1.
Push
menu
, then select STOP ANALYZER.
Note: If START ANALYZER shows, the analyzer is already in standby mode.
2.
Push
cal
, then select FLOW RATE CAL.
3.
Wait for the calibration to complete (approximately 5 minutes).
4.
Push
enter
to go to the measurement screen.
5.
Push
menu
, then select START ANALYZER.
5.7 Calibrate the 4-20 mA analog outputs
If an analog output in the analyzer is connected to an external devices, calibrate the analog output as
necessary. The analog outputs are factory-calibrated. The adjustment range for analog output
calibration is ± 2 mA.
Note: If an analog output is configured to be 0–20 mA, 4 mA and 20 mA are calibrated.
1.
Push
menu
, then select SETUP SYSTEM > CONFIGURE ANALYZER > SETUP OUTPUTS >
OUTPUT CALIBRATION > [select an output].
2.
Select an option.
Option
Description
CAL 4mA
With a calibrated digital multimeter, measure the actual value supplied at the analog output.
Adjust the value shown until the signal at the analog output is 4.00 mA.
CAL 20mA
With a calibrated digital multimeter, measure the actual value supplied at the analog output.
Adjust the value shown until the signal at the analog output is 20.00 mA.
Section 6 Operation
6.1 Show the details of the current and the last measurement
Push
menu
and select VIEW DATA > MEASUREMENT DATA. Refer to
.
Table 4 Measurement data descriptions
Item
Description
LAST MEAS TIME
The time the last measurement was completed.
LAST MEAS CHANNEL
The last channel measured.
NEXT MEAS TIME
The time the next measurement will be completed.
NEXT MEAS CHANNEL
The next channel to be measured.
SAMPLE TEMPERATURE
The temperature of the channel in use.
FLOW RATE
The flow rate of the channel in use.
LAST CONC
The sodium concentration of the last channel measured.
CONCENTRATION
The sodium concentration of the channel in use.
RAW POTENTIAL
The real time mV signal. The potential between the two electrodes.
AVERAGE POTENTIAL
The six second average (approximately) of the mV signal.
COMPENSATED POTENTIAL The temperature compensated mV value (potential) at 25 °C.
MEAS STABLE
Identifies if the measurement is stable (0 to 100). The larger the value, the more
stable the measurement.
pH
4
The adjusted pH of the channel in use.
4
The pH value does not show if the optional cationic pump is installed.
English
25
Содержание NA5600 sc Na+
Страница 1: ...DOC023 97 80591 NA5600 sc Na 02 2020 Edition 3 Operations Fonctionnement Operaciones Opera es...
Страница 2: ...Table of Contents English 3 Fran ais 30 Espa ol 59 Portugu s 88 118 145 172 2...
Страница 119: ...5 2 6 138 1 1 3 2 4 1 2 1 3 3 1 2 1 2 2 119...
Страница 120: ...2 1 6 diag DIAG TEST 2 122 7 cal Calibrate 3 8 back 4 home 9 menu 5 enter 3 2 3 1 1 diag DIAGNOSTICS 4 120...
Страница 132: ...6 1 4 7 2 5 8 3 6 X 9 Y 132...
Страница 134: ...8 1 Y 3 5 OnMax 2 4 X 6 OffMax 9 OnMin OffMin 1 3 X 5 OnMin 2 4 OffMin 6 Y 134...
Страница 146: ...1 1 3 2 4 1 2 1 3 3 1 2 2 1 6 Diag 2 148 7 Cal 3 8 Back 4 Home 9 Menu 5 Enter 1 2 146...
Страница 159: ...6 1 4 7 2 5 8 3 6 x 9 y 159...
Страница 161: ...8 1 y 3 5 OnMax 2 4 x 6 OffMax 9 OnMin OffMin 1 3 x 5 OnMin 2 4 OffMin 6 y 161...
Страница 198: ......
Страница 199: ......