3.3 Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.

Electrical equipment marked with this symbol may not be disposed of in European
domestic or public disposal systems. Return old or end-of-life equipment to the
manufacturer for disposal at no charge to the user.

3.4 Product hazards

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.

C A U T I O N

 

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.

Section 4 Preparation for use

Prepare the probe for calibration and measurement as follows.

1.

Remove the sensor protection cap from the probe.

2.

Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not wipe the tip of the
probe.

3.

Condition the probe before use. To condition the probe, soak the probe in approximately 25 mL of
the lowest concentration standard solution used for calibration for 30 to 60 minutes. Do not add
the ISA to the standard solution.

Note: Condition the probe each day for 15 to 30 minutes or for 30 to 60 minutes after long-term storage.

4.

Make sure that the meter has the correct date and time settings. The service-life time stamp in
the probe comes from the date and time settings in the meter.

Note: Some meters automatically open the date and time settings when the meter starts for the first time, or
after battery replacement.

5.

Connect the probe to the meter.

Section 5 Calibration

The procedure that follows is applicable to meters that can connect to Intellical ISE probes. Refer to
the applicable meter documentation for meter operation and probe-specific settings.

5.1 Calibration notes

Read the notes that follow before calibration.

• Measure the standard solutions from lowest to highest concentration for best results.
• Keep all of the solutions (standard solutions and samples) at the same temperature (± 2 °C

(± 3.6 °F)) for best results.

• Stir the standards and samples at a slow and constant rate to prevent the formation of a vortex.
• Use the default calibration options or change the options in the probe settings menu.

English

   

5

Содержание ISENH4181

Страница 1: ...DOC022 53 80029 ISENH4181 05 2021 Edition 5 User Manual...

Страница 2: ......

Страница 3: ...n 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 8 6 4 Interferences...

Страница 4: ...Table of Contents 2...

Страница 5: ...2 Specifications Specifications are subject to change without notice Specifications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sen...

Страница 6: ...r Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellic...

Страница 7: ...probe before use To condition the probe soak the probe in approximately 25 mL of the lowest concentration standard solution used for calibration for 30 to 60 minutes Do not add the ISA to the standard...

Страница 8: ...e 1 Go to the calibrate menu Select the probe if applicable The display shows the standard solutions to use for calibration 2 Add 25 mL of each standard solution to different beakers 3 Add one Ammoniu...

Страница 9: ...lete traceability is necessary enter a sample ID and operator ID before measurement Refer to the meter manual for instructions Stir the standards and samples at a slow and constant rate to prevent the...

Страница 10: ...ow apply The sample does not contain interferences The sample pH is within the range given in the specifications Omission of the ISA is accepted by the regulatory reporting agency if the measurement i...

Страница 11: ...o the acceptance criteria 7 1 Verification procedure 1 Go to the verification menu The display shows the standard solution to use for verification Note Menu name for HQd meters Run check standard 2 Po...

Страница 12: ...rd solution 8 2 Storage Keep the sensor protection cap on the probe during storage to prevent damage to the sensing element Note Make sure to condition the probe after long term storage Refer to Prepa...

Страница 13: ...contamination Use the specified standard solution of good quality Too much time passed after the ISA was added Measure the sample or standard solutions immediately after the ISA is added 9 1 Slope ch...

Страница 14: ...tandard addition A percent recovery of 100 5 is an indication that the sample measurements are accurate Percent recovery CM CE x 100 Where CM measured concentration of the sample after the addition of...

Страница 15: ...scription Quantity Item no Beaker polypropylene 50 mL low form 1 108041 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard Intellical probes 1...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 rout...

Отзывы: