Section 5 Startup
Make sure that the flow rate and pressure do not exceed the values in
on page 3.
1.
Open the valve on the sample line to let sample flow through the analyzer.
2.
Turn the knob on the flow meter to set the flow rate.
3.
Examine the plumbing for leaks and stop any leaks if found.
4.
Apply power to the controller.
5.
Make the applicable menu selections when the controller starts.
Section 6 User navigation
Refer to the controller documentation for keypad description and navigation information.
Push the
RIGHT
arrow key on the controller multiple times to show more information on the home
screen and to show a graphical display.
Section 7 Operation
7.1 System configuration
Refer to the controller documentation for system configuration, general controller settings and setup
for outputs and communications.
7.2 Configure the sensor
Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.
1.
Push the
menu
key, select SENSOR SETUP>CONFIGURE.
2.
Use the arrow keys to select an option and push
enter
. To enter numbers, characters or
punctuation, push and hold the
up
or
down
arrow keys. Push the
right
arrow key to advance to
the next space.
Option
Description
EDIT NAME
Changes the name that corresponds to the sensor on the top of the measure screen.
The name is limited to 16 characters in any combination of letters, numbers, spaces or
punctuation.
SENSOR S/N
Allows the user to enter the serial number of the sensor, limited to 16 characters in any
combination of letters, numbers, spaces or punctuation.
MEAS UNITS
Changes the measurement units—Select the unit from the list available.
PRESSURE UNITS
Sets the units for atmospheric pressure—Select the unit from the list available.
TEMP UNITS
Sets the temperature units to °C (default) or °F
FILTER
Sets a time constant to increase signal stability. The time constant calculates the
average value during a specified time—0 (no effect, default) to 60 seconds (average of
signal value for 60 seconds). The filter increases the time for the sensor signal to
respond to actual changes in the process.
LOG SETUP
Sets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10,
15 (default), 30, 60 minutes.
RESET DEFAULTS
Sets the configuration menu to the default settings. All sensor information is lost.