Section 8 Maintenance

Regular maintenance is necessary for the best accuracy, stabilization time and life of the probe. Fully
rinse the probe with deionized water between measurements to keep the probe clean.

8.1 Clean the probe

Rinse the probe with deionized water regularly to remove mineral or sample buildup on the sensing
element. Symptoms of contamination:

• Incorrect or irregular readings
• Slow stabilization times
• Calibration errors
• Sample material stays on the probe

1.

Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the tip of the
probe.

2.

Polish the white reference junction on the probe tip with a soft cloth or cotton swab. Refer to 

Product overview

 on page 3. Do not polish the PVC sensing element in the center of the probe

tip. Do not use cleaning solutions. Make sure to fully rinse the probe with deionized water
between sample measurements to prevent mineral or sample buildup on the sensing element.

3.

Soak the probe for 30 minutes in a 100-mg/L ammonium standard solution.

8.2 Storage

Keep the sensor protection cap on the probe during storage to prevent damage to the sensing
element.

Note: Make sure to condition the probe after long-term storage. Refer to 

Preparation for use

 on page 5.

1.

Rinse the probe with deionized water. Dry the probe with a lint-free cloth.

2.

Make sure that the probe sensor is dry, then install the sensor protection cap. Refer to 

Product

overview

 on page 3.

Section 9 Troubleshooting

Refer to 

Table 2

 for general troubleshooting information. To check the probe performance, refer to 

Slope check

 on page 11. To check the accuracy of sample measurements, refer to 

Standard

additions check

 on page 11.

Table 2   Troubleshooting information

Problem

Possible cause

Solution

Decreased probe performance
causes slow stabilization and
prevents accurate calibrations or
measurements.

The probe has contamination
on the sensing element.

Rinse and condition the probe.
Refer to 

Clean the probe

on page 10.

The reference junction is
clogged.

Fully rinse the reference junction
with deionized water. Shake the
probe down to remove air
bubbles.

The probe is not conditioned
to the sample sufficiently.

Condition the probe. Refer to 

Preparation for use

 on page 5.

10

   

English

Содержание ISENH4181

Страница 1: ...DOC022 53 80029 ISENH4181 05 2021 Edition 5 User Manual...

Страница 2: ......

Страница 3: ...n 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 8 6 4 Interferences...

Страница 4: ...Table of Contents 2...

Страница 5: ...2 Specifications Specifications are subject to change without notice Specifications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sen...

Страница 6: ...r Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellic...

Страница 7: ...probe before use To condition the probe soak the probe in approximately 25 mL of the lowest concentration standard solution used for calibration for 30 to 60 minutes Do not add the ISA to the standard...

Страница 8: ...e 1 Go to the calibrate menu Select the probe if applicable The display shows the standard solutions to use for calibration 2 Add 25 mL of each standard solution to different beakers 3 Add one Ammoniu...

Страница 9: ...lete traceability is necessary enter a sample ID and operator ID before measurement Refer to the meter manual for instructions Stir the standards and samples at a slow and constant rate to prevent the...

Страница 10: ...ow apply The sample does not contain interferences The sample pH is within the range given in the specifications Omission of the ISA is accepted by the regulatory reporting agency if the measurement i...

Страница 11: ...o the acceptance criteria 7 1 Verification procedure 1 Go to the verification menu The display shows the standard solution to use for verification Note Menu name for HQd meters Run check standard 2 Po...

Страница 12: ...rd solution 8 2 Storage Keep the sensor protection cap on the probe during storage to prevent damage to the sensing element Note Make sure to condition the probe after long term storage Refer to Prepa...

Страница 13: ...contamination Use the specified standard solution of good quality Too much time passed after the ISA was added Measure the sample or standard solutions immediately after the ISA is added 9 1 Slope ch...

Страница 14: ...tandard addition A percent recovery of 100 5 is an indication that the sample measurements are accurate Percent recovery CM CE x 100 Where CM measured concentration of the sample after the addition of...

Страница 15: ...scription Quantity Item no Beaker polypropylene 50 mL low form 1 108041 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard Intellical probes 1...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 rout...

Отзывы: