Option
Description
SET FUNCTION
ALARM
(default)—Sets the relay to on when the measurement value is more than the
high alarm value or less than the low alarm value.
FEEDER CONTROL
—Sets the relay
to on if a measurement value is more (or less) than the setpoint value.
EVENT
CONTROL
—The relay toggles if a process value reaches an upper or lower limit.
SCHEDULER
—Sets the relay to on at selected times regardless of the measurement
value.
WARNING
—Sets the relay to on when there is a warning or error condition.
PROCESS EVENT
—Sets the relay to on when the analyzer does a selected operation.
SET TRANSFER
Sets the relay to active (on) or inactive (off) when an error occurs if the ERROR HOLD
MODE setting is set to TRANSFER OUTPUTS. The default setting is INACTIVE (off).
Refer to
on page 21.
FAIL SAFE
YES
—Sets the normal condition for the relays to be active (on).
NO
—Sets the normal
condition for the relays to be inactive (off).
•
ALARM function
)
Option
Description
LOW ALARM
Sets the value where the relay is set to on in response to decreasing measured value.
For example, if the low alarm is set for 1.0 and the measured value drops to 0.9, the
relay activates.
HIGH ALARM
Sets the value where the relay is set to on in response to increasing measured value.
For example, if the high alarm is set for 1.0 and the measured value increases to 1.1,
the relay activates.
LOW DEADBAND
Sets the range where the relay stays on after the measured value increases above the
low alarm value. For example, if the low alarm is set for 1.0 and the low deadband is set
for 0.5, the relay stays on between 1.0 and 1.5.
HIGH DEADBAND
Sets the range where the relay stays on after the measured value decreases below the
high alarm value. For example, if the high alarm is set for 4.0 and the high deadband is
set for 0.5, the relay stays on between 3.5 and 4.0.
OFF DELAY
Sets a delay time for the relay to be set to off.
ON DELAY
Sets a delay time for the relay to be set to on.
English
15
Содержание NA5600 sc Na+
Страница 1: ...DOC023 97 80591 NA5600 sc Na 02 2020 Edition 3 Operations Fonctionnement Operaciones Opera es...
Страница 2: ...Table of Contents English 3 Fran ais 30 Espa ol 59 Portugu s 88 118 145 172 2...
Страница 119: ...5 2 6 138 1 1 3 2 4 1 2 1 3 3 1 2 1 2 2 119...
Страница 120: ...2 1 6 diag DIAG TEST 2 122 7 cal Calibrate 3 8 back 4 home 9 menu 5 enter 3 2 3 1 1 diag DIAGNOSTICS 4 120...
Страница 132: ...6 1 4 7 2 5 8 3 6 X 9 Y 132...
Страница 134: ...8 1 Y 3 5 OnMax 2 4 X 6 OffMax 9 OnMin OffMin 1 3 X 5 OnMin 2 4 OffMin 6 Y 134...
Страница 146: ...1 1 3 2 4 1 2 1 3 3 1 2 2 1 6 Diag 2 148 7 Cal 3 8 Back 4 Home 9 Menu 5 Enter 1 2 146...
Страница 159: ...6 1 4 7 2 5 8 3 6 x 9 y 159...
Страница 161: ...8 1 y 3 5 OnMax 2 4 x 6 OffMax 9 OnMin OffMin 1 3 x 5 OnMin 2 4 OffMin 6 y 161...
Страница 198: ......
Страница 199: ......