6.3 Low-level measurements

Use the guidelines that follow for measurements at low concentrations (<1 mg/L NH

4

+

–N).

• Fully rinse the probe with deionized water and blot dry between measurements.
• Soak the probe in the lowest concentration standard solution for 1 hour maximum before

calibration and measurement.

• Use the meter to set the stability criteria for the probe to a low value.
• Stir the standards and samples at a slow and steady rate to prevent the formation of a vortex.
• Use a dilute ionic strength adjustor (ISA) solution for calibration and measurements:

1.

Dissolve the contents of one ionic strength adjustor powder pillow in 50 mL of deionized water.

2.

Add 5 mL of the dilute (ISA) solution to each 25 mL of standard solution or sample.

Note: The ISA is not necessary when all of the conditions that follow apply:

• The sample does not contain interferences.
• The sample pH is within the range given in the specifications.
• Omission of the ISA is accepted by the regulatory reporting agency (if the measurement is for regulatory

reporting).

6.4 Interferences

The sensing element will measure some other ions that are known to interfere with the method. The
probe response to other ions usually increases the mV potential and causes a positive error. The
response to other ions can be semi-quantitatively calculated through the Nikolsky equation, an
extended Nernst equation:
E = Eº + (RT/(zF))ln[aN

a

 + KN

a

x × ax]

Where

• ax = the activity of the interfering ion
• KN

a

x = the selectivity coefficient for the interfering ion relative to the primary ion

The primary interferences for ammonium ion-selective electrodes are perchlorate (ClO

4

), chlorate

(ClO

3

) and iodide (I

). There is no procedure to remove the interference from these ions. Other

interferences are shown in 

Table 1

. Addition of the Ammonium ISA adjusts the pH and prevents most

of the interferences from 

Table 1

.

The selectivity coefficient is the approximate apparent increase in the measured concentration that is
caused by one unit of the interfering ion (e.g., 1 unit of K

+

 increases the ammonium concentration by

0.1). The approximate selectivity coefficients for some ions with the Intellical Ammonium ISE are
shown in 

Table 1

.

8

   

English

Содержание ISENH4181

Страница 1: ...DOC022 53 80029 ISENH4181 05 2021 Edition 5 User Manual...

Страница 2: ......

Страница 3: ...n 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 8 6 4 Interferences...

Страница 4: ...Table of Contents 2...

Страница 5: ...2 Specifications Specifications are subject to change without notice Specifications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sen...

Страница 6: ...r Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellic...

Страница 7: ...probe before use To condition the probe soak the probe in approximately 25 mL of the lowest concentration standard solution used for calibration for 30 to 60 minutes Do not add the ISA to the standard...

Страница 8: ...e 1 Go to the calibrate menu Select the probe if applicable The display shows the standard solutions to use for calibration 2 Add 25 mL of each standard solution to different beakers 3 Add one Ammoniu...

Страница 9: ...lete traceability is necessary enter a sample ID and operator ID before measurement Refer to the meter manual for instructions Stir the standards and samples at a slow and constant rate to prevent the...

Страница 10: ...ow apply The sample does not contain interferences The sample pH is within the range given in the specifications Omission of the ISA is accepted by the regulatory reporting agency if the measurement i...

Страница 11: ...o the acceptance criteria 7 1 Verification procedure 1 Go to the verification menu The display shows the standard solution to use for verification Note Menu name for HQd meters Run check standard 2 Po...

Страница 12: ...rd solution 8 2 Storage Keep the sensor protection cap on the probe during storage to prevent damage to the sensing element Note Make sure to condition the probe after long term storage Refer to Prepa...

Страница 13: ...contamination Use the specified standard solution of good quality Too much time passed after the ISA was added Measure the sample or standard solutions immediately after the ISA is added 9 1 Slope ch...

Страница 14: ...tandard addition A percent recovery of 100 5 is an indication that the sample measurements are accurate Percent recovery CM CE x 100 Where CM measured concentration of the sample after the addition of...

Страница 15: ...scription Quantity Item no Beaker polypropylene 50 mL low form 1 108041 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard Intellical probes 1...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 rout...

Отзывы: