Hach Polymetron 9582sc Скачать руководство пользователя страница 12

Section 5 Startup

Make sure that the flow rate and pressure do not exceed the values in 

Specifications

 on page 3.

1.

Open the valve on the sample line to let sample flow through the analyzer.

2.

Turn the knob on the flow meter to set the flow rate.

3.

Examine the plumbing for leaks and stop any leaks if found.

4.

Apply power to the controller.

5.

Make the applicable menu selections when the controller starts.

Section 6 User navigation

Refer to the controller documentation for keypad description and navigation information.
Push the 

RIGHT

 arrow key on the controller multiple times to show more information on the home

screen and to show a graphical display.

Section 7 Operation

7.1 System configuration

Refer to the controller documentation for system configuration, general controller settings and setup
for outputs and communications.

7.2 Configure the sensor

Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.

1.

Push the 

menu

 key, select SENSOR SETUP>CONFIGURE.

2.

Use the arrow keys to select an option and push 

enter

. To enter numbers, characters or

punctuation, push and hold the 

up

 or 

down

 arrow keys. Push the 

right

 arrow key to advance to

the next space.

Option

Description

EDIT NAME

Changes the name that corresponds to the sensor on the top of the measure screen.
The name is limited to 16 characters in any combination of letters, numbers, spaces or
punctuation.

SENSOR S/N

Allows the user to enter the serial number of the sensor, limited to 16 characters in any
combination of letters, numbers, spaces or punctuation.

MEAS UNITS

Changes the measurement units—Select the unit from the list available.

PRESSURE UNITS

Sets the units for atmospheric pressure—Select the unit from the list available.

TEMP UNITS

Sets the temperature units to °C (default) or °F

FILTER

Sets a time constant to increase signal stability. The time constant calculates the
average value during a specified time—0 (no effect, default) to 60 seconds (average of
signal value for 60 seconds). The filter increases the time for the sensor signal to
respond to actual changes in the process.

LOG SETUP

Sets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10,
15 (default), 30, 60 minutes.

RESET DEFAULTS

Sets the configuration menu to the default settings. All sensor information is lost.

10

   

English

Содержание Polymetron 9582sc

Страница 1: ...DOC023 53 93067 Hach 9582sc Dissolved Oxygen 02 2020 Edition 5 User Manual...

Страница 2: ......

Страница 3: ...iguration 10 7 2 Configure the sensor 10 7 3 Calibrate the sensor 11 7 3 1 About sensor calibration 11 7 3 2 Temperature calibration 11 7 3 3 Zero calibration 11 7 3 4 Calibration in air 12 7 3 5 Cali...

Страница 4: ...Table of Contents 2...

Страница 5: ...pressure Atmospheric pressure Relative humidity 0 90 Certifications EN 61326 1 2006 EN 61010 1 2010 Sensor module Specification Details Measuring range 0 2 ppm Sensitivity 0 5 ppb Repeatability 0 5 p...

Страница 6: ...if not avoided could result in death or serious injury C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury N O T I C E Indicates a situation which if...

Страница 7: ...t transfer liquid 3 3 Product components Make sure that all components have been received If any items are missing or damaged contact the manufacturer or a sales representative immediately Components...

Страница 8: ...il there is resistance Do not force the electrode 4 Install and tighten the holding nut by hand 5 Install the gasket and filling screw to prevent electrolyte leakage or external contamination from the...

Страница 9: ...connector not provided 3 Assembled sensor 7 M4 screws not provided 4 Gasket 8 1 4 NPT sample output connector not provided 4 4 Install the controller Refer to the controller documentation to see mount...

Страница 10: ...c components can be damaged by static electricity resulting in degraded performance or eventual failure Refer to the illustrated steps that follow and Table 1 Be sure to connect all sensor ground shie...

Страница 11: ...English 9...

Страница 12: ...on and push enter To enter numbers characters or punctuation push and hold the up or down arrow keys Push the right arrow key to advance to the next space Option Description EDIT NAME Changes the name...

Страница 13: ...tilled water 2 Push the menu key and select SENSOR SETUP CALIBRATE 3 If the pass code is enabled in the security menu for the controller enter the pass code 4 Select ZERO CAL and push enter 5 Select t...

Страница 14: ...rops of water Place the sensor vertically in the calibration cap with the membrane downwards Tighten the calibration cap and push enter 8 Wait for the value to stabilize and push enter 9 Review the ca...

Страница 15: ...libration options on page 13 10 On the NEW SENSOR screen select whether the sensor is new Option Description YES The sensor was not calibrated previously with this controller The days of operation and...

Страница 16: ...STANDBY Section 8 Maintenance W A R N I N G Multiple hazards Only qualified personnel must conduct the tasks described in this section of the document 8 1 Membrane replacement C A U T I O N Chemical e...

Страница 17: ...embrane 1 Unscrew the filling screw and remove along with the gasket 2 Unscrew the holding nut 3 Carefully remove the electrode from the sensor body 4 Empty any electrolyte left in the sensor body 5 U...

Страница 18: ...e and membrane visually check the silver anode If more than 2 3 of the surface is coated with silver bromide then an electrode rejuvenation is required To do this very softly polish the areas coated w...

Страница 19: ...nd install finger tight The gold surface is scratched or damaged Change the electrode Sludge or particles on the cathode Clean the cathode with a soft and absorbent tissue Rinse the membrane Cable or...

Страница 20: ...membrane filling screw and change the electrolyte and the membrane Electrolyte leaking Check the screw in parts membrane filling screw and change the electrolyte and the membrane Interferences mainly...

Страница 21: ...number for the sensor module SENSOR INFORMATION Shows the name and serial number that was entered by the user CAL DAYS Shows the number of days since the last calibration CAL HISTORY Shows a list of t...

Страница 22: ...he sensor CAL IN PROGRESS A calibration was started but not completed Return to calibration 9 3 2 Error list Errors may occur for various reasons An error icon consists of an exclamation point within...

Страница 23: ...uaranty of the described product properties 1 IDENTIFICATION OF THE SUBSTANCE Product code 09181 A 3600 Product name Electrolyte 2 COMPOSITION INFORMATION ON INGREDIENTS Water 90 CAS No 108 18 9 Potas...

Страница 24: ...cautions protective equipment and emergency procedures Use personal protective equipment Environmental precautions Do not flush into surface water or sanitary sewer system Methods and material for con...

Страница 25: ...f Specific effects in experiment on an animal No data is available on the product itself 12 ECOLOGICAL INFORMATION Toxicity No data is available on the product itself Do not flush into surface water o...

Страница 26: ...REGULATORY INFORMATION EU regulatory information The product does not need to be labelled in accordance with EC directives or respective national laws National regulatory information Water contaminati...

Страница 27: ......

Страница 28: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: