![Hach NA5600 sc Na+ Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/hach/na5600-sc-na/na5600-sc-na_installation-manual_577078027.webp)
b.
Use the supplied tubing cutter to cut a piece of 6 mm OD (smaller) tubing for the sample inlet
line. Make sure that the tube length is sufficiently long to connect the sample inlet to the
sample source. Keep the sample inlet line as short as possible.
c.
Use the supplied tubing cutter to cut a piece of 6 mm OD (smaller) tubing for the sample
bypass line. Make sure that the tube length is sufficiently long to connect the sample bypass
drain to an open chemical drain.
Note: As an alternative, use ¼-in. OD tubing and tubing adapters (6 mm to ¼-in. OD) to plumb the sample
inlet line(s) and sample bypass line(s).
d.
Push the tubes into the sample inlet and sample bypass drain. Push the tubes in 14 mm
(0.55 in.) to make sure that the tubes are pushed to the stop.
e.
Do step 1 again for another channel(s) as necessary.
For analyzers
with
an enclosure, refer to
on page 30 to identify the sample inlet
and sample bypass drain for each channel.
For analyzers
without
an enclosure, refer to
on page 30 to identify the sample
inlet and sample bypass drain for each channel.
2.
To keep the enclosure rating, install the supplied, red plugs in the sample inlets and sample
bypass drains that are not used.
Do not install a red plug in the DIPA exhaust port.
3.
Connect the sample inlet lines to the optional heat exchanger if the temperature difference
between the samples is more than 15 °C (27 °F). Refer to the documentation supplied with the
heat exchanger for instructions.
4.
Install a pressure regulator on each sample inlet line. For analyzers
with
an enclosure, refer to
For analyzers
without
an enclosure, refer to
.
5.
Make sure that the water pressure to the pressure regulator is less than 6 bar (87 psi) or a
blockage at the pressure regulator can occur.
6.
Install a shut-off valve on each sample inlet line before the pressure regulator.
7.
If the sample turbidity is more than 2 NTU or the sample contains iron particulates, oil or grease,
install a 100 um filter on each sample inlet line. Refer to
Replacement parts and accessories
in
the maintenance and troubleshooting manual for ordering information.
8.
Connect each sample line to a sample source.
9.
Turn the shut-off valve(s) to the open position.
10.
Make sure that there are no leaks at the tubing connections. If there is a leak at a fitting, push the
tube farther into the fitting.
English
27
Содержание NA5600 sc Na+
Страница 17: ...English 17 ...
Страница 34: ...Figure 20 DIPA bottle installation Analyzer without enclosure 34 English ...
Страница 52: ...3 5 1 Abnehmen der Stromabdeckplatte Berücksichtigen Sie dabei die folgenden bebilderten Schritte 52 Deutsch ...
Страница 54: ...54 Deutsch ...
Страница 62: ...Abbildung 10 Führen des Kabels durch einen internen Anschlussstecker 62 Deutsch ...
Страница 72: ...Abbildung 20 Installation der DIPA Flasche Analysator ohne Gehäuse 72 Deutsch ...
Страница 92: ...92 Italiano ...
Страница 100: ...Figura 10 Far passare il cavo attraverso il tappo di una porta interna 100 Italiano ...
Страница 110: ...Figura 20 Installazione del flacone di DIPA Analizzatore senza alloggiamento 110 Italiano ...
Страница 128: ...3 5 1 Dépose du couvercle d accès électrique Reportez vous aux étapes illustrées suivantes 128 Français ...
Страница 130: ...130 Français ...
Страница 138: ...Figure 10 Acheminement du câble par un connecteur de port interne 138 Français ...
Страница 148: ...Figure 20 Installation du flacon de DIPA Analyseur sans boîtier 148 Français ...
Страница 167: ...Español 167 ...
Страница 175: ...Figura 10 Pasar el cable a través de un tapón de puerto interno Español 175 ...
Страница 185: ...Figura 20 Instalación de una botella de DIPA en un analizador sin carcasa Español 185 ...
Страница 202: ...3 5 1 Remover a tampa de acesso eléctrico Consulte os passos ilustrados abaixo 202 Português ...
Страница 204: ...204 Português ...
Страница 212: ...Figura 10 Passar o cabo através de uma tampa de porta interna 212 Português ...
Страница 222: ...Figura 20 Instalação do frasco de DIPA Analisador sem estrutura 222 Português ...
Страница 239: ...3 5 1 De afdekkap voor elektronica verwijderen Raadpleeg de volgende afgebeelde stappen Nederlands 239 ...
Страница 241: ...Nederlands 241 ...
Страница 249: ...Afbeelding 10 Steek de kabel door de tule van een interne poort Nederlands 249 ...
Страница 259: ...Afbeelding 20 DIPA fles plaatsen Analyser zonder behuizing Nederlands 259 ...
Страница 277: ...3 5 1 Zdejmowanie pokrywy dostępu elektrycznego Sprawdzić listę kolejnych czynności na rysunku Polski 277 ...
Страница 279: ...Polski 279 ...
Страница 287: ...Rysunek 10 Poprowadzenie przewodu przez zatyczkę portu wewnętrznego Polski 287 ...
Страница 297: ...Rysunek 20 Instalacja butelki z DIPA analizator bez obudowy Polski 297 ...
Страница 314: ...3 5 1 Ta bort el åtkomstskyddet Se de illustrerade stegen som följer 314 Svenska ...
Страница 316: ...316 Svenska ...
Страница 333: ...Figur 20 Installation av DIPA flaska analysator utan kapsling Svenska 333 ...
Страница 351: ...Suomi 351 ...
Страница 368: ...Kuva 20 DIPA pullon asennus analysaattori jossa ei ole koteloa 368 Suomi ...
Страница 385: ...3 5 1 Az elektromos szerelőfedél leszerelése Lásd az alábbi képeken bemutatott lépéseket Magyar 385 ...
Страница 387: ...Magyar 387 ...
Страница 395: ...10 ábra Vezesse át a kábelt az egyik belső csatlakozó dugóján Magyar 395 ...
Страница 405: ...20 ábra DIPA palack beszerelése ház nélküli analizátor Magyar 405 ...
Страница 423: ...3 5 1 Снимите крышку доступа к электрооборудованию Следуйте инструкциям на представленных ниже рисунках Русский 423 ...
Страница 425: ...Русский 425 ...
Страница 433: ...Рисунок 10 Пропустите кабель через заглушку внутреннего порта Русский 433 ...
Страница 443: ...Рисунок 20 Установка бутылки DIPA Анализатор без корпуса Русский 443 ...
Страница 461: ...Türkçe 461 ...
Страница 478: ...Şekil 20 DIPA şişesinin takılması Muhafazasız analiz cihazı 478 Türkçe ...
Страница 482: ......
Страница 483: ......