6.3.2 Calibration in line
The in line calibration procedure can be used to calibrate the sensor directly in line, against a liquid
sample with a known dissolved oxygen concentration. To perform this type of calibration, the
instrument must be set for calibration
In line
. Switch on the instrument, if necessary, and wait a
minute or so for the displayed measurement to settle. Expose the sensor to a liquid sample with a
known gas concentration.
1.
Press the
CAL
button. Remember, this button may have been locked out to prevent an accidental
Locking out the instrument's CAL button
2.
A brief clearing
[- - -]
message appears.
3.
Press
CAL
again within a 3 second period.
4.
A measurement will flash on the LCD, showing the oxygen concentration of the calibration
sample.
5.
Assuming you know the gas content to be a certain value, you can adjust the displayed value with
the
Up/Down Arrow
keys.
6.
When the reading is adjusted to the known concentration, press
CAL
again.
7.
The instrument displays
CAL
and returns to the measurement mode.
6.3.3 Calibration in air (oxygen sensors only)
The oxygen sensor can be accurately calibrated in air. To perform this type of calibration, the
instrument must be set for calibration
In air
.
In order to calibrate the sensor in air, you will need to extract it from its mounting or flow chamber,
and wipe dry the sensor protection grille (if applicable).
Calibration is best achieved using the storage cap that protected the sensor during shipment. Put
several drops of tap water in the cap, shake out the excess, and then attach it to the sensor by
means of its collar. It is best to leave the cap slightly loose, to avoid compressing the air inside. The
cap and sensor should be about the same temperature.
Switch on the instrument, if necessary, and wait a minute or so for the displayed measurement to
settle.
1.
Press the
CAL
button. Remember, this button may have been locked out to prevent an accidental
Locking out the instrument's CAL button
on page 12 for details).
2.
A brief clearing
[- - -]
message appears.
3.
Press
CAL
again within a 3 second period.
4.
The percentage of the measured current to the ideal one is displayed.
5.
When the reading is stable, press
CAL
again.
If the new calibration current is within 25% to 175% of the ideal current, the instrument displays
CAL
and returns to the measurement mode. Calibration is now complete, the sensor is
calibrated, and you can proceed with your measurements. If the new calibration current is not
within this range, the instrument displays
Err
and returns to measurement mode. The system will
not calibrate, and it is likely that a sensor service will be required.
Section 7 Maintenance
7.1 Instrument
If there are problems with the instrument, please contact your local Hach Lange service
representative.
Section 8 Troubleshooting
If your analyzer is behaving strangely (failing to calibrate, giving inappropriate measurement values,
etc.) and you have attempted to rectify the problem by servicing the sensor, but to no avail, you may
14
English
Содержание ORBISPHERE 29122
Страница 2: ...Table of Contents English 3 Deutsch 16 Italiano 29 Fran ais 42 Portugu s 55 Nederlands 68 81 2...
Страница 82: ...29122 EN 61326 1 2 2 1 2 2 2 3 29122 29122 82...
Страница 83: ...2 4 2 5 29122 ATEX 3650EX 83...
Страница 94: ...1 CAL 91 2 3 3 CAL 4 5 CAL 25 175 CAL Err 7 7 1 Hach Lange 8 Troubleshooting WinLog97 94...
Страница 95: ......