Problem

Probable cause

Resolution

Important instability
in measuring mode.

There is water or humidity in the sensor
cable connector.

Dry the sensor cable connector inside and
out, and install finger tight.

Incorrect connection.

Check the sensor connection to the
controller module.

There are bubbles close to the cathode.

Refill with electrolyte and check there are
no bubbles at the bottom of the sensor
body.

The sensor has been violently shaken.

Check the fixation and stability of the
sensor.

Electromagnetic interferences close to the
sensor or controller cable.

Find a better place for the cable and check
the EMC levels.

Temporary interference with other gases.

Mainly with H

2

S.

Flow rate too low (4 mL/h minimum).

Increase sample flow.

Sludge from a heterogeneous sample has
damaged the membrane.

Install a deflector or change the sensor
location.

Pressure variation in the line.

Make sure the sensor is used at
atmospheric pressure.

Lack of accuracy.

The permeability of the membrane has
changed (dirt deposits).

Calibrate the analyzer and check if the
concentration is back to normal.

Electrolyte pollution.

Check the screw-in parts (membrane, filling
screw) and change the electrolyte and the
membrane.

Electrolyte leaking.

Check the screw-in parts (membrane, filling
screw) and change the electrolyte and the
membrane.

Interferences, mainly with H

2

S.

If the level of H

2

S (or other pollutant) is

stable, take its concentration into account to
determine the dissolved O

2

 concentration.

Error during calibration or incorrect
calibration.

Calibrate again to check the parameters. If
the error persists check the calibration
current (too high, too low or unstable) and
the concentration in air. Refer to the
problems described above.

There are bubbles close to the cathode.

Refill with electrolyte and check there are
no bubbles at the bottom of the sensor
body.

Flow rate too low (4 mL/h minimum).

Increase sample flow.

The sample temperature or pressure is out
of specification.

Change the sensor location or modify the
sample so that it is within specification.

Sludge or particles on the cathode.

Clean the cathode with a soft and
absorbent tissue. Change the membrane.

The sensor current
is null during
measurement.

There is no electrolyte in the sensor
(leakage).

Check the screw-in parts (membrane, filling
screw) and change the electrolyte.

Sensor cable is disconnected or poorly
connected.

Check cable connections on the module
and the sensor connector in the controller.

18

   

English

Содержание Polymetron 9582sc

Страница 1: ...DOC023 53 93067 Hach 9582sc Dissolved Oxygen 02 2020 Edition 5 User Manual...

Страница 2: ......

Страница 3: ...iguration 10 7 2 Configure the sensor 10 7 3 Calibrate the sensor 11 7 3 1 About sensor calibration 11 7 3 2 Temperature calibration 11 7 3 3 Zero calibration 11 7 3 4 Calibration in air 12 7 3 5 Cali...

Страница 4: ...Table of Contents 2...

Страница 5: ...pressure Atmospheric pressure Relative humidity 0 90 Certifications EN 61326 1 2006 EN 61010 1 2010 Sensor module Specification Details Measuring range 0 2 ppm Sensitivity 0 5 ppb Repeatability 0 5 p...

Страница 6: ...if not avoided could result in death or serious injury C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury N O T I C E Indicates a situation which if...

Страница 7: ...t transfer liquid 3 3 Product components Make sure that all components have been received If any items are missing or damaged contact the manufacturer or a sales representative immediately Components...

Страница 8: ...il there is resistance Do not force the electrode 4 Install and tighten the holding nut by hand 5 Install the gasket and filling screw to prevent electrolyte leakage or external contamination from the...

Страница 9: ...connector not provided 3 Assembled sensor 7 M4 screws not provided 4 Gasket 8 1 4 NPT sample output connector not provided 4 4 Install the controller Refer to the controller documentation to see mount...

Страница 10: ...c components can be damaged by static electricity resulting in degraded performance or eventual failure Refer to the illustrated steps that follow and Table 1 Be sure to connect all sensor ground shie...

Страница 11: ...English 9...

Страница 12: ...on and push enter To enter numbers characters or punctuation push and hold the up or down arrow keys Push the right arrow key to advance to the next space Option Description EDIT NAME Changes the name...

Страница 13: ...tilled water 2 Push the menu key and select SENSOR SETUP CALIBRATE 3 If the pass code is enabled in the security menu for the controller enter the pass code 4 Select ZERO CAL and push enter 5 Select t...

Страница 14: ...rops of water Place the sensor vertically in the calibration cap with the membrane downwards Tighten the calibration cap and push enter 8 Wait for the value to stabilize and push enter 9 Review the ca...

Страница 15: ...libration options on page 13 10 On the NEW SENSOR screen select whether the sensor is new Option Description YES The sensor was not calibrated previously with this controller The days of operation and...

Страница 16: ...STANDBY Section 8 Maintenance W A R N I N G Multiple hazards Only qualified personnel must conduct the tasks described in this section of the document 8 1 Membrane replacement C A U T I O N Chemical e...

Страница 17: ...embrane 1 Unscrew the filling screw and remove along with the gasket 2 Unscrew the holding nut 3 Carefully remove the electrode from the sensor body 4 Empty any electrolyte left in the sensor body 5 U...

Страница 18: ...e and membrane visually check the silver anode If more than 2 3 of the surface is coated with silver bromide then an electrode rejuvenation is required To do this very softly polish the areas coated w...

Страница 19: ...nd install finger tight The gold surface is scratched or damaged Change the electrode Sludge or particles on the cathode Clean the cathode with a soft and absorbent tissue Rinse the membrane Cable or...

Страница 20: ...membrane filling screw and change the electrolyte and the membrane Electrolyte leaking Check the screw in parts membrane filling screw and change the electrolyte and the membrane Interferences mainly...

Страница 21: ...number for the sensor module SENSOR INFORMATION Shows the name and serial number that was entered by the user CAL DAYS Shows the number of days since the last calibration CAL HISTORY Shows a list of t...

Страница 22: ...he sensor CAL IN PROGRESS A calibration was started but not completed Return to calibration 9 3 2 Error list Errors may occur for various reasons An error icon consists of an exclamation point within...

Страница 23: ...uaranty of the described product properties 1 IDENTIFICATION OF THE SUBSTANCE Product code 09181 A 3600 Product name Electrolyte 2 COMPOSITION INFORMATION ON INGREDIENTS Water 90 CAS No 108 18 9 Potas...

Страница 24: ...cautions protective equipment and emergency procedures Use personal protective equipment Environmental precautions Do not flush into surface water or sanitary sewer system Methods and material for con...

Страница 25: ...f Specific effects in experiment on an animal No data is available on the product itself 12 ECOLOGICAL INFORMATION Toxicity No data is available on the product itself Do not flush into surface water o...

Страница 26: ...REGULATORY INFORMATION EU regulatory information The product does not need to be labelled in accordance with EC directives or respective national laws National regulatory information Water contaminati...

Страница 27: ......

Страница 28: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: