Figure 5 Alarm function
1
High alarm
4
Low alarm
7
Time (x-axis)
2
High deadband
5
ON delay
8
Source (y-axis)
3
Low deadband
6
OFF delay
•
FEEDER CONTROL function
(refer to
Option
Description
PHASE
HIGH
—Sets the relay to on when the measured value is more than the setpoint value.
LOW
—Sets the relay to on when the measured value is less than the setpoint value.
SET SETPOINT
Sets a measurement value as the setpoint value.
DEADBAND
Sets the deadband value for the relay. If PHASE is set to LOW, the relay stays on until
the measurement value increases to more than the setpoint value plus the deadband
value. If PHASE is set to HIGH, the relay stays on until the measurement value
decreases to less than the setpoint value minus the deadband value.
OVERFEED TIMER
Sets a time limit for how long the relay can stay on. Once an overfeed alarm is present,
it must be manually reset. Refer to
Reset the overfeed timer for relays
on page 21.
OFF DELAY
Sets a delay time before the relay is set to off.
ON DELAY
Sets a delay time before the relay is set to on.
16
English
Содержание NA5600 sc Na+
Страница 1: ...DOC023 97 80591 NA5600 sc Na 02 2020 Edition 3 Operations Fonctionnement Operaciones Opera es...
Страница 2: ...Table of Contents English 3 Fran ais 30 Espa ol 59 Portugu s 88 118 145 172 2...
Страница 119: ...5 2 6 138 1 1 3 2 4 1 2 1 3 3 1 2 1 2 2 119...
Страница 120: ...2 1 6 diag DIAG TEST 2 122 7 cal Calibrate 3 8 back 4 home 9 menu 5 enter 3 2 3 1 1 diag DIAGNOSTICS 4 120...
Страница 132: ...6 1 4 7 2 5 8 3 6 X 9 Y 132...
Страница 134: ...8 1 Y 3 5 OnMax 2 4 X 6 OffMax 9 OnMin OffMin 1 3 X 5 OnMin 2 4 OffMin 6 Y 134...
Страница 146: ...1 1 3 2 4 1 2 1 3 3 1 2 2 1 6 Diag 2 148 7 Cal 3 8 Back 4 Home 9 Menu 5 Enter 1 2 146...
Страница 159: ...6 1 4 7 2 5 8 3 6 x 9 y 159...
Страница 161: ...8 1 y 3 5 OnMax 2 4 x 6 OffMax 9 OnMin OffMin 1 3 x 5 OnMin 2 4 OffMin 6 y 161...
Страница 198: ......
Страница 199: ......