screw in parts membrane filling screw and change the electrolyte and the membrane Interferences mainly...

Страница 21: ...number for the sensor module SENSOR INFORMATION Shows the name and serial number that was entered by the user CAL DAYS Shows the number of days since the last calibration CAL HISTORY Shows a list of t...

Страница 22: ...he sensor CAL IN PROGRESS A calibration was started but not completed Return to calibration 9 3 2 Error list Errors may occur for various reasons An error icon consists of an exclamation point within...

Страница 23: ...uaranty of the described product properties 1 IDENTIFICATION OF THE SUBSTANCE Product code 09181 A 3600 Product name Electrolyte 2 COMPOSITION INFORMATION ON INGREDIENTS Water 90 CAS No 108 18 9 Potas...

Страница 24: ...cautions protective equipment and emergency procedures Use personal protective equipment Environmental precautions Do not flush into surface water or sanitary sewer system Methods and material for con...

Страница 25: ...f Specific effects in experiment on an animal No data is available on the product itself 12 ECOLOGICAL INFORMATION Toxicity No data is available on the product itself Do not flush into surface water o...

Страница 26: ...REGULATORY INFORMATION EU regulatory information The product does not need to be labelled in accordance with EC directives or respective national laws National regulatory information Water contaminati...

Страница 27: ......

Страница 28: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: