4.1.4 Saving values
To save this list of measurements as a text (
.txt
) file, capable of being recalled by the
WinLog97 program or imported as a file into other Windows programs, choose the
Save As
command from the
File
menu. A dialog box appears, with a space to fill in with an eight-letter name.
(The program automatically attaches a
.txt
suffix to these files.) If you have saved previous files, a
grayed-out list of these names appears as well. Typical to Windows programs,
Directories
and
Drives
boxes can be used to locate other places to save. You may also type the drive and directory
yourself when saving the file.
4.1.5 Printing values
To place this list of measurements into a tabular format and send it to the Windows printer, choose
the
command from the
File
menu. The program asks you to enter
Title
and
Author
information. Note that the
Date
is fixed by your operating system. The resulting printed list will
include this information on each page.
4.1.6 Clearing stored values
To clear all the values stored in the instrument via the WinLog97 program, choose the
Clear Data
command from the
Logger
menu. Since this action will clear the storage memory of the instrument, a
warning appears first. Choose
OK
to bring up the next dialog box to confirm the clear action. Choose
Clear
to start the memory clear operation. A message,
Reset should be completed
appears in this
box when the task is finished.
Note: You can accomplish the same thing passively, by simply allowing the analyzer to overwrite a set of stored
values with new ones.
4.2 Monitoring measurements in real-time
You may wish to analyze a particular sampling point via the WinLog97 program's
Monitoring
menu.
To use this Monitoring chart, the instrument must be connected to your PC. Choose
Monitoring
from
the WinLog97 menu to bring up a chart display.
The Monitoring chart shows the gas concentration (in blue), temperature (in red), and pressure (in
green) as the sample is being measured by the instrument. The chart is updated directly from
instrument measurements, at a rate determined by the time scale set in the
TIMEBASE
box at the
lower right corner of the chart.
Click the
TIMEBASE
up/down pointers to change the time scale of the divisions of the chart. Each
division mark along the baseline (1, 2, ...10) can be made to represent from 30 seconds to 2½ hours,
providing from 5 minutes to 25 hours of continuously displayed samples. The chart updating rate is
determined by the time scale selected.
Timebase
Updating rate*
Maximum samples
(10 divisions)
30 Seconds/Division
5 Seconds/Sample
60
1 Minute/Division
5 Seconds/Sample
120
10 Minutes/Division
5 Seconds/Sample
1,200
30 Minutes/Division
9 Seconds/Sample
2,000
1 Hour/Division
18 Seconds/Sample
2,000
2.5 Hours/Division
45 Seconds/Sample
2,000
*This chart's updating rate is independent from the acquisition rate (see
Automatic data acquisition - setting
Click on the
Continuous
box, in the lower right corner, to enable or disable continuous charting.
When this box is checked, the chart scrolls continuously after reaching the
10
division, and the oldest
samples are lost off the left of the chart. When
Continuous
is not checked, the chart stops displaying
new results after reaching the
10
division, and all subsequent measurements are lost.
Click the up/down pointers for each measurement variable (
GAS
,
TEMPERATURE
and
PRESSURE
) at the right of the chart to change the scaling of that value on the chart. The display of
English
9
Содержание ORBISPHERE 29122
Страница 2: ...Table of Contents English 3 Deutsch 16 Italiano 29 Fran ais 42 Portugu s 55 Nederlands 68 81 2...
Страница 82: ...29122 EN 61326 1 2 2 1 2 2 2 3 29122 29122 82...
Страница 83: ...2 4 2 5 29122 ATEX 3650EX 83...
Страница 94: ...1 CAL 91 2 3 3 CAL 4 5 CAL 25 175 CAL Err 7 7 1 Hach Lange 8 Troubleshooting WinLog97 94...
Страница 95: ......